Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis
The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8305786/ https://www.ncbi.nlm.nih.gov/pubmed/34206675 http://dx.doi.org/10.3390/diagnostics11071139 |
_version_ | 1783727655395262464 |
---|---|
author | Corte, Giulia Bayat, Sara Tascilar, Koray Valor-Mendez, Larissa Schuster, Louis Knitza, Johannes Fagni, Filippo Schett, Georg Kleyer, Arnd Simon, David |
author_facet | Corte, Giulia Bayat, Sara Tascilar, Koray Valor-Mendez, Larissa Schuster, Louis Knitza, Johannes Fagni, Filippo Schett, Georg Kleyer, Arnd Simon, David |
author_sort | Corte, Giulia |
collection | PubMed |
description | The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US was performed on the clinically affected joints using a cart-based sonographic device (Samsung HS40) and a HHUS device (Butterfly iQ). One blinded reader scored all images for the presence of erosions, bony enlargement, synovial hypertrophy, joint effusion, bursitis, tenosynovitis, and enthesitis. Synovitis was graded (B mode and power Doppler (PD)) by the 4-level EULAR-OMERACT scale. To avoid bias by the blinded reader, we included 67 joints of two healthy volunteers in the evaluation. We calculated the overall concordance and the concordance by type of joint and pathological finding. We also measured the time required for the US examination per joint with both devices. Thirty-two patients (20 with RA, 10 with PsA, and one each with gout and SLE-associated arthritis) were included, and 186 joints were examined. The overall raw concordance in B mode was 97% (κappa 0.90, 95% CI (0.89, 0.94)). In B mode, no significant differences were found in relation to type of joint or pathological finding examined. The PD mode of the HHUS device did not detect any PD signal, whereas the cart-based device detected a PD signal in 61 joints (33%). The portable device did not offer any time savings compared to the cart-based device (47.0 versus 46.3 s). The HHUS device was accurate in the assessment of structural damage and inflammation in patients with IA, but only in the B mode. Significant improvements are still needed for HHUS to reliably demonstrate blood flow detection in PD mode. |
format | Online Article Text |
id | pubmed-8305786 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-83057862021-07-25 Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis Corte, Giulia Bayat, Sara Tascilar, Koray Valor-Mendez, Larissa Schuster, Louis Knitza, Johannes Fagni, Filippo Schett, Georg Kleyer, Arnd Simon, David Diagnostics (Basel) Article The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound (HHUS) machine in comparison to a conventional cart-based sonographic machine in patients with inflammatory arthritis (IA). IA patients with at least one tender and swollen joint count were enrolled. US was performed on the clinically affected joints using a cart-based sonographic device (Samsung HS40) and a HHUS device (Butterfly iQ). One blinded reader scored all images for the presence of erosions, bony enlargement, synovial hypertrophy, joint effusion, bursitis, tenosynovitis, and enthesitis. Synovitis was graded (B mode and power Doppler (PD)) by the 4-level EULAR-OMERACT scale. To avoid bias by the blinded reader, we included 67 joints of two healthy volunteers in the evaluation. We calculated the overall concordance and the concordance by type of joint and pathological finding. We also measured the time required for the US examination per joint with both devices. Thirty-two patients (20 with RA, 10 with PsA, and one each with gout and SLE-associated arthritis) were included, and 186 joints were examined. The overall raw concordance in B mode was 97% (κappa 0.90, 95% CI (0.89, 0.94)). In B mode, no significant differences were found in relation to type of joint or pathological finding examined. The PD mode of the HHUS device did not detect any PD signal, whereas the cart-based device detected a PD signal in 61 joints (33%). The portable device did not offer any time savings compared to the cart-based device (47.0 versus 46.3 s). The HHUS device was accurate in the assessment of structural damage and inflammation in patients with IA, but only in the B mode. Significant improvements are still needed for HHUS to reliably demonstrate blood flow detection in PD mode. MDPI 2021-06-22 /pmc/articles/PMC8305786/ /pubmed/34206675 http://dx.doi.org/10.3390/diagnostics11071139 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Corte, Giulia Bayat, Sara Tascilar, Koray Valor-Mendez, Larissa Schuster, Louis Knitza, Johannes Fagni, Filippo Schett, Georg Kleyer, Arnd Simon, David Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title | Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title_full | Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title_fullStr | Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title_full_unstemmed | Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title_short | Performance of a Handheld Ultrasound Device to Assess Articular and Periarticular Pathologies in Patients with Inflammatory Arthritis |
title_sort | performance of a handheld ultrasound device to assess articular and periarticular pathologies in patients with inflammatory arthritis |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8305786/ https://www.ncbi.nlm.nih.gov/pubmed/34206675 http://dx.doi.org/10.3390/diagnostics11071139 |
work_keys_str_mv | AT cortegiulia performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT bayatsara performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT tascilarkoray performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT valormendezlarissa performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT schusterlouis performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT knitzajohannes performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT fagnifilippo performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT schettgeorg performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT kleyerarnd performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis AT simondavid performanceofahandheldultrasounddevicetoassessarticularandperiarticularpathologiesinpatientswithinflammatoryarthritis |