Cargando…

Mechanical Properties of Porous Structures for Dental Implants: Experimental Study and Computational Homogenization

A combined experimental and numerical study on titanium porous microstructures intended to interface the bone tissue and the solid homogeneous part of a modern dental implant is presented. A specific class of trabecular geometries is compared to a gyroid structure. Limitations associated with the ap...

Descripción completa

Detalles Bibliográficos
Autores principales: Jíra, Aleš, Šejnoha, Michal, Krejčí, Tomáš, Vorel, Jan, Řehounek, Luboš, Marseglia, Guido
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8398312/
https://www.ncbi.nlm.nih.gov/pubmed/34443120
http://dx.doi.org/10.3390/ma14164592
Descripción
Sumario:A combined experimental and numerical study on titanium porous microstructures intended to interface the bone tissue and the solid homogeneous part of a modern dental implant is presented. A specific class of trabecular geometries is compared to a gyroid structure. Limitations associated with the application of the adopted selective laser melting technology to small microstructures with a pore size of 500 [Formula: see text] m are first examined experimentally. The measured effective elastic properties of trabecular structures made of Ti6Al4V material support the computational framework based on homogenization with the difference between the measured and predicted Young’s moduli of the Dode Thick structure being less than 5%. In this regard, the extended finite element method is promoted, particularly in light of the complex sheet gyroid studied next. While for plastic material-based structures a close match between experiments and simulations was observed, an order of magnitude difference was encountered for titanium specimens. This calls for further study and we expect to reconcile this inconsistency with the help of computational microtomography.