Cargando…
Clinical Utility of Artificial Intelligence Algorithms to Enhance Wide-Field Optical Coherence Tomography Angiography Images
The aim of this paper is to investigate the clinical utility of the application of deep learning denoise algorithms on standard wide-field Optical Coherence Tomography Angiography (OCT-A) images. This was a retrospective case-series assessing forty-nine 10 × 10 mm OCT-A1 macula scans of 49 consecuti...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8321271/ https://www.ncbi.nlm.nih.gov/pubmed/34460631 http://dx.doi.org/10.3390/jimaging7020032 |
Sumario: | The aim of this paper is to investigate the clinical utility of the application of deep learning denoise algorithms on standard wide-field Optical Coherence Tomography Angiography (OCT-A) images. This was a retrospective case-series assessing forty-nine 10 × 10 mm OCT-A1 macula scans of 49 consecutive patients attending a medical retina clinic over a 6-month period. Thirty-seven patients had pathology; 13 had none. Retinal vascular layers were categorised into superficial or deep capillary plexus. For each category, the retinal experts compared the original standard image with the same image that had intelligent denoise applied. When analysing the Superficial Capillary Plexus (SCP), the denoised image was selected as “best for clinical assessment” in 98% of comparisons. No difference was established in the remaining 2%. On evaluating the Deep Capillary Plexus (DCP), the denoised image was preferred in 35% of comparisons. No difference was found in 65%. There was no evidence of new artefactual features nor loss of anatomical detail in denoised compared to the standard images. The wide-field denoise feature of the Canon Xephilio OCT-A1 produced scans that were clinically preferable over their original OCT-A images, especially for SCP assessment, without evidence for causing a new artefactual error. |
---|