Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique
INTRODUCTION AND METHODS: Dynamic fluorescence quenching is a technique that may overcome some of the limitations associated with measurement of tissue partial oxygen tension (PO(2)). We compared this technique with a polarographic Eppendorf needle electrode method using a saline tonometer in which...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2002
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC83850/ https://www.ncbi.nlm.nih.gov/pubmed/11940270 |
_version_ | 1782120184704663552 |
---|---|
author | Shaw, Andrew D Li, Zheng Thomas, Zach Stevens, Craig W |
author_facet | Shaw, Andrew D Li, Zheng Thomas, Zach Stevens, Craig W |
author_sort | Shaw, Andrew D |
collection | PubMed |
description | INTRODUCTION AND METHODS: Dynamic fluorescence quenching is a technique that may overcome some of the limitations associated with measurement of tissue partial oxygen tension (PO(2)). We compared this technique with a polarographic Eppendorf needle electrode method using a saline tonometer in which the PO(2) could be controlled. We also tested the fluorescence quenching system in a rodent model of skeletal muscle ischemiahypoxia. RESULTS: Both systems measured PO(2) accurately in the tonometer, and there was excellent correlation between them (r(2) = 0.99). The polarographic system exhibited proportional bias that was not evident with the fluorescence method. In vivo, the fluorescence quenching technique provided a readily recordable signal that varied as expected. DISCUSSION: Measurement of tissue PO(2) using fluorescence quenching is at least as accurate as measurement using the Eppendorf needle electrode in vitro, and may prove useful in vivo for assessment of tissue oxygenation. |
format | Text |
id | pubmed-83850 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2002 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-838502002-03-15 Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique Shaw, Andrew D Li, Zheng Thomas, Zach Stevens, Craig W Crit Care Research INTRODUCTION AND METHODS: Dynamic fluorescence quenching is a technique that may overcome some of the limitations associated with measurement of tissue partial oxygen tension (PO(2)). We compared this technique with a polarographic Eppendorf needle electrode method using a saline tonometer in which the PO(2) could be controlled. We also tested the fluorescence quenching system in a rodent model of skeletal muscle ischemiahypoxia. RESULTS: Both systems measured PO(2) accurately in the tonometer, and there was excellent correlation between them (r(2) = 0.99). The polarographic system exhibited proportional bias that was not evident with the fluorescence method. In vivo, the fluorescence quenching technique provided a readily recordable signal that varied as expected. DISCUSSION: Measurement of tissue PO(2) using fluorescence quenching is at least as accurate as measurement using the Eppendorf needle electrode in vitro, and may prove useful in vivo for assessment of tissue oxygenation. BioMed Central 2002 2002-01-10 /pmc/articles/PMC83850/ /pubmed/11940270 Text en Copyright © 2002 BioMed Central Ltd |
spellingShingle | Research Shaw, Andrew D Li, Zheng Thomas, Zach Stevens, Craig W Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title | Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title_full | Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title_fullStr | Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title_full_unstemmed | Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title_short | Assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
title_sort | assessment of tissue oxygen tension: comparison of dynamic fluorescence quenching and polarographic electrode technique |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC83850/ https://www.ncbi.nlm.nih.gov/pubmed/11940270 |
work_keys_str_mv | AT shawandrewd assessmentoftissueoxygentensioncomparisonofdynamicfluorescencequenchingandpolarographicelectrodetechnique AT lizheng assessmentoftissueoxygentensioncomparisonofdynamicfluorescencequenchingandpolarographicelectrodetechnique AT thomaszach assessmentoftissueoxygentensioncomparisonofdynamicfluorescencequenchingandpolarographicelectrodetechnique AT stevenscraigw assessmentoftissueoxygentensioncomparisonofdynamicfluorescencequenchingandpolarographicelectrodetechnique |