Cargando…

Sensitivity and Specificity of New Visual Field Screening Software for Diagnosing Hemianopia

PURPOSE: To assess the diagnostic accuracy of visual field results generated by the newly developed software (CU-VF) and the standard automated perimetry (SAP) for detecting hemianopia. PATIENTS AND METHODS: Forty-three subjects with hemianopia and 33 controls were tested with the CU-VF software on...

Descripción completa

Detalles Bibliográficos
Autores principales: Jariyakosol, Supharat, Jaru-ampornpan, Patcharaporn, Manassakorn, Anita, Itthipanichpong, Rath, Hirunwiwatkul, Parima, Tantisevi, Visanee, Somkijrungroj, Thanapong, Rojanapongpun, Prin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8412821/
https://www.ncbi.nlm.nih.gov/pubmed/34512063
http://dx.doi.org/10.2147/EB.S315403
Descripción
Sumario:PURPOSE: To assess the diagnostic accuracy of visual field results generated by the newly developed software (CU-VF) and the standard automated perimetry (SAP) for detecting hemianopia. PATIENTS AND METHODS: Forty-three subjects with hemianopia and 33 controls were tested with the CU-VF software on a personal computer and SAP. Hemianopia was defined as the presence of a hemianopic field respecting the vertical meridian on SAP with the corresponding neuroimaging pathology as evaluated by 2 neuro-ophthalmologists. Results of CU-VF were independently evaluated by 2 neuro-ophthalmologists, 1 general ophthalmologist, and 1 general practitioner in terms of the presence of hemianopia. Sensitivity, specificity, and kappa coefficient for inter-observer reliability were calculated. Satisfaction and ease of use were evaluated with a visual analog-scale questionnaire and analyzed using paired t-test. RESULTS: The sensitivity (95% CI) and specificity (95% CI) of the CU-VF to detect hemianopia was 74.42% (58.53–85.96) and 93.94% (78.38–99.94). Kappa coefficient between neuro-ophthalmologists versus general ophthalmologist and general practitioner were 0.71 and 0.84, respectively. The mean (SD) test duration was 2.25 (0.002) minutes for the CU-VF and 5.38 (1.34) minutes for SAP (p < 0.001). Subjects reported significantly higher satisfaction and comfort using the CU-VF software compared to SAP. CONCLUSION: The CU-VF screening software showed good validity and reliability to detect hemianopia, with shorter test duration and higher subject satisfaction compared to SAP.