Vapourized hydrogen peroxide decontamination in a hospital setting inactivates SARS-CoV-2 and HCoV-229E without compromising filtration efficiency of unexpired N95 respirators

BACKGROUND: The COVID-19 pandemic highlighted the need for evidence-based approaches to decontamination and reuse of N95 filtering facepiece respirators (FFRs). We sought to determine whether vapourized hydrogen peroxide (VHP) reduced SARS-CoV-2 bioburden on FFRs without compromising filtration effi...

Descripción completa

Detalles Bibliográficos
Autores principales: Christie-Holmes, Natasha, Tyli, Rachel, Budylowski, Patrick, Guvenc, Furkan, Weiner, Amit, Poon, Betty, Speck, Mary, Naugler, Stephenie, Rainville, Allen, Ghalami, Ayoob, McCaw, Shannon, Hayes, Steven, Mubareka, Samira, Gray-Owen, Scott D., Rotstein, Ori D., Kandel, Rita A., Scott, James A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8310574/
https://www.ncbi.nlm.nih.gov/pubmed/34320409
http://dx.doi.org/10.1016/j.ajic.2021.07.012
Descripción
Sumario:BACKGROUND: The COVID-19 pandemic highlighted the need for evidence-based approaches to decontamination and reuse of N95 filtering facepiece respirators (FFRs). We sought to determine whether vapourized hydrogen peroxide (VHP) reduced SARS-CoV-2 bioburden on FFRs without compromising filtration efficiency. We also investigated coronavirus HCoV-229E as a surrogate for decontamination validation testing. METHODS: N95 FFRs were laced with SARS-CoV-2 or HCoV-229E and treated with VHP in a hospital reprocessing facility. After sterilization, viral burden was determined using viral outgrowth in a titration assay, and filtration efficiency of FFRs was tested against ATSM F2299 and NIOSH TEB-STP-APR-0059. RESULTS: Viable SARS-CoV-2 virus was not detected after VHP treatment. One replicate of the HCoV-229E laced FFRs yielded virus after processing. Unexpired N95 FFRs retained full filtration efficiency after VHP processing. Expired FFRs failed to meet design-specified filtration efficiency and therefore are unsuitable for reprocessing. DISCUSSION: In-hospital VHP is an effective decontaminant for SARS-CoV-2 on FFRs. Further, filtration efficiency of unexpired respirators is not affected by this decontamination process. CONCLUSIONS: VHP is effective in inactivating SARS-CoV-2 on FFRs without compromising filtration efficiency. HCoV-229E is a suitable surrogate for SARS-CoV-2 for disinfection studies.