A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications

The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocent...

Descripción completa

Detalles Bibliográficos
Autores principales: Chojnowski, Jacek M., Sykes, Jonathan R., Thwaites, David I.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292690/
https://www.ncbi.nlm.nih.gov/pubmed/34056850
http://dx.doi.org/10.1002/acm2.13257
_version_ 1783724875176738816
author Chojnowski, Jacek M.
Sykes, Jonathan R.
Thwaites, David I.
author_facet Chojnowski, Jacek M.
Sykes, Jonathan R.
Thwaites, David I.
author_sort Chojnowski, Jacek M.
collection PubMed
description The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocenter, but its method of localization is not strictly defined, and it depends on the linac‐specific beam steering parameters. A novel method is presented for determining the linac mechanical isocenter position and size based on the localization of the collimator axis of rotation at arbitrary gantry angle. The collimator axis of rotation position is determined from the radiation beam center position corrected for the focal spot offset. The focal spot offset is determined using the image center shift method with a custom‐design rigid phantom with two sets of ball‐bearings. Three specific quality assurance (QA) applications and assessment methods are also presented to demonstrate the functionality of linac mechanical isocenter position and size determination in clinical practice. The first is a mechanical and radiation isocenters coincidence test suitable for quick congruence assessment of these two isocenters for a selected energy, usually required after a nonroutine linac repair and/or energy adjustment. The second is a stereotactic beam isocentricity assessment suitable for pretreatment stereotactic QA. The third is a comprehensive linac geometrical performance test suitable for routine linac QA. The uncertainties of the method for determining mechanical isocenter position and size were measured to be 0.05 mm and 0.04 mm, respectively, using four available photon energies, and were significantly smaller than those of determining the radiation isocenter position and size, which were 0.36 mm and 0.12 mm respectively. It is therefore recommended that the mechanical isocenter position and size be used as the reference linac treatment isocenter and a linac mechanical characteristic parameter respectively.
format Online
Article
Text
id pubmed-8292690
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-82926902021-07-22 A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications Chojnowski, Jacek M. Sykes, Jonathan R. Thwaites, David I. J Appl Clin Med Phys Radiation Oncology Physics The most important geometric characteristic of stereotactic treatment is the accuracy of positioning the target at the treatment isocenter and the accuracy of directing the radiation beam at the treatment isocenter. Commonly, the radiation isocenter is used as the reference for the treatment isocenter, but its method of localization is not strictly defined, and it depends on the linac‐specific beam steering parameters. A novel method is presented for determining the linac mechanical isocenter position and size based on the localization of the collimator axis of rotation at arbitrary gantry angle. The collimator axis of rotation position is determined from the radiation beam center position corrected for the focal spot offset. The focal spot offset is determined using the image center shift method with a custom‐design rigid phantom with two sets of ball‐bearings. Three specific quality assurance (QA) applications and assessment methods are also presented to demonstrate the functionality of linac mechanical isocenter position and size determination in clinical practice. The first is a mechanical and radiation isocenters coincidence test suitable for quick congruence assessment of these two isocenters for a selected energy, usually required after a nonroutine linac repair and/or energy adjustment. The second is a stereotactic beam isocentricity assessment suitable for pretreatment stereotactic QA. The third is a comprehensive linac geometrical performance test suitable for routine linac QA. The uncertainties of the method for determining mechanical isocenter position and size were measured to be 0.05 mm and 0.04 mm, respectively, using four available photon energies, and were significantly smaller than those of determining the radiation isocenter position and size, which were 0.36 mm and 0.12 mm respectively. It is therefore recommended that the mechanical isocenter position and size be used as the reference linac treatment isocenter and a linac mechanical characteristic parameter respectively. John Wiley and Sons Inc. 2021-05-30 /pmc/articles/PMC8292690/ /pubmed/34056850 http://dx.doi.org/10.1002/acm2.13257 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Chojnowski, Jacek M.
Sykes, Jonathan R.
Thwaites, David I.
A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title_full A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title_fullStr A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title_full_unstemmed A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title_short A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
title_sort novel method to determine linac mechanical isocenter position and size and examples of specific qa applications
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292690/
https://www.ncbi.nlm.nih.gov/pubmed/34056850
http://dx.doi.org/10.1002/acm2.13257
work_keys_str_mv AT chojnowskijacekm anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications
AT sykesjonathanr anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications
AT thwaitesdavidi anovelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications
AT chojnowskijacekm novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications
AT sykesjonathanr novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications
AT thwaitesdavidi novelmethodtodeterminelinacmechanicalisocenterpositionandsizeandexamplesofspecificqaapplications