Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers

BACKGROUND: To conserve personal protective equipment (PPE) and reduce exposure to potentially infected COVID‐19 patients, several Californian facilities independently implemented a method of acquiring portable chest radiographs through glass barriers that was originally developed by the University...

Descripción completa

Detalles Bibliográficos
Autores principales: McKenney, Sarah E., Wait, John M. S., Cooper, Virgil N., Johnson, Amirh M., Wang, Jia, Leung, Ann N., Clements, Jessica
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8364281/
https://www.ncbi.nlm.nih.gov/pubmed/34216091
http://dx.doi.org/10.1002/acm2.13330
_version_ 1783738500693098496
author McKenney, Sarah E.
Wait, John M. S.
Cooper, Virgil N.
Johnson, Amirh M.
Wang, Jia
Leung, Ann N.
Clements, Jessica
author_facet McKenney, Sarah E.
Wait, John M. S.
Cooper, Virgil N.
Johnson, Amirh M.
Wang, Jia
Leung, Ann N.
Clements, Jessica
author_sort McKenney, Sarah E.
collection PubMed
description BACKGROUND: To conserve personal protective equipment (PPE) and reduce exposure to potentially infected COVID‐19 patients, several Californian facilities independently implemented a method of acquiring portable chest radiographs through glass barriers that was originally developed by the University of Washington. METHODS: This work quantifies the transmission of radiation through a glass barrier using six radiographic systems at five facilities. Patient entrance air kerma (EAK) and effective dose were estimated both with and without the glass barrier. Beam penetrability and resulting exposure index (EI) and deviation index (DI) were measured and used to adjust the tube current‐time product (mAs) for glass barriers. Because of beam hardening, the contrast‐to‐noise ratio (CNR) was measured with image quality phantoms to ensure diagnostic integrity. Finally, scatter surveys were performed to assess staff radiation exposure both inside and outside the exam room. RESULTS: The glass barriers attenuated a mean of 61% of the normal X‐ray beams. When the mAs was increased to match EI values, there was no discernible degradation of image quality as determined by the CNR. This was corroborated with subjective assessments of image quality by chest radiologists. The glass‐hardened beams acted as a filter for low energy X‐rays, and some facilities observed slight changes in patient effective doses. There was scattering from both the phantoms and the glass barriers within the room. CONCLUSIONS: Glass barriers require an approximate 2.5 times increase in beam intensity, with all other technique factors held constant. Further refinements are necessary for increased source‐to‐image distance and beam quality in order to adequately match EI values. This does not result in a significant increase in the radiation dose delivered to the patient. The use of lead aprons, mobile shields, and increased distance from scattering sources should be employed where practicable in order to keep staff radiation doses as low as reasonably achievable.
format Online
Article
Text
id pubmed-8364281
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-83642812021-08-23 Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers McKenney, Sarah E. Wait, John M. S. Cooper, Virgil N. Johnson, Amirh M. Wang, Jia Leung, Ann N. Clements, Jessica J Appl Clin Med Phys Medical Imaging BACKGROUND: To conserve personal protective equipment (PPE) and reduce exposure to potentially infected COVID‐19 patients, several Californian facilities independently implemented a method of acquiring portable chest radiographs through glass barriers that was originally developed by the University of Washington. METHODS: This work quantifies the transmission of radiation through a glass barrier using six radiographic systems at five facilities. Patient entrance air kerma (EAK) and effective dose were estimated both with and without the glass barrier. Beam penetrability and resulting exposure index (EI) and deviation index (DI) were measured and used to adjust the tube current‐time product (mAs) for glass barriers. Because of beam hardening, the contrast‐to‐noise ratio (CNR) was measured with image quality phantoms to ensure diagnostic integrity. Finally, scatter surveys were performed to assess staff radiation exposure both inside and outside the exam room. RESULTS: The glass barriers attenuated a mean of 61% of the normal X‐ray beams. When the mAs was increased to match EI values, there was no discernible degradation of image quality as determined by the CNR. This was corroborated with subjective assessments of image quality by chest radiologists. The glass‐hardened beams acted as a filter for low energy X‐rays, and some facilities observed slight changes in patient effective doses. There was scattering from both the phantoms and the glass barriers within the room. CONCLUSIONS: Glass barriers require an approximate 2.5 times increase in beam intensity, with all other technique factors held constant. Further refinements are necessary for increased source‐to‐image distance and beam quality in order to adequately match EI values. This does not result in a significant increase in the radiation dose delivered to the patient. The use of lead aprons, mobile shields, and increased distance from scattering sources should be employed where practicable in order to keep staff radiation doses as low as reasonably achievable. John Wiley and Sons Inc. 2021-07-02 /pmc/articles/PMC8364281/ /pubmed/34216091 http://dx.doi.org/10.1002/acm2.13330 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
McKenney, Sarah E.
Wait, John M. S.
Cooper, Virgil N.
Johnson, Amirh M.
Wang, Jia
Leung, Ann N.
Clements, Jessica
Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title_full Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title_fullStr Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title_full_unstemmed Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title_short Multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
title_sort multi‐institution consensus paper for acquisition of portable chest radiographs through glass barriers
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8364281/
https://www.ncbi.nlm.nih.gov/pubmed/34216091
http://dx.doi.org/10.1002/acm2.13330
work_keys_str_mv AT mckenneysarahe multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT waitjohnms multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT coopervirgiln multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT johnsonamirhm multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT wangjia multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT leungannn multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers
AT clementsjessica multiinstitutionconsensuspaperforacquisitionofportablechestradiographsthroughglassbarriers