High-Resolution Multi-Channel Frequency Standard Comparator Using Digital Frequency Measurement

The rapid improvement accuracy of the atomic frequency standard puts forward higher requirements for the measurement resolution of the atomic frequency standard comparison system. To overcome the defect that the single zero-crossing point detection is sensitive to noise in the traditional dual mixer...

Descripción completa

Detalles Bibliográficos
Autores principales: Xiao, Bo, Liu, Ya, Li, Xiaohui, Deng, Zhifeng, Xue, Yanrong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8402556/
https://www.ncbi.nlm.nih.gov/pubmed/34451064
http://dx.doi.org/10.3390/s21165626
Descripción
Sumario:The rapid improvement accuracy of the atomic frequency standard puts forward higher requirements for the measurement resolution of the atomic frequency standard comparison system. To overcome the defect that the single zero-crossing point detection is sensitive to noise in the traditional dual mixer time difference measurement method, a digital frequency measurement method is proposed. This method combines sinusoidal beat technology, multi-channel synchronous acquisition technology, and digital frequency measurement technology, and uses differential compensation of system error to realize the precision measurement of atomic frequency standard. The frequency measurement accuracy is less than 2.5 × 10(−14) and the noise floor is better than 6.5 × 10(−15)/τ = 1 s. The system has a high frequency measurement accuracy and a low noise floor, which can realize the precise measurement of a highly stable frequency source.