Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults

OBJECTIVE: To evaluate whether measurement of the hemoglobin (Hb) concentration with a blood gas analyzer approximates that determined by a conventional coulter counter in critically ill adults. DESIGN: Prospective patient series. SETTING: A 32-bed cardiovascular, neurosurgical, trauma and medical-s...

Descripción completa

Detalles Bibliográficos
Autores principales: Ray, Joel G, Post, Julia R, Hamielec, Cindy
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2002
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC83849/
https://www.ncbi.nlm.nih.gov/pubmed/11940269
_version_ 1782120184483414016
author Ray, Joel G
Post, Julia R
Hamielec, Cindy
author_facet Ray, Joel G
Post, Julia R
Hamielec, Cindy
author_sort Ray, Joel G
collection PubMed
description OBJECTIVE: To evaluate whether measurement of the hemoglobin (Hb) concentration with a blood gas analyzer approximates that determined by a conventional coulter counter in critically ill adults. DESIGN: Prospective patient series. SETTING: A 32-bed cardiovascular, neurosurgical, trauma and medical-surgical intensive care unit in a single Canadian center. PATIENTS: We consecutively recruited 202 critically ill adults, the majority of whom had recent cardiac or vascular surgery, neurosurgery or trauma. MEASUREMENTS: The nurse obtained a single arterial blood sample within a few hours of the patient's admission to the intensive care unit. The Hb concentration was determined from each blood sample in a masked fashion, using both a blood gas analyzer and a conventional laboratory coulter counter. MAIN RESULTS: A total of 202 consecutive paired analyses were conducted. There was a highly significant correlation between the coulter counter and blood gas analyzer methods of Hb measurement (r(2) = 0.98, 95% confidence interval [CI] = 0.97–0.99; P < 0.0001). Using the method of Bland and Altman, the overall mean difference in Hb concentration between the coulter counter and the blood gas analyzer was -4.3 g/l (95% CI = -11.0 to 2.4). Of the 11 (5.4%) Hb measurements that extended beyond the upper and lower 95% CI, 10 (5.0%) were within ± 3 g/l of these confidence limits. CONCLUSIONS: An arterial blood gas analyzer may provide a valid alternative method to the traditional coulter counter for the rapid assessment of Hb concentration among critically ill adults. Since issues related to its safety, quality control, data entry and cost savings have yet to be addressed, however, use of such point of care testing should be viewed as a supplement to conventional laboratory testing.
format Text
id pubmed-83849
institution National Center for Biotechnology Information
language English
publishDate 2002
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-838492002-03-15 Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults Ray, Joel G Post, Julia R Hamielec, Cindy Crit Care Research OBJECTIVE: To evaluate whether measurement of the hemoglobin (Hb) concentration with a blood gas analyzer approximates that determined by a conventional coulter counter in critically ill adults. DESIGN: Prospective patient series. SETTING: A 32-bed cardiovascular, neurosurgical, trauma and medical-surgical intensive care unit in a single Canadian center. PATIENTS: We consecutively recruited 202 critically ill adults, the majority of whom had recent cardiac or vascular surgery, neurosurgery or trauma. MEASUREMENTS: The nurse obtained a single arterial blood sample within a few hours of the patient's admission to the intensive care unit. The Hb concentration was determined from each blood sample in a masked fashion, using both a blood gas analyzer and a conventional laboratory coulter counter. MAIN RESULTS: A total of 202 consecutive paired analyses were conducted. There was a highly significant correlation between the coulter counter and blood gas analyzer methods of Hb measurement (r(2) = 0.98, 95% confidence interval [CI] = 0.97–0.99; P < 0.0001). Using the method of Bland and Altman, the overall mean difference in Hb concentration between the coulter counter and the blood gas analyzer was -4.3 g/l (95% CI = -11.0 to 2.4). Of the 11 (5.4%) Hb measurements that extended beyond the upper and lower 95% CI, 10 (5.0%) were within ± 3 g/l of these confidence limits. CONCLUSIONS: An arterial blood gas analyzer may provide a valid alternative method to the traditional coulter counter for the rapid assessment of Hb concentration among critically ill adults. Since issues related to its safety, quality control, data entry and cost savings have yet to be addressed, however, use of such point of care testing should be viewed as a supplement to conventional laboratory testing. BioMed Central 2002 2001-11-19 /pmc/articles/PMC83849/ /pubmed/11940269 Text en Copyright © 2002 Ray et al., licensee BioMed Central Ltd
spellingShingle Research
Ray, Joel G
Post, Julia R
Hamielec, Cindy
Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title_full Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title_fullStr Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title_full_unstemmed Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title_short Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
title_sort use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC83849/
https://www.ncbi.nlm.nih.gov/pubmed/11940269
work_keys_str_mv AT rayjoelg useofarapidarterialbloodgasanalyzertoestimatebloodhemoglobinconcentrationamongcriticallyilladults
AT postjuliar useofarapidarterialbloodgasanalyzertoestimatebloodhemoglobinconcentrationamongcriticallyilladults
AT hamieleccindy useofarapidarterialbloodgasanalyzertoestimatebloodhemoglobinconcentrationamongcriticallyilladults