Cargando…
An Unusual Case of Bioprosthetic Mitral Valve Failure
We report a case of sudden-onset pulmonary edema due to failure of a bioprosthetic mitral valve. Gross inspection revealed a leaflet tear at a stent post without calcification or pannus formation and no evidence of sutures. This case highlights the mechanical failure of a bioprosthetic mitral valve...
Autores principales: | Miura, Shiro, Yamashita, Takehiro, Murata, Masaki, Iwano, Hiroyuki, Ohkawa, Youhei |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8302175/ https://www.ncbi.nlm.nih.gov/pubmed/34317020 http://dx.doi.org/10.1016/j.jaccas.2020.05.090 |
Ejemplares similares
-
Transcatheter Mitral Valve-In-Valve Implantation: An Option for Failed Bioprosthetic Mitral Valve Stenosis During Pregnancy
por: Ribeyrolles, Sophie, et al.
Publicado: (2020) -
Parachute Mitral Valve: A Case of Isolated Accessory Mitral Valve Tissue
por: Akanya, Deborah Tosin, et al.
Publicado: (2020) -
Preemptive Alcohol Septal Ablation Prior to Valve-in-Valve Transcatheter Mitral Valve Replacement With Bioprosthetic Balloon Fracture
por: Engel Gonzalez, Pedro, et al.
Publicado: (2021) -
Fluttering Bioprosthetic Valve Leaflet Detected by Intravascular Ultrasound During Valve-in-Valve Transcatheter Aortic Valve Replacement
por: Iseki, Yosei, et al.
Publicado: (2021) -
Sudden Cardiac Arrest in a Patient With Mitral Valve Prolapse and LMNA and SCN5A Mutations
por: Mahajan, Asha M., et al.
Publicado: (2021)