Fast MCsquare-Based Independent Dose Verification Platform for Pencil Beam Scanning Proton Therapy
PURPOSE: To commission MCsquare (a multi-cores CPU-based dose calculation engine) for pencil beam scanning (PBS) proton therapy, integrate it into RayStation treatment plan system (TPS) to create a dedicated platform for fast independent dose verification. METHOD: A MCsquare-based independent dose v...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8326813/ https://www.ncbi.nlm.nih.gov/pubmed/34338058 http://dx.doi.org/10.1177/15330338211033076 |
Sumario: | PURPOSE: To commission MCsquare (a multi-cores CPU-based dose calculation engine) for pencil beam scanning (PBS) proton therapy, integrate it into RayStation treatment plan system (TPS) to create a dedicated platform for fast independent dose verification. METHOD: A MCsquare-based independent dose verification platform (MC2InRS) was developed to realize automatic dose re-calculation for clinical use, including data preparation, dose calculation, 2D/3D gamma analysis. MCsquare was commissioned based on in-air lateral dose profiles, integrated depth dose, and the absolute dose of different beam energies for Proteus(®)ONE. MC2InRS was validated with measurement data using various targets and depths in a water phantom. This study also investigated 15 clinical cases to demonstrate the feasibility and effectiveness of MC2InRS platform in clinic practice. RESULTS: Between simulation and measurement, the distal range differences at 80% (R80) and 20% (R20) dose levels for each energy were below 0.05 mm, and 0.1 mm, respectively, and the absolute dose differences were below 0.5%. 29 out of 36 QA planes reached a 100% gamma passing rate (GPR) for 2%/2mm criteria, and a minimum of 98.3% gamma was obtained in water phantom between simulation and measurement. For the 15 clinical cases investigated, the average 2D GPR (2%/2mm) was 95.4%, 99.3% for MCsquare vs. measurement, MCsquare vs. TPS, respectively. The average 3D GPR (2%/2mm) was 98.9%, 95.3% for MCsquare vs. TPS in water, and computed tomography (CT), respectively. CONCLUSION: MC2InRS, a fast, independent dose verification platform, has been developed to perform dose verification with high accuracy and efficiency for Pencil Bream Scanning (PBS). Its potential to be applied in routine clinical practice has also been discussed. |
---|