Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001

BACKGROUND: BioSystems has developed a histamine kit for automated procedure, through the use of a Y15 analyzer, for quantification of histamine in fish and fishery products. OBJECTIVE: Validate the method under the specific guidelines of the AOAC Research Institute Performance Tested Method(SM) (PT...

Descripción completa

Detalles Bibliográficos
Autores principales: Tobeña, Andreu, Dueñas, Sabina, Boix, Mercè
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8372101/
https://www.ncbi.nlm.nih.gov/pubmed/33367653
http://dx.doi.org/10.1093/jaoacint/qsaa139
_version_ 1783739762221252608
author Tobeña, Andreu
Dueñas, Sabina
Boix, Mercè
author_facet Tobeña, Andreu
Dueñas, Sabina
Boix, Mercè
author_sort Tobeña, Andreu
collection PubMed
description BACKGROUND: BioSystems has developed a histamine kit for automated procedure, through the use of a Y15 analyzer, for quantification of histamine in fish and fishery products. OBJECTIVE: Validate the method under the specific guidelines of the AOAC Research Institute Performance Tested Method(SM) (PTM) program. METHOD: Samples are extracted with boiling water. The enzymatic method is based on histamine dehydrogenase that catalyzes the oxidation of histamine in the presence of an electron mediator that reduces a dye that is measured at 420 nm. The increase of absorbance is proportional to the histamine concentration. Dispensing of reagents and sample, absorbance readings, calibration, and calculation of results are performed automatically in the analyzer BioSystems Y15. RESULTS: The linearity ranges from 0 to 200 mg/kg (r(2) > 0.99). The LOQ is 10 mg/kg in all the matrixes. Recoveries range from 75 to 107% at concentrations from 5 to 200 mg/kg, with repeatability precision values between 0.8 and 5.5%. Comparisons with the HPLC reference method shows a good correlation. Cross-reactivity of the assay is negligible for all biogenic amines tested except for agmatine (6.3%). Product consistency was verified by validating lot-to-lot variations and variations within the same lot. Shelf life was verified by real-time stability testing during 40 months at 2–8°C. No differences in histamine detection were observed in robustness testing, in which minor changes are introduced to the assay protocol. CONCLUSIONS: The automated, simple, and rapid BioSystems Y15 Histamine Dehydrogenase Kit has been successfully validated. HIGHLIGHTS: The method is qualified for PTM certification No. 072001.
format Online
Article
Text
id pubmed-8372101
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-83721012021-12-28 Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001 Tobeña, Andreu Dueñas, Sabina Boix, Mercè J AOAC Int Food Chemical Contaminants BACKGROUND: BioSystems has developed a histamine kit for automated procedure, through the use of a Y15 analyzer, for quantification of histamine in fish and fishery products. OBJECTIVE: Validate the method under the specific guidelines of the AOAC Research Institute Performance Tested Method(SM) (PTM) program. METHOD: Samples are extracted with boiling water. The enzymatic method is based on histamine dehydrogenase that catalyzes the oxidation of histamine in the presence of an electron mediator that reduces a dye that is measured at 420 nm. The increase of absorbance is proportional to the histamine concentration. Dispensing of reagents and sample, absorbance readings, calibration, and calculation of results are performed automatically in the analyzer BioSystems Y15. RESULTS: The linearity ranges from 0 to 200 mg/kg (r(2) > 0.99). The LOQ is 10 mg/kg in all the matrixes. Recoveries range from 75 to 107% at concentrations from 5 to 200 mg/kg, with repeatability precision values between 0.8 and 5.5%. Comparisons with the HPLC reference method shows a good correlation. Cross-reactivity of the assay is negligible for all biogenic amines tested except for agmatine (6.3%). Product consistency was verified by validating lot-to-lot variations and variations within the same lot. Shelf life was verified by real-time stability testing during 40 months at 2–8°C. No differences in histamine detection were observed in robustness testing, in which minor changes are introduced to the assay protocol. CONCLUSIONS: The automated, simple, and rapid BioSystems Y15 Histamine Dehydrogenase Kit has been successfully validated. HIGHLIGHTS: The method is qualified for PTM certification No. 072001. Oxford University Press 2020-10-14 /pmc/articles/PMC8372101/ /pubmed/33367653 http://dx.doi.org/10.1093/jaoacint/qsaa139 Text en © AOAC INTERNATIONAL 2020. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Food Chemical Contaminants
Tobeña, Andreu
Dueñas, Sabina
Boix, Mercè
Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title_full Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title_fullStr Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title_full_unstemmed Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title_short Validation Study of BioSystems(®) Y15 Histamine Dehydrogenase Kit for the Detection of Histamine in Fish and Fishery Products: AOAC Performance Tested Method(SM) 072001
title_sort validation study of biosystems(®) y15 histamine dehydrogenase kit for the detection of histamine in fish and fishery products: aoac performance tested method(sm) 072001
topic Food Chemical Contaminants
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8372101/
https://www.ncbi.nlm.nih.gov/pubmed/33367653
http://dx.doi.org/10.1093/jaoacint/qsaa139
work_keys_str_mv AT tobenaandreu validationstudyofbiosystemsy15histaminedehydrogenasekitforthedetectionofhistamineinfishandfisheryproductsaoacperformancetestedmethodsm072001
AT duenassabina validationstudyofbiosystemsy15histaminedehydrogenasekitforthedetectionofhistamineinfishandfisheryproductsaoacperformancetestedmethodsm072001
AT boixmerce validationstudyofbiosystemsy15histaminedehydrogenasekitforthedetectionofhistamineinfishandfisheryproductsaoacperformancetestedmethodsm072001