Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement

BACKGROUND: Since 2017, automated assays have been used in most clinical laboratories for anti-Müllerian hormone (AMH) level measurement. We evaluated the analytical performance of the newly developed automated fluorescent immunoassay system (AFIAS) AMH assay (Boditech Med, Gangwon-do, Korea) in com...

Descripción completa

Detalles Bibliográficos
Autores principales: Han, Aera, Suh, Borum, Yi, Gwang, Lee, Yoo Jin, Kim, Sung Eun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8368231/
https://www.ncbi.nlm.nih.gov/pubmed/34374348
http://dx.doi.org/10.3343/alm.2022.42.1.47
_version_ 1783739165773398016
author Han, Aera
Suh, Borum
Yi, Gwang
Lee, Yoo Jin
Kim, Sung Eun
author_facet Han, Aera
Suh, Borum
Yi, Gwang
Lee, Yoo Jin
Kim, Sung Eun
author_sort Han, Aera
collection PubMed
description BACKGROUND: Since 2017, automated assays have been used in most clinical laboratories for anti-Müllerian hormone (AMH) level measurement. We evaluated the analytical performance of the newly developed automated fluorescent immunoassay system (AFIAS) AMH assay (Boditech Med, Gangwon-do, Korea) in comparison with the Roche Elecsys and Beckman Coulter Access 2 AMH assays. METHODS: Analytical performance of the AFIAS AMH assay was assessed in terms of linearity, repeatability, and within-laboratory precision (CV%) using human recombinant AMH samples according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP05 and EP06. Using 293 serum samples collected from an infertility clinic, the AMH levels were compared across AFIAS, Elecsys, and Access 2 AMH assays according to the CLSI EP09 guidelines. RESULTS: The AFIAS AMH assay results were linear across the measurement range of 0.420–72.386 pmol/L AMH, with repeatability of 6.341%. CV% of the AFIAS AMH assay for three levels of control, 1.786, 7.143, and 56.857 pmol/L, were 5.801%, 5.714%, and 6.228%, respectively. The results of the three AMH assays showed strong correlation AFIAS and Elecsys [slope, 1.055 (95% confidence interval (CI), 1.022–1.088) and Spearman’s rho, 0.978 (95% CI, 0.973–0.983)], Elecsys and Access 2 [slope, 0.813 (95% CI, 0.791–0.834) and Spearman’s rho, 0.986 (95% CI, 0.983–0.989)], and AFIAS and Access 2 [slope, 0.836 (95% CI, 0.821–0.853) and Spearman’s rho, 0.984 (95% CI, 0.980–0.988)]. CONCLUSIONS: The AFIAS AMH assay may be an alternative to the Roche Elecsys and Beckman Coulter Access 2 AMH assays.
format Online
Article
Text
id pubmed-8368231
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-83682312022-01-01 Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement Han, Aera Suh, Borum Yi, Gwang Lee, Yoo Jin Kim, Sung Eun Ann Lab Med Original Article BACKGROUND: Since 2017, automated assays have been used in most clinical laboratories for anti-Müllerian hormone (AMH) level measurement. We evaluated the analytical performance of the newly developed automated fluorescent immunoassay system (AFIAS) AMH assay (Boditech Med, Gangwon-do, Korea) in comparison with the Roche Elecsys and Beckman Coulter Access 2 AMH assays. METHODS: Analytical performance of the AFIAS AMH assay was assessed in terms of linearity, repeatability, and within-laboratory precision (CV%) using human recombinant AMH samples according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP05 and EP06. Using 293 serum samples collected from an infertility clinic, the AMH levels were compared across AFIAS, Elecsys, and Access 2 AMH assays according to the CLSI EP09 guidelines. RESULTS: The AFIAS AMH assay results were linear across the measurement range of 0.420–72.386 pmol/L AMH, with repeatability of 6.341%. CV% of the AFIAS AMH assay for three levels of control, 1.786, 7.143, and 56.857 pmol/L, were 5.801%, 5.714%, and 6.228%, respectively. The results of the three AMH assays showed strong correlation AFIAS and Elecsys [slope, 1.055 (95% confidence interval (CI), 1.022–1.088) and Spearman’s rho, 0.978 (95% CI, 0.973–0.983)], Elecsys and Access 2 [slope, 0.813 (95% CI, 0.791–0.834) and Spearman’s rho, 0.986 (95% CI, 0.983–0.989)], and AFIAS and Access 2 [slope, 0.836 (95% CI, 0.821–0.853) and Spearman’s rho, 0.984 (95% CI, 0.980–0.988)]. CONCLUSIONS: The AFIAS AMH assay may be an alternative to the Roche Elecsys and Beckman Coulter Access 2 AMH assays. Korean Society for Laboratory Medicine 2022-01-01 2022-01-01 /pmc/articles/PMC8368231/ /pubmed/34374348 http://dx.doi.org/10.3343/alm.2022.42.1.47 Text en © Korean Society for Laboratory Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Han, Aera
Suh, Borum
Yi, Gwang
Lee, Yoo Jin
Kim, Sung Eun
Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title_full Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title_fullStr Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title_full_unstemmed Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title_short Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement
title_sort comparison of the automated fluorescent immunoassay system with roche elecsys and beckman coulter access 2 assays for anti-müllerian hormone measurement
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8368231/
https://www.ncbi.nlm.nih.gov/pubmed/34374348
http://dx.doi.org/10.3343/alm.2022.42.1.47
work_keys_str_mv AT hanaera comparisonoftheautomatedfluorescentimmunoassaysystemwithrocheelecsysandbeckmancoulteraccess2assaysforantimullerianhormonemeasurement
AT suhborum comparisonoftheautomatedfluorescentimmunoassaysystemwithrocheelecsysandbeckmancoulteraccess2assaysforantimullerianhormonemeasurement
AT yigwang comparisonoftheautomatedfluorescentimmunoassaysystemwithrocheelecsysandbeckmancoulteraccess2assaysforantimullerianhormonemeasurement
AT leeyoojin comparisonoftheautomatedfluorescentimmunoassaysystemwithrocheelecsysandbeckmancoulteraccess2assaysforantimullerianhormonemeasurement
AT kimsungeun comparisonoftheautomatedfluorescentimmunoassaysystemwithrocheelecsysandbeckmancoulteraccess2assaysforantimullerianhormonemeasurement