Convenience of Hgb-O detected by optical method in XN-series hematology analyzers in evaluating hemoglobin concentration in samples with chylous turbidity

The chylous turbidity of blood samples is one of the causes of false-high hemoglobin (Hgb) concentration measurements by the colorimetric method, which has been widely applied in hematology analyzers. In such cases, additional manual procedures are required to correct Hgb concentrations. We therefor...

Descripción completa

Detalles Bibliográficos
Autores principales: Aruga, Yu, Ikeda, Chiaki, Hanai, Arisa, Yoshimura, Sakiko, Kito, Momoko, Miyaki, Satoe, Tsubokura, Misato, Yasuno, Yuka, Hayashi, Chiaki, Miyakoshi, Motoi, Nishino, Takahiro, Kawamura, Kimihiko, Matsushita, Hiromichi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8298502/
https://www.ncbi.nlm.nih.gov/pubmed/34294786
http://dx.doi.org/10.1038/s41598-021-94394-z
Descripción
Sumario:The chylous turbidity of blood samples is one of the causes of false-high hemoglobin (Hgb) concentration measurements by the colorimetric method, which has been widely applied in hematology analyzers. In such cases, additional manual procedures are required to correct Hgb concentrations. We therefore examined the effectiveness of an optical method for measuring Hgb concentrations in samples with chylous turbidity using Hgb-O in the reticulocyte channel equipped in XN-series analyzers (Sysmex, Kobe, Japan). Hgb-O showed excellent basic performance, including linear correlation and invariability with sodium lauryl sulfate (SLS)-Hgb detected by the colorimetric method. In the analysis of samples from healthy volunteers supplemented with fat emulsion, chylous turbidity did not affect Hgb-O but SLS-Hgb, which was falsely increased according to the dose of fat emulsion. Actually, SLS-Hgb was falsely elevated in 34 of 40 chylous turbidity 3+ samples. The remaining 6 samples were measured in hematology analyzers where Hgb-O was inconsistent with SLS-Hgb in the internal quality control records. For these samples, the correction factors calculated from the internal quality control records could contribute to providing the corrected Hgb-O value. These findings suggested that the optical method was effective and convenient for accurately evaluating Hgb concentrations in samples with extremely chylous turbidity.