Cargando…
Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results
Extracellular vesicles (EVs) from liquid biopsies are extensively analyzed by flow cytometry, a technology that is continuously evolving. Thresholding utilizing a violet 405 nm laser side scatter (VSSC) has recently been implemented. Here, we collected set of large EV (lEV) samples from cord blood,...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8392526/ https://www.ncbi.nlm.nih.gov/pubmed/34441255 http://dx.doi.org/10.3390/diagnostics11081320 |
_version_ | 1783743524085170176 |
---|---|
author | Pekárková, Kristýna Soukup, Jakub Kostelanská, Marie Širc, Jan Straňák, Zbyněk Holada, Karel |
author_facet | Pekárková, Kristýna Soukup, Jakub Kostelanská, Marie Širc, Jan Straňák, Zbyněk Holada, Karel |
author_sort | Pekárková, Kristýna |
collection | PubMed |
description | Extracellular vesicles (EVs) from liquid biopsies are extensively analyzed by flow cytometry, a technology that is continuously evolving. Thresholding utilizing a violet 405 nm laser side scatter (VSSC) has recently been implemented. Here, we collected set of large EV (lEV) samples from cord blood, which we analyzed using a standard flow cytometer improved via a 405 nm laser side scatter. Samples were analyzed using two distinct thresholding methods—one based on VSSC, and one based on VSSC combined with fluorescence thresholding on stained phosphatidylserine. Through these thresholding methods, we compared lEVs from pre-term births and control cord blood. Double-labeled lEVs with platelet CD36+/CD41+, activated platelet CD41+/CD62P+ and endothelial CD31+/CD105+ antibodies were used. Apart from comparing the two groups together, we also correlated measured lEVs with the thresholding methods. We also correlated the results of this study with data analyzed in our previous study in which we used a conventional 488 nm laser SSC. We did not find any difference between the two cord blood groups. However, we found highly concurrent data via our correlation of the thresholding methods, with correlation coefficients ranging from 0.80 to 0.96 even though the numbers of detected lEVs differed between thresholding methods. In conclusion, our approaches to thresholding provided concurrent data and it seems that improving the cytometer with the use of a VSSC increases its sensitivity, despite not being particularly critical to the validity of flow cytometric studies that compare pathological and physiological conditions in liquid biopsies. |
format | Online Article Text |
id | pubmed-8392526 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-83925262021-08-28 Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results Pekárková, Kristýna Soukup, Jakub Kostelanská, Marie Širc, Jan Straňák, Zbyněk Holada, Karel Diagnostics (Basel) Technical Note Extracellular vesicles (EVs) from liquid biopsies are extensively analyzed by flow cytometry, a technology that is continuously evolving. Thresholding utilizing a violet 405 nm laser side scatter (VSSC) has recently been implemented. Here, we collected set of large EV (lEV) samples from cord blood, which we analyzed using a standard flow cytometer improved via a 405 nm laser side scatter. Samples were analyzed using two distinct thresholding methods—one based on VSSC, and one based on VSSC combined with fluorescence thresholding on stained phosphatidylserine. Through these thresholding methods, we compared lEVs from pre-term births and control cord blood. Double-labeled lEVs with platelet CD36+/CD41+, activated platelet CD41+/CD62P+ and endothelial CD31+/CD105+ antibodies were used. Apart from comparing the two groups together, we also correlated measured lEVs with the thresholding methods. We also correlated the results of this study with data analyzed in our previous study in which we used a conventional 488 nm laser SSC. We did not find any difference between the two cord blood groups. However, we found highly concurrent data via our correlation of the thresholding methods, with correlation coefficients ranging from 0.80 to 0.96 even though the numbers of detected lEVs differed between thresholding methods. In conclusion, our approaches to thresholding provided concurrent data and it seems that improving the cytometer with the use of a VSSC increases its sensitivity, despite not being particularly critical to the validity of flow cytometric studies that compare pathological and physiological conditions in liquid biopsies. MDPI 2021-07-22 /pmc/articles/PMC8392526/ /pubmed/34441255 http://dx.doi.org/10.3390/diagnostics11081320 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Technical Note Pekárková, Kristýna Soukup, Jakub Kostelanská, Marie Širc, Jan Straňák, Zbyněk Holada, Karel Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title | Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title_full | Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title_fullStr | Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title_full_unstemmed | Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title_short | Cord Blood Extracellular Vesicles Analyzed by Flow Cytometry with Thresholding Using 405 nm or 488 nm Laser Leads to Concurrent Results |
title_sort | cord blood extracellular vesicles analyzed by flow cytometry with thresholding using 405 nm or 488 nm laser leads to concurrent results |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8392526/ https://www.ncbi.nlm.nih.gov/pubmed/34441255 http://dx.doi.org/10.3390/diagnostics11081320 |
work_keys_str_mv | AT pekarkovakristyna cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults AT soukupjakub cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults AT kostelanskamarie cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults AT sircjan cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults AT stranakzbynek cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults AT holadakarel cordbloodextracellularvesiclesanalyzedbyflowcytometrywiththresholdingusing405nmor488nmlaserleadstoconcurrentresults |