Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter

PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT pe...

Descripción completa

Detalles Bibliográficos
Autores principales: Lasiyah, Nani, Anam, Choirul, Hidayanto, Eko, Dougherty, Geoff
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292687/
https://www.ncbi.nlm.nih.gov/pubmed/34109738
http://dx.doi.org/10.1002/acm2.13317
_version_ 1783724874438541312
author Lasiyah, Nani
Anam, Choirul
Hidayanto, Eko
Dougherty, Geoff
author_facet Lasiyah, Nani
Anam, Choirul
Hidayanto, Eko
Dougherty, Geoff
author_sort Lasiyah, Nani
collection PubMed
description PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full‐width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso‐center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers. RESULTS: The differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso‐center position by up to 4 cm above and 4 cm below the iso‐center. The automated results were similar (to within 0.1 mm) for various reconstruction filters. CONCLUSIONS: We successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness.
format Online
Article
Text
id pubmed-8292687
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-82926872021-07-22 Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter Lasiyah, Nani Anam, Choirul Hidayanto, Eko Dougherty, Geoff J Appl Clin Med Phys Technical Note PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso‐center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full‐width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso‐center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers. RESULTS: The differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso‐center position by up to 4 cm above and 4 cm below the iso‐center. The automated results were similar (to within 0.1 mm) for various reconstruction filters. CONCLUSIONS: We successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness. John Wiley and Sons Inc. 2021-06-09 /pmc/articles/PMC8292687/ /pubmed/34109738 http://dx.doi.org/10.1002/acm2.13317 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Lasiyah, Nani
Anam, Choirul
Hidayanto, Eko
Dougherty, Geoff
Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title_full Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title_fullStr Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title_full_unstemmed Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title_short Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso‐center, and reconstruction filter
title_sort automated procedure for slice thickness verification of computed tomography images: variations of slice thickness, position from iso‐center, and reconstruction filter
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292687/
https://www.ncbi.nlm.nih.gov/pubmed/34109738
http://dx.doi.org/10.1002/acm2.13317
work_keys_str_mv AT lasiyahnani automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter
AT anamchoirul automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter
AT hidayantoeko automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter
AT doughertygeoff automatedprocedureforslicethicknessverificationofcomputedtomographyimagesvariationsofslicethicknesspositionfromisocenterandreconstructionfilter