Cargando…

The effects of rotational setup errors in total body irradiation using helical tomotherapy

PURPOSE: Helical tomotherapy (HT) is a form of intensity‐modulated radiation therapy that is employed in total body irradiation (TBI). Because TBI targets the whole body, accurate setup positioning at the edge of the treatment volume is made difficult by the whole‐body rotational posture. The purpos...

Descripción completa

Detalles Bibliográficos
Autores principales: Isobe, Akira, Usui, Keisuke, Hara, Naoya, Sasai, Keisuke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8292714/
https://www.ncbi.nlm.nih.gov/pubmed/34028944
http://dx.doi.org/10.1002/acm2.13271
Descripción
Sumario:PURPOSE: Helical tomotherapy (HT) is a form of intensity‐modulated radiation therapy that is employed in total body irradiation (TBI). Because TBI targets the whole body, accurate setup positioning at the edge of the treatment volume is made difficult by the whole‐body rotational posture. The purpose of this study is to clarify the tolerance for rotational setup error (SE) in the vertical direction. In addition, we perform a retrospective analysis of actually irradiated dose distributions using previous patients’ irradiation data. METHODS: To clarify the effects of rotational SE on the dose distribution, the planned CT images of 10 patients were rotated by 1–5° in the vertical (pitch) direction to create a pseudo‐rotational SE image. Then, the effect of the magnitude of the rotational SE on the dose distribution was simulated. In addition, the irradiated dose to the patients was analyzed by obtaining recalculated dose distributions using megavoltage CT images acquired before treatment. RESULTS: The simulation results showed that the average value of the lung volume receiving at least 10 Gy did not exceed the allowable value when the SE value was ≤2°. When the rotational SE was ≤3°, it was possible to maintain the clinical target volume dose heterogeneity within ±10% of the prescribed dose, which is acceptable according to the guidelines. A retrospective analysis of previous patients’ irradiation data showed their daily irradiation dose distribution. The dose to the clinical target volume was reduced by up to 3.4% as a result of the residual rotational SE. Although whole‐course retrospective analyses showed a statistically significant increase in high‐dose areas, the increase was only approximately 1.0%. CONCLUSIONS: Dose errors induced by rotational SEs of ≤2° were acceptable in this study.