Cargando…

Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study

PURPOSE: Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-reference...

Descripción completa

Detalles Bibliográficos
Autores principales: Freigang, Viola, Gottsauner, Maximilian, Rupp, Markus, Pfeifer, Christian, Grechenig, Stephan, Kerner, Alexander, Alt, Volker, Baumann, Florian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8331300/
https://www.ncbi.nlm.nih.gov/pubmed/34355040
http://dx.doi.org/10.1155/2021/2321504
Descripción
Sumario:PURPOSE: Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP). METHODS: We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler's teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface. RESULTS: The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm. CONCLUSIONS: A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.