Analytical Evaluation of the Ideal Strategy for High-Throughput Flow Injection Analysis by Tandem Mass Spectrometry in Routine Newborn Screening

The introduction of tandem mass spectrometry (MS/MS) to clinical laboratories and the advent of expanded newborn screening (NBS) were crucial changes to public health programs worldwide. Speed, robustness, accuracy, selectivity, and specificity of analysis are all requirements of expanded NBS and ar...

Descripción completa

Detalles Bibliográficos
Autores principales: Cicalini, Ilaria, Valentinuzzi, Silvia, Pieragostino, Damiana, Consalvo, Ada, Zucchelli, Mirco, Donzelli, Simone, Ambrogi, Davide, Brown, Heather A., Calton, Lisa J., Stuppia, Liborio, De Laurenzi, Vincenzo, Del Boccio, Piero, Rossi, Claudia
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8399422/
https://www.ncbi.nlm.nih.gov/pubmed/34436414
http://dx.doi.org/10.3390/metabo11080473
Descripción
Sumario:The introduction of tandem mass spectrometry (MS/MS) to clinical laboratories and the advent of expanded newborn screening (NBS) were crucial changes to public health programs worldwide. Speed, robustness, accuracy, selectivity, and specificity of analysis are all requirements of expanded NBS and are needed to minimize false positive results risks, to possibly eliminate false negatives, and to improve the positive predictive value of NBS. In this study, we firstly evaluated the analytical performances of the RenataDX Screening System, a fully integrated flow-injection MS/MS (FIA-MS/MS) IVD system for high-throughput dried blood spot (DBS) analysis in a routine NBS laboratory. Since a choice of several commercial NBS kits is available, we sought to compare NeoBase(TM) 2 (PerkinElmer(®)) and MassChrom(®) (Chromsystems) non-derivatized kits on the RenataDX platform by evaluating their analytical performances. Moreover, we verified the degree of correlation between data obtained by the two different NBS MS/MS kits by FIA-MS/MS of over 500 samples. Our data suggest that both methods correlate well with clinically insignificant differences that do not impact the NBS result. Finally, while NeoBase™ 2 offers an easier and faster sample preparation, MassChrom(®) provides a cleaner sample extract which empirically should improve instrument reliability.