Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport
We present and compare the designs of three types of neutron microscopes for high-resolution neutron imaging. Like optical microscopes, and unlike standard neutron imaging instruments, these microscopes have both condenser and image-forming objective optics. The optics are glancing-incidence axisymm...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8321192/ https://www.ncbi.nlm.nih.gov/pubmed/34460541 http://dx.doi.org/10.3390/jimaging6100100 |
_version_ | 1783730792422178816 |
---|---|
author | Abir, Muhammad Hussey, Daniel S. Khaykovich, Boris |
author_facet | Abir, Muhammad Hussey, Daniel S. Khaykovich, Boris |
author_sort | Abir, Muhammad |
collection | PubMed |
description | We present and compare the designs of three types of neutron microscopes for high-resolution neutron imaging. Like optical microscopes, and unlike standard neutron imaging instruments, these microscopes have both condenser and image-forming objective optics. The optics are glancing-incidence axisymmetric mirrors and therefore suitable for polychromatic neutron beams. The mirrors are designed to provide a magnification of 10 to achieve a spatial resolution of better than 10 [Formula: see text] m. The resolution of the microscopes is determined by the mirrors rather than by the L/Dratio as in conventional pinhole imaging, leading to possible dramatic improvements in the signal rate. We predict the increase in the signal rate by at least two orders of magnitude for very high-resolution imaging, which is always flux limited. Furthermore, in contrast to pinhole imaging, in the microscope, the samples are placed far from the detector to allow for a bulky sample environment without sacrificing spatial resolution. |
format | Online Article Text |
id | pubmed-8321192 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-83211922021-08-26 Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport Abir, Muhammad Hussey, Daniel S. Khaykovich, Boris J Imaging Article We present and compare the designs of three types of neutron microscopes for high-resolution neutron imaging. Like optical microscopes, and unlike standard neutron imaging instruments, these microscopes have both condenser and image-forming objective optics. The optics are glancing-incidence axisymmetric mirrors and therefore suitable for polychromatic neutron beams. The mirrors are designed to provide a magnification of 10 to achieve a spatial resolution of better than 10 [Formula: see text] m. The resolution of the microscopes is determined by the mirrors rather than by the L/Dratio as in conventional pinhole imaging, leading to possible dramatic improvements in the signal rate. We predict the increase in the signal rate by at least two orders of magnitude for very high-resolution imaging, which is always flux limited. Furthermore, in contrast to pinhole imaging, in the microscope, the samples are placed far from the detector to allow for a bulky sample environment without sacrificing spatial resolution. MDPI 2020-09-27 /pmc/articles/PMC8321192/ /pubmed/34460541 http://dx.doi.org/10.3390/jimaging6100100 Text en © 2020 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ). |
spellingShingle | Article Abir, Muhammad Hussey, Daniel S. Khaykovich, Boris Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title | Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title_full | Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title_fullStr | Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title_full_unstemmed | Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title_short | Design of Neutron Microscopes Equipped with Wolter Mirror Condenser and Objective Optics for High-Fidelity Imaging and Beam Transport |
title_sort | design of neutron microscopes equipped with wolter mirror condenser and objective optics for high-fidelity imaging and beam transport |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8321192/ https://www.ncbi.nlm.nih.gov/pubmed/34460541 http://dx.doi.org/10.3390/jimaging6100100 |
work_keys_str_mv | AT abirmuhammad designofneutronmicroscopesequippedwithwoltermirrorcondenserandobjectiveopticsforhighfidelityimagingandbeamtransport AT husseydaniels designofneutronmicroscopesequippedwithwoltermirrorcondenserandobjectiveopticsforhighfidelityimagingandbeamtransport AT khaykovichboris designofneutronmicroscopesequippedwithwoltermirrorcondenserandobjectiveopticsforhighfidelityimagingandbeamtransport |