Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound

PURPOSE: The purpose of this study is to evaluate biomechanical accuracy of 3D printed anatomical vessels using a material jetting printer (J750, Stratasys, Rehovot, Israel) by measuring distensibility via intravascular ultrasound. MATERIALS AND METHODS: The test samples are 3D printed tubes to simu...

Descripción completa

Detalles Bibliográficos
Autores principales: Sparks, Adam J., Smith, Cody M., Allman, Ariana B., Senko, Jillian L., Meess, Karen M., Ducharme, Richard W., Springer, Michael E., Waqas, Muhammad, Siddiqui, Adnan H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8414686/
https://www.ncbi.nlm.nih.gov/pubmed/34477997
http://dx.doi.org/10.1186/s41205-021-00114-8
_version_ 1783747829273985024
author Sparks, Adam J.
Smith, Cody M.
Allman, Ariana B.
Senko, Jillian L.
Meess, Karen M.
Ducharme, Richard W.
Springer, Michael E.
Waqas, Muhammad
Siddiqui, Adnan H.
author_facet Sparks, Adam J.
Smith, Cody M.
Allman, Ariana B.
Senko, Jillian L.
Meess, Karen M.
Ducharme, Richard W.
Springer, Michael E.
Waqas, Muhammad
Siddiqui, Adnan H.
author_sort Sparks, Adam J.
collection PubMed
description PURPOSE: The purpose of this study is to evaluate biomechanical accuracy of 3D printed anatomical vessels using a material jetting printer (J750, Stratasys, Rehovot, Israel) by measuring distensibility via intravascular ultrasound. MATERIALS AND METHODS: The test samples are 3D printed tubes to simulate arterial vessels (aorta, carotid artery, and coronary artery). Each vessel type is defined by design geometry of the vessel inner diameter and wall thickness. Vessel inner diameters are aorta = 30mm, carotid = 7mm, and coronary = 3mm. Vessel wall thickness are aorta = 3mm, carotid = 1.5mm, and coronary = 1mm. Each vessel type was printed in 3 different material options. Material options are user-selected from the J750 printer software graphical user interface as blood vessel wall anatomy elements in ‘compliant’, ‘slightly compliant’, and ‘rigid’ options. Three replicates of each vessel type were printed in each of the three selected material options, for a total of 27 models. The vessels were connected to a flow loop system where pressure was monitored via a pressure wire and cross-sectional area was measured with intravascular ultrasound (IVUS). Distensibility was calculated by comparing the % difference in cross-sectional area vs. pulse pressure to clinical literature values. Target clinical ranges for normal and diseased population distensibility are 10.3-44 % for the aorta, 5.1-10.1 % for carotid artery, and 0.5-6 % for coronary artery. RESULTS: Aorta test vessels had the most clinically representative distensibility when printed in user-selected ‘compliant’ and ‘slightly compliant’ material. All aorta test vessels of ‘compliant’ material (n = 3) and 2 of 3 ‘slightly compliant’ vessels evaluated were within target range. Carotid vessels were most clinically represented in distensibility when printed in ‘compliant’ and ‘slightly compliant’ material. For carotid test vessels, 2 of 3 ‘compliant’ material samples and 1 of 3 ‘slightly compliant’ material samples were within target range. Coronary arteries were most clinically represented in distensibility when printed in ‘slightly compliant’ and ‘rigid’ material. For coronary test vessels, 1 of 3 ‘slightly compliant’ materials and 3 of 3 ‘rigid’ material samples fell within target range. CONCLUSIONS: This study suggests that advancements in materials and 3D printing technology introduced with the J750 Digital Anatomy 3D Printer can enable anatomical models with clinically relevant distensibility.
format Online
Article
Text
id pubmed-8414686
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-84146862021-09-09 Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound Sparks, Adam J. Smith, Cody M. Allman, Ariana B. Senko, Jillian L. Meess, Karen M. Ducharme, Richard W. Springer, Michael E. Waqas, Muhammad Siddiqui, Adnan H. 3D Print Med Research PURPOSE: The purpose of this study is to evaluate biomechanical accuracy of 3D printed anatomical vessels using a material jetting printer (J750, Stratasys, Rehovot, Israel) by measuring distensibility via intravascular ultrasound. MATERIALS AND METHODS: The test samples are 3D printed tubes to simulate arterial vessels (aorta, carotid artery, and coronary artery). Each vessel type is defined by design geometry of the vessel inner diameter and wall thickness. Vessel inner diameters are aorta = 30mm, carotid = 7mm, and coronary = 3mm. Vessel wall thickness are aorta = 3mm, carotid = 1.5mm, and coronary = 1mm. Each vessel type was printed in 3 different material options. Material options are user-selected from the J750 printer software graphical user interface as blood vessel wall anatomy elements in ‘compliant’, ‘slightly compliant’, and ‘rigid’ options. Three replicates of each vessel type were printed in each of the three selected material options, for a total of 27 models. The vessels were connected to a flow loop system where pressure was monitored via a pressure wire and cross-sectional area was measured with intravascular ultrasound (IVUS). Distensibility was calculated by comparing the % difference in cross-sectional area vs. pulse pressure to clinical literature values. Target clinical ranges for normal and diseased population distensibility are 10.3-44 % for the aorta, 5.1-10.1 % for carotid artery, and 0.5-6 % for coronary artery. RESULTS: Aorta test vessels had the most clinically representative distensibility when printed in user-selected ‘compliant’ and ‘slightly compliant’ material. All aorta test vessels of ‘compliant’ material (n = 3) and 2 of 3 ‘slightly compliant’ vessels evaluated were within target range. Carotid vessels were most clinically represented in distensibility when printed in ‘compliant’ and ‘slightly compliant’ material. For carotid test vessels, 2 of 3 ‘compliant’ material samples and 1 of 3 ‘slightly compliant’ material samples were within target range. Coronary arteries were most clinically represented in distensibility when printed in ‘slightly compliant’ and ‘rigid’ material. For coronary test vessels, 1 of 3 ‘slightly compliant’ materials and 3 of 3 ‘rigid’ material samples fell within target range. CONCLUSIONS: This study suggests that advancements in materials and 3D printing technology introduced with the J750 Digital Anatomy 3D Printer can enable anatomical models with clinically relevant distensibility. Springer International Publishing 2021-09-03 /pmc/articles/PMC8414686/ /pubmed/34477997 http://dx.doi.org/10.1186/s41205-021-00114-8 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Sparks, Adam J.
Smith, Cody M.
Allman, Ariana B.
Senko, Jillian L.
Meess, Karen M.
Ducharme, Richard W.
Springer, Michael E.
Waqas, Muhammad
Siddiqui, Adnan H.
Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title_full Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title_fullStr Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title_full_unstemmed Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title_short Compliant vascular models 3D printed with the Stratasys J750: a direct characterization of model distensibility using intravascular ultrasound
title_sort compliant vascular models 3d printed with the stratasys j750: a direct characterization of model distensibility using intravascular ultrasound
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8414686/
https://www.ncbi.nlm.nih.gov/pubmed/34477997
http://dx.doi.org/10.1186/s41205-021-00114-8
work_keys_str_mv AT sparksadamj compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT smithcodym compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT allmanarianab compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT senkojillianl compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT meesskarenm compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT ducharmerichardw compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT springermichaele compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT waqasmuhammad compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound
AT siddiquiadnanh compliantvascularmodels3dprintedwiththestratasysj750adirectcharacterizationofmodeldistensibilityusingintravascularultrasound