Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT pe...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292687/ https://www.ncbi.nlm.nih.gov/pubmed/34109738 http://dx.doi.org/10.1002/acm2.13317 |
_version_ | 1783724874438541312 |
---|---|
author | Lasiyah, Nani Anam, Choirul Hidayanto, Eko Dougherty, Geoff |
author_facet | Lasiyah, Nani Anam, Choirul Hidayanto, Eko Dougherty, Geoff |
author_sort | Lasiyah, Nani |
collection | PubMed |
description | PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full‐width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso‐center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers. RESULTS: The differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso‐center position by up to 4 cm above and 4 cm below the iso‐center. The automated results were similar (to within 0.1 mm) for various reconstruction filters. CONCLUSIONS: We successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness. |
format | Online Article Text |
id | pubmed-8292687 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-82926872021-07-22 Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter Lasiyah, Nani Anam, Choirul Hidayanto, Eko Dougherty, Geoff J Appl Clin Med Phys Technical Note PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full‐width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso‐center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers. RESULTS: The differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso‐center position by up to 4 cm above and 4 cm below the iso‐center. The automated results were similar (to within 0.1 mm) for various reconstruction filters. CONCLUSIONS: We successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness. John Wiley and Sons Inc. 2021-06-09 /pmc/articles/PMC8292687/ /pubmed/34109738 http://dx.doi.org/10.1002/acm2.13317 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Lasiyah, Nani Anam, Choirul Hidayanto, Eko Dougherty, Geoff Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title | Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title_full | Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title_fullStr | Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title_full_unstemmed | Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title_short | Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter |
title_sort | automated procedure for slice thickness verification of computed tomography images: variations of slice thickness, position from iso‐center, and reconstruction filter |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292687/ https://www.ncbi.nlm.nih.gov/pubmed/34109738 http://dx.doi.org/10.1002/acm2.13317 |
work_keys_str_mv | AT lasiyahnani automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter AT anamchoirul automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter AT hidayantoeko automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter AT doughertygeoff automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter |