Cargando…

A Simple Ventilator Designed To Be Used in Shortage Crises: Construction and Verification Testing

BACKGROUND: The COVID-19 pandemic has demonstrated the possibility of severe ventilator shortages in the near future. OBJECTIVE: We aimed to develop an acute shortage ventilator. METHODS: The ventilator was designed to mechanically compress a self-inflating bag resuscitator, using a modified ventila...

Descripción completa

Detalles Bibliográficos
Autores principales: Akerib, Daniel S, Ames, Andrew, Breidenbach, Martin, Bressack, Michael, Breur, Pieter A, Charles, Eric, Gaba, David M, Herbst, Ryan, Ignarra, Christina M, Luitz, Steffen, Miller, Eric H, Mong, Brian, Shutt, Tom A, Wittgen, Matthias
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8371616/
https://www.ncbi.nlm.nih.gov/pubmed/34458681
http://dx.doi.org/10.2196/26047
Descripción
Sumario:BACKGROUND: The COVID-19 pandemic has demonstrated the possibility of severe ventilator shortages in the near future. OBJECTIVE: We aimed to develop an acute shortage ventilator. METHODS: The ventilator was designed to mechanically compress a self-inflating bag resuscitator, using a modified ventilator patient circuit, which is controlled by a microcontroller and an optional laptop. It was designed to operate in both volume-controlled mode and pressure-controlled assist modes. We tested the ventilator in 4 modes using an artificial lung while measuring the volume, flow, and pressure delivered over time by the ventilator. RESULTS: The ventilator was successful in reaching the desired tidal volume and respiratory rates specified in national emergency use resuscitator system guidelines. The ventilator responded to simulated spontaneous breathing. CONCLUSIONS: The key design goals were achieved. We developed a simple device with high performance for short-term use, made primarily from common hospital parts and generally available nonmedical components to avoid any compatibility or safety issues with the patient, and at low cost, with a unit cost per ventilator is less than $400 US excluding the patient circuit parts, that can be easily manufactured.