Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT

PURPOSE: In this study, we evaluate and compare single isocenter multiple target VMAT (SIMT) and Conformal Arc Informed VMAT (CAVMAT) radiosurgery's sensitivity to uncertainties in dosimetric leaf gap (DLG) and treatment delivery. CAVMAT is a novel planning technique that uses multiple target c...

Descripción completa

Detalles Bibliográficos
Autores principales: Cullom, Edward T., Xia, Yuqing, Chuang, Kai‐Cheng, Gude, Zachary W., Zlateva, Yana, Adamson, Justus D., Giles, William M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292691/
https://www.ncbi.nlm.nih.gov/pubmed/34165217
http://dx.doi.org/10.1002/acm2.13248
_version_ 1783724875404279808
author Cullom, Edward T.
Xia, Yuqing
Chuang, Kai‐Cheng
Gude, Zachary W.
Zlateva, Yana
Adamson, Justus D.
Giles, William M.
author_facet Cullom, Edward T.
Xia, Yuqing
Chuang, Kai‐Cheng
Gude, Zachary W.
Zlateva, Yana
Adamson, Justus D.
Giles, William M.
author_sort Cullom, Edward T.
collection PubMed
description PURPOSE: In this study, we evaluate and compare single isocenter multiple target VMAT (SIMT) and Conformal Arc Informed VMAT (CAVMAT) radiosurgery's sensitivity to uncertainties in dosimetric leaf gap (DLG) and treatment delivery. CAVMAT is a novel planning technique that uses multiple target conformal arcs as the starting point for limited inverse VMAT optimization. METHODS: All VMAT and CAVMAT plans were recalculated with DLG values of 0.4, 0.8, and 1.2 mm. DLG effect on V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc], and target dose was evaluated. Plans were delivered to a Delta(4) (ScandiDos, Madison, WI) phantom and gamma analysis performed with varying criteria. Log file analysis was performed to evaluate MLC positional error. Sixteen targets were delivered to a SRS MapCHECK (Sun Nuclear Corp., Melbourne, FL) to evaluate VMAT and CAVMAT's dose difference (DD) as a function of DLG. RESULTS: VMAT's average maximum and minimum target dose sensitivity to DLG was 9.08 ±3.50%/mm and 9.50 ± 3.30%/mm, compared to 3.20 ± 1.60%/mm and 4.72 ± 1.60%/mm for CAVMAT. For VMAT, V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc] sensitivity was 35.83 ± 9.50%/mm, 34.12 ± 6.60%/mm, and 39.23 ± 8.40%/mm. In comparison, CAVMAT's sensitivity was 23.19 ± 4.50%/mm, 22.45 ± 4.40%/mm, and 24.88 ± 4.90%/mm, respectively. Upon delivery to the Delta(4), CAVMAT offered superior dose agreement compared to VMAT. For a 1%/1 mm gamma analysis, VMAT and CAVMAT had a passing rate of 94.53 ± 4.40% and 99.28 ± 1.70%, respectively. CAVMAT was more robust to DLG variation, with the SRS MapCHECK plans yielding an absolute average DD sensitivity of 2.99 ± 1.30%/mm compared to 5.07 ± 1.10%/mm for VMAT. Log files demonstrated minimal differences in MLC positional error for both techniques. CONCLUSIONS: CAVMAT remains robust to delivery uncertainties while offering a target dose sensitivity to DLG less than half that of VMAT, and 65% of that of VMAT for V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc]. The superior dose agreement and reduced sensitivity of CAVMAT to DLG uncertainties indicate promise as a robust alternative to VMAT for SIMT SRS.
format Online
Article
Text
id pubmed-8292691
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-82926912021-07-22 Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT Cullom, Edward T. Xia, Yuqing Chuang, Kai‐Cheng Gude, Zachary W. Zlateva, Yana Adamson, Justus D. Giles, William M. J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: In this study, we evaluate and compare single isocenter multiple target VMAT (SIMT) and Conformal Arc Informed VMAT (CAVMAT) radiosurgery's sensitivity to uncertainties in dosimetric leaf gap (DLG) and treatment delivery. CAVMAT is a novel planning technique that uses multiple target conformal arcs as the starting point for limited inverse VMAT optimization. METHODS: All VMAT and CAVMAT plans were recalculated with DLG values of 0.4, 0.8, and 1.2 mm. DLG effect on V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc], and target dose was evaluated. Plans were delivered to a Delta(4) (ScandiDos, Madison, WI) phantom and gamma analysis performed with varying criteria. Log file analysis was performed to evaluate MLC positional error. Sixteen targets were delivered to a SRS MapCHECK (Sun Nuclear Corp., Melbourne, FL) to evaluate VMAT and CAVMAT's dose difference (DD) as a function of DLG. RESULTS: VMAT's average maximum and minimum target dose sensitivity to DLG was 9.08 ±3.50%/mm and 9.50 ± 3.30%/mm, compared to 3.20 ± 1.60%/mm and 4.72 ± 1.60%/mm for CAVMAT. For VMAT, V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc] sensitivity was 35.83 ± 9.50%/mm, 34.12 ± 6.60%/mm, and 39.23 ± 8.40%/mm. In comparison, CAVMAT's sensitivity was 23.19 ± 4.50%/mm, 22.45 ± 4.40%/mm, and 24.88 ± 4.90%/mm, respectively. Upon delivery to the Delta(4), CAVMAT offered superior dose agreement compared to VMAT. For a 1%/1 mm gamma analysis, VMAT and CAVMAT had a passing rate of 94.53 ± 4.40% and 99.28 ± 1.70%, respectively. CAVMAT was more robust to DLG variation, with the SRS MapCHECK plans yielding an absolute average DD sensitivity of 2.99 ± 1.30%/mm compared to 5.07 ± 1.10%/mm for VMAT. Log files demonstrated minimal differences in MLC positional error for both techniques. CONCLUSIONS: CAVMAT remains robust to delivery uncertainties while offering a target dose sensitivity to DLG less than half that of VMAT, and 65% of that of VMAT for V(6Gy)[cc], V(12Gy)[cc], and V(16Gy)[cc]. The superior dose agreement and reduced sensitivity of CAVMAT to DLG uncertainties indicate promise as a robust alternative to VMAT for SIMT SRS. John Wiley and Sons Inc. 2021-06-24 /pmc/articles/PMC8292691/ /pubmed/34165217 http://dx.doi.org/10.1002/acm2.13248 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Cullom, Edward T.
Xia, Yuqing
Chuang, Kai‐Cheng
Gude, Zachary W.
Zlateva, Yana
Adamson, Justus D.
Giles, William M.
Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title_full Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title_fullStr Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title_full_unstemmed Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title_short Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT
title_sort single isocenter srs using cavmat offers improved robustness to commissioning and treatment delivery uncertainty compared to vmat
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292691/
https://www.ncbi.nlm.nih.gov/pubmed/34165217
http://dx.doi.org/10.1002/acm2.13248
work_keys_str_mv AT cullomedwardt singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT xiayuqing singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT chuangkaicheng singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT gudezacharyw singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT zlatevayana singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT adamsonjustusd singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat
AT gileswilliamm singleisocentersrsusingcavmatoffersimprovedrobustnesstocommissioningandtreatmentdeliveryuncertaintycomparedtovmat