Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts

Background and objective: Cardiac magnetic resonance (CMR) is a key tool for cardiac work-up. However, arrhythmia can be responsible for arrhythmia-related artifacts (ARA) and increased scan time using segmented sequences. The aim of this study is to evaluate the effect of cardiac arrhythmia on imag...

Descripción completa

Detalles Bibliográficos
Autores principales: Longère, Benjamin, Allard, Paul-Edouard, Gkizas, Christos V, Coisne, Augustin, Hennicaux, Justin, Simeone, Arianna, Schmidt, Michaela, Forman, Christoph, Toupin, Solenn, Montaigne, David, Pontana, François
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8348071/
https://www.ncbi.nlm.nih.gov/pubmed/34362058
http://dx.doi.org/10.3390/jcm10153274
_version_ 1783735248313384960
author Longère, Benjamin
Allard, Paul-Edouard
Gkizas, Christos V
Coisne, Augustin
Hennicaux, Justin
Simeone, Arianna
Schmidt, Michaela
Forman, Christoph
Toupin, Solenn
Montaigne, David
Pontana, François
author_facet Longère, Benjamin
Allard, Paul-Edouard
Gkizas, Christos V
Coisne, Augustin
Hennicaux, Justin
Simeone, Arianna
Schmidt, Michaela
Forman, Christoph
Toupin, Solenn
Montaigne, David
Pontana, François
author_sort Longère, Benjamin
collection PubMed
description Background and objective: Cardiac magnetic resonance (CMR) is a key tool for cardiac work-up. However, arrhythmia can be responsible for arrhythmia-related artifacts (ARA) and increased scan time using segmented sequences. The aim of this study is to evaluate the effect of cardiac arrhythmia on image quality in a comparison of a compressed sensing real-time (CS(rt)) cine sequence with the reference prospectively gated segmented balanced steady-state free precession (Cine(ref)) technique regarding ARA. Methods: A total of 71 consecutive adult patients (41 males; mean age = 59.5 ± 20.1 years (95% CI: 54.7–64.2 years)) referred for CMR examination with concomitant irregular heart rate (defined by an RR interval coefficient of variation >10%) during scanning were prospectively enrolled. For each patient, two cine sequences were systematically acquired: first, the reference prospectively triggered multi-breath-hold Cine(ref) sequence including a short-axis stack, one four-chamber slice, and a couple of two-chamber slices; second, an additional single breath-hold CS(rt) sequence providing the same slices as the reference technique. Two radiologists independently assessed ARA and image quality (overall, acquisition, and edge sharpness) for both techniques. Results: The mean heart rate was 71.8 ± 19.0 (SD) beat per minute (bpm) (95% CI: 67.4–76.3 bpm) and its coefficient of variation was 25.0 ± 9.4 (SD) % (95% CI: 22.8–27.2%). Acquisition was significantly faster with CS(rt) than with Cine(ref) (Cine(ref): 556.7 ± 145.4 (SD) s (95% CI: 496.7–616.7 s); CS(rt): 23.9 ± 7.9 (SD) s (95% CI: 20.6–27.1 s); p < 0.0001). A total of 599 pairs of cine slices were evaluated (median: 8 (range: 6–14) slices per patient). The mean proportion of ARA-impaired slices per patient was 85.9 ± 22.7 (SD) % using Cine(ref), but this was figure was zero using CS(rt) (p < 0.0001). The European CMR registry artifact score was lower with CS(rt) (median: 1 (range: 0–5)) than with Cine(ref) (median: 3 (range: 0–3); p < 0.0001). Subjective image quality was higher in CS(rt) than in Cine(ref) (median: 3 (range: 1–3) versus 2 (range: 1–4), respectively; p < 0.0001). In line, edge sharpness was higher on CS(rt) cine than on Cine(ref) images (0.054 ± 0.016 pixel(−1) (95% CI: 0.050–0.057 pixel(−1)) versus 0.042 ± 0.022 pixel(−1) (95% CI: 0.037–0.047 pixel(−1)), respectively; p = 0.0001). Conclusion: Compressed sensing real-time cine drastically reduces arrhythmia-related artifacts and thus improves cine image quality in patients with arrhythmia.
format Online
Article
Text
id pubmed-8348071
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-83480712021-08-08 Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts Longère, Benjamin Allard, Paul-Edouard Gkizas, Christos V Coisne, Augustin Hennicaux, Justin Simeone, Arianna Schmidt, Michaela Forman, Christoph Toupin, Solenn Montaigne, David Pontana, François J Clin Med Article Background and objective: Cardiac magnetic resonance (CMR) is a key tool for cardiac work-up. However, arrhythmia can be responsible for arrhythmia-related artifacts (ARA) and increased scan time using segmented sequences. The aim of this study is to evaluate the effect of cardiac arrhythmia on image quality in a comparison of a compressed sensing real-time (CS(rt)) cine sequence with the reference prospectively gated segmented balanced steady-state free precession (Cine(ref)) technique regarding ARA. Methods: A total of 71 consecutive adult patients (41 males; mean age = 59.5 ± 20.1 years (95% CI: 54.7–64.2 years)) referred for CMR examination with concomitant irregular heart rate (defined by an RR interval coefficient of variation >10%) during scanning were prospectively enrolled. For each patient, two cine sequences were systematically acquired: first, the reference prospectively triggered multi-breath-hold Cine(ref) sequence including a short-axis stack, one four-chamber slice, and a couple of two-chamber slices; second, an additional single breath-hold CS(rt) sequence providing the same slices as the reference technique. Two radiologists independently assessed ARA and image quality (overall, acquisition, and edge sharpness) for both techniques. Results: The mean heart rate was 71.8 ± 19.0 (SD) beat per minute (bpm) (95% CI: 67.4–76.3 bpm) and its coefficient of variation was 25.0 ± 9.4 (SD) % (95% CI: 22.8–27.2%). Acquisition was significantly faster with CS(rt) than with Cine(ref) (Cine(ref): 556.7 ± 145.4 (SD) s (95% CI: 496.7–616.7 s); CS(rt): 23.9 ± 7.9 (SD) s (95% CI: 20.6–27.1 s); p < 0.0001). A total of 599 pairs of cine slices were evaluated (median: 8 (range: 6–14) slices per patient). The mean proportion of ARA-impaired slices per patient was 85.9 ± 22.7 (SD) % using Cine(ref), but this was figure was zero using CS(rt) (p < 0.0001). The European CMR registry artifact score was lower with CS(rt) (median: 1 (range: 0–5)) than with Cine(ref) (median: 3 (range: 0–3); p < 0.0001). Subjective image quality was higher in CS(rt) than in Cine(ref) (median: 3 (range: 1–3) versus 2 (range: 1–4), respectively; p < 0.0001). In line, edge sharpness was higher on CS(rt) cine than on Cine(ref) images (0.054 ± 0.016 pixel(−1) (95% CI: 0.050–0.057 pixel(−1)) versus 0.042 ± 0.022 pixel(−1) (95% CI: 0.037–0.047 pixel(−1)), respectively; p = 0.0001). Conclusion: Compressed sensing real-time cine drastically reduces arrhythmia-related artifacts and thus improves cine image quality in patients with arrhythmia. MDPI 2021-07-24 /pmc/articles/PMC8348071/ /pubmed/34362058 http://dx.doi.org/10.3390/jcm10153274 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Longère, Benjamin
Allard, Paul-Edouard
Gkizas, Christos V
Coisne, Augustin
Hennicaux, Justin
Simeone, Arianna
Schmidt, Michaela
Forman, Christoph
Toupin, Solenn
Montaigne, David
Pontana, François
Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title_full Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title_fullStr Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title_full_unstemmed Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title_short Compressed Sensing Real-Time Cine Reduces CMR Arrhythmia-Related Artifacts
title_sort compressed sensing real-time cine reduces cmr arrhythmia-related artifacts
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8348071/
https://www.ncbi.nlm.nih.gov/pubmed/34362058
http://dx.doi.org/10.3390/jcm10153274
work_keys_str_mv AT longerebenjamin compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT allardpauledouard compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT gkizaschristosv compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT coisneaugustin compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT hennicauxjustin compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT simeonearianna compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT schmidtmichaela compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT formanchristoph compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT toupinsolenn compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT montaignedavid compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts
AT pontanafrancois compressedsensingrealtimecinereducescmrarrhythmiarelatedartifacts