Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis

The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US...

Descripción completa

Detalles Bibliográficos
Autores principales: Corte, Giulia, Bayat, Sara, Tascilar, Koray, Valor-Mendez, Larissa, Schuster, Louis, Knitza, Johannes, Fagni, Filippo, Schett, Georg, Kleyer, Arnd, Simon, David
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8305786/
https://www.ncbi.nlm.nih.gov/pubmed/34206675
http://dx.doi.org/10.3390/diagnostics11071139
_version_ 1783727655395262464
author Corte, Giulia
Bayat, Sara
Tascilar, Koray
Valor-Mendez, Larissa
Schuster, Louis
Knitza, Johannes
Fagni, Filippo
Schett, Georg
Kleyer, Arnd
Simon, David
author_facet Corte, Giulia
Bayat, Sara
Tascilar, Koray
Valor-Mendez, Larissa
Schuster, Louis
Knitza, Johannes
Fagni, Filippo
Schett, Georg
Kleyer, Arnd
Simon, David
author_sort Corte, Giulia
collection PubMed
description The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US was performed on the clinically affected joints using a cart-based sonographic device (Samsung HS40) and a HHUS device (Butterfly iQ). One blinded reader scored all images for the presence of erosions, bony enlargement, synovial hypertrophy, joint effusion, bursitis, tenosynovitis, and enthesitis. Synovitis was graded (B mode and power Doppler (PD)) by the 4-level EULAR-OMERACT scale. To avoid bias by the blinded reader, we included 67 joints of two healthy volunteers in the evaluation. We calculated the overall concordance and the concordance by type of joint and pathological finding. We also measured the time required for the US examination per joint with both devices. Thirty-two patients (20 with RA, 10 with PsA, and one each with gout and SLE-associated arthritis) were included, and 186 joints were examined. The overall raw concordance in B mode was 97% (κappa 0.90, 95% CI (0.89, 0.94)). In B mode, no significant differences were found in relation to type of joint or pathological finding examined. The PD mode of the HHUS device did not detect any PD signal, whereas the cart-based device detected a PD signal in 61 joints (33%). The portable device did not offer any time savings compared to the cart-based device (47.0 versus 46.3 s). The HHUS device was accurate in the assessment of structural damage and inflammation in patients with IA, but only in the B mode. Significant improvements are still needed for HHUS to reliably demonstrate blood flow detection in PD mode.
format Online
Article
Text
id pubmed-8305786
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-83057862021-07-25 Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis Corte, Giulia Bayat, Sara Tascilar, Koray Valor-Mendez, Larissa Schuster, Louis Knitza, Johannes Fagni, Filippo Schett, Georg Kleyer, Arnd Simon, David Diagnostics (Basel) Article The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US was performed on the clinically affected joints using a cart-based sonographic device (Samsung HS40) and a HHUS device (Butterfly iQ). One blinded reader scored all images for the presence of erosions, bony enlargement, synovial hypertrophy, joint effusion, bursitis, tenosynovitis, and enthesitis. Synovitis was graded (B mode and power Doppler (PD)) by the 4-level EULAR-OMERACT scale. To avoid bias by the blinded reader, we included 67 joints of two healthy volunteers in the evaluation. We calculated the overall concordance and the concordance by type of joint and pathological finding. We also measured the time required for the US examination per joint with both devices. Thirty-two patients (20 with RA, 10 with PsA, and one each with gout and SLE-associated arthritis) were included, and 186 joints were examined. The overall raw concordance in B mode was 97% (κappa 0.90, 95% CI (0.89, 0.94)). In B mode, no significant differences were found in relation to type of joint or pathological finding examined. The PD mode of the HHUS device did not detect any PD signal, whereas the cart-based device detected a PD signal in 61 joints (33%). The portable device did not offer any time savings compared to the cart-based device (47.0 versus 46.3 s). The HHUS device was accurate in the assessment of structural damage and inflammation in patients with IA, but only in the B mode. Significant improvements are still needed for HHUS to reliably demonstrate blood flow detection in PD mode. MDPI 2021-06-22 /pmc/articles/PMC8305786/ /pubmed/34206675 http://dx.doi.org/10.3390/diagnostics11071139 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Corte, Giulia
Bayat, Sara
Tascilar, Koray
Valor-Mendez, Larissa
Schuster, Louis
Knitza, Johannes
Fagni, Filippo
Schett, Georg
Kleyer, Arnd
Simon, David
Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title_full Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title_fullStr Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title_full_unstemmed Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title_short Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
title_sort performance of a handheld ultrasound device to assess articular and periarticular pathologies in patients with inflammatory arthritis
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8305786/
https://www.ncbi.nlm.nih.gov/pubmed/34206675
http://dx.doi.org/10.3390/diagnostics11071139
work_keys_str_mv AT cortegiulia performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT bayatsara performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT tascilarkoray performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT valormendezlarissa performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT schusterlouis performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT knitzajohannes performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT fagnifilippo performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT schettgeorg performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT kleyerarnd performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis
AT simondavid performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis