Pragmatic Micrometre to Millimetre Calibration Using Multiple Methods for Low-Coherence Interferometer in Embedded Metrology Applications
In-situ metrology utilised for surface topography, texture and form analysis along with quality control processes requires a high-level of reliability. Hence, a traceable method for calibrating the measurement system’s transfer function is required at regular intervals. This paper compares three met...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8347976/ https://www.ncbi.nlm.nih.gov/pubmed/34372336 http://dx.doi.org/10.3390/s21155101 |
Sumario: | In-situ metrology utilised for surface topography, texture and form analysis along with quality control processes requires a high-level of reliability. Hence, a traceable method for calibrating the measurement system’s transfer function is required at regular intervals. This paper compares three methods of dimensional calibration for a spectral domain low coherence interferometer using a reference laser interferometer versus two types of single material measure. Additionally, the impact of dataset sparsity is shown along with the effect of using a singular calibration dataset for system performance when operating across different media. |
---|