Cargando…

Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta

This case report underlines the complexity of the transcatheter aortic valve replacement (TAVR) procedure where rare complications sometimes are inevitable, even in experienced hands. Supra-annular dislocation of the balloon-expandable prosthesis was caused by loss of capture of the temporary transv...

Descripción completa

Detalles Bibliográficos
Autores principales: Vendrik, Jeroen, van den Boogert, Thomas P.W., Koch, Karel T., Baan, Jan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8301255/
https://www.ncbi.nlm.nih.gov/pubmed/34316759
http://dx.doi.org/10.1016/j.jaccas.2019.05.026
_version_ 1783726627036856320
author Vendrik, Jeroen
van den Boogert, Thomas P.W.
Koch, Karel T.
Baan, Jan
author_facet Vendrik, Jeroen
van den Boogert, Thomas P.W.
Koch, Karel T.
Baan, Jan
author_sort Vendrik, Jeroen
collection PubMed
description This case report underlines the complexity of the transcatheter aortic valve replacement (TAVR) procedure where rare complications sometimes are inevitable, even in experienced hands. Supra-annular dislocation of the balloon-expandable prosthesis was caused by loss of capture of the temporary transvenous pacemaker lead and treated successfully by retracting it towards the abdominal aorta. (Level of Difficulty: Advanced.)
format Online
Article
Text
id pubmed-8301255
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-83012552021-07-26 Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta Vendrik, Jeroen van den Boogert, Thomas P.W. Koch, Karel T. Baan, Jan JACC Case Rep Mini-Focus Issue: Interventional Complications and Their Management This case report underlines the complexity of the transcatheter aortic valve replacement (TAVR) procedure where rare complications sometimes are inevitable, even in experienced hands. Supra-annular dislocation of the balloon-expandable prosthesis was caused by loss of capture of the temporary transvenous pacemaker lead and treated successfully by retracting it towards the abdominal aorta. (Level of Difficulty: Advanced.) Elsevier 2019-08-21 /pmc/articles/PMC8301255/ /pubmed/34316759 http://dx.doi.org/10.1016/j.jaccas.2019.05.026 Text en © 2019 Published by Elsevier on behalf of the American College of Cardiology Foundation. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Mini-Focus Issue: Interventional Complications and Their Management
Vendrik, Jeroen
van den Boogert, Thomas P.W.
Koch, Karel T.
Baan, Jan
Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title_full Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title_fullStr Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title_full_unstemmed Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title_short Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta
title_sort balloon-expandable tavr prosthesis dislocates into the ascending aorta
topic Mini-Focus Issue: Interventional Complications and Their Management
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8301255/
https://www.ncbi.nlm.nih.gov/pubmed/34316759
http://dx.doi.org/10.1016/j.jaccas.2019.05.026
work_keys_str_mv AT vendrikjeroen balloonexpandabletavrprosthesisdislocatesintotheascendingaorta
AT vandenboogertthomaspw balloonexpandabletavrprosthesisdislocatesintotheascendingaorta
AT kochkarelt balloonexpandabletavrprosthesisdislocatesintotheascendingaorta
AT baanjan balloonexpandabletavrprosthesisdislocatesintotheascendingaorta