A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocent...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292690/ https://www.ncbi.nlm.nih.gov/pubmed/34056850 http://dx.doi.org/10.1002/acm2.13257 |
_version_ | 1783724875176738816 |
---|---|
author | Chojnowski, Jacek M. Sykes, Jonathan R. Thwaites, David I. |
author_facet | Chojnowski, Jacek M. Sykes, Jonathan R. Thwaites, David I. |
author_sort | Chojnowski, Jacek M. |
collection | PubMed |
description | The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocenter, but its method of localization is not strictly defined, and it depends on the linac‐specific beam steering parameters. A novel method is presented for determining the linac mechanical isocenter position and size based on the localization of the collimator axis of rotation at arbitrary gantry angle. The collimator axis of rotation position is determined from the radiation beam center position corrected for the focal spot offset. The focal spot offset is determined using the image center shift method with a custom‐design rigid phantom with two sets of ball‐bearings. Three specific quality assurance (QA) applications and assessment methods are also presented to demonstrate the functionality of linac mechanical isocenter position and size determination in clinical practice. The first is a mechanical and radiation isocenters coincidence test suitable for quick congruence assessment of these two isocenters for a selected energy, usually required after a nonroutine linac repair and/or energy adjustment. The second is a stereotactic beam isocentricity assessment suitable for pretreatment stereotactic QA. The third is a comprehensive linac geometrical performance test suitable for routine linac QA. The uncertainties of the method for determining mechanical isocenter position and size were measured to be 0.05 mm and 0.04 mm, respectively, using four available photon energies, and were significantly smaller than those of determining the radiation isocenter position and size, which were 0.36 mm and 0.12 mm respectively. It is therefore recommended that the mechanical isocenter position and size be used as the reference linac treatment isocenter and a linac mechanical characteristic parameter respectively. |
format | Online Article Text |
id | pubmed-8292690 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-82926902021-07-22 A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications Chojnowski, Jacek M. Sykes, Jonathan R. Thwaites, David I. J Appl Clin Med Phys Radiation Oncology Physics The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocenter, but its method of localization is not strictly defined, and it depends on the linac‐specific beam steering parameters. A novel method is presented for determining the linac mechanical isocenter position and size based on the localization of the collimator axis of rotation at arbitrary gantry angle. The collimator axis of rotation position is determined from the radiation beam center position corrected for the focal spot offset. The focal spot offset is determined using the image center shift method with a custom‐design rigid phantom with two sets of ball‐bearings. Three specific quality assurance (QA) applications and assessment methods are also presented to demonstrate the functionality of linac mechanical isocenter position and size determination in clinical practice. The first is a mechanical and radiation isocenters coincidence test suitable for quick congruence assessment of these two isocenters for a selected energy, usually required after a nonroutine linac repair and/or energy adjustment. The second is a stereotactic beam isocentricity assessment suitable for pretreatment stereotactic QA. The third is a comprehensive linac geometrical performance test suitable for routine linac QA. The uncertainties of the method for determining mechanical isocenter position and size were measured to be 0.05 mm and 0.04 mm, respectively, using four available photon energies, and were significantly smaller than those of determining the radiation isocenter position and size, which were 0.36 mm and 0.12 mm respectively. It is therefore recommended that the mechanical isocenter position and size be used as the reference linac treatment isocenter and a linac mechanical characteristic parameter respectively. John Wiley and Sons Inc. 2021-05-30 /pmc/articles/PMC8292690/ /pubmed/34056850 http://dx.doi.org/10.1002/acm2.13257 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Chojnowski, Jacek M. Sykes, Jonathan R. Thwaites, David I. A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title | A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title_full | A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title_fullStr | A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title_full_unstemmed | A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title_short | A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications |
title_sort | novel method to determine linac mechanical isocenter position and size and examples of specific qa applications |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292690/ https://www.ncbi.nlm.nih.gov/pubmed/34056850 http://dx.doi.org/10.1002/acm2.13257 |
work_keys_str_mv | AT chojnowskijacekm anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications AT sykesjonathanr anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications AT thwaitesdavidi anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications AT chojnowskijacekm novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications AT sykesjonathanr novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications AT thwaitesdavidi novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications |