Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years
Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8415254/ https://www.ncbi.nlm.nih.gov/pubmed/34566293 http://dx.doi.org/10.4103/jmp.JMP_42_21 |
_version_ | 1783747931468201984 |
---|---|
author | Semwal, Manoj Kumar Gupta, Harishchandra Singhal, Sakshi Kumar, Ashok Kaushik, Aruna |
author_facet | Semwal, Manoj Kumar Gupta, Harishchandra Singhal, Sakshi Kumar, Ashok Kaushik, Aruna |
author_sort | Semwal, Manoj Kumar |
collection | PubMed |
description | Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration factor (N(DW)) over a period of time. In this study, the N(DW)factor for two ion chambers, thimble type (Farmer) and parallel plate type (Markus), used in most of the radiotherapy centers, were retrospectively reviewed for 20 years. The calibration on all occasions except once was carried out at the Secondary Standards Dosimetry Laboratory, Bhabha Atomic Research Centre, Mumbai. The change in the N(DW)factor over this period was < ±3%. We, therefore, believe that a dosimeter with no history of repairs showed reasonable stability in the N(DW)factor over a long period. |
format | Online Article Text |
id | pubmed-8415254 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-84152542021-09-24 Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years Semwal, Manoj Kumar Gupta, Harishchandra Singhal, Sakshi Kumar, Ashok Kaushik, Aruna J Med Phys Technical Note Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration factor (N(DW)) over a period of time. In this study, the N(DW)factor for two ion chambers, thimble type (Farmer) and parallel plate type (Markus), used in most of the radiotherapy centers, were retrospectively reviewed for 20 years. The calibration on all occasions except once was carried out at the Secondary Standards Dosimetry Laboratory, Bhabha Atomic Research Centre, Mumbai. The change in the N(DW)factor over this period was < ±3%. We, therefore, believe that a dosimeter with no history of repairs showed reasonable stability in the N(DW)factor over a long period. Wolters Kluwer - Medknow 2021 2021-08-07 /pmc/articles/PMC8415254/ /pubmed/34566293 http://dx.doi.org/10.4103/jmp.JMP_42_21 Text en Copyright: © 2021 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Technical Note Semwal, Manoj Kumar Gupta, Harishchandra Singhal, Sakshi Kumar, Ashok Kaushik, Aruna Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title | Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title_full | Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title_fullStr | Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title_full_unstemmed | Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title_short | Variation in Dosimeter Calibration Factor (N(DW)) Over a Period of 20 Years |
title_sort | variation in dosimeter calibration factor (n(dw)) over a period of 20 years |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8415254/ https://www.ncbi.nlm.nih.gov/pubmed/34566293 http://dx.doi.org/10.4103/jmp.JMP_42_21 |
work_keys_str_mv | AT semwalmanojkumar variationindosimetercalibrationfactorndwoveraperiodof20years AT guptaharishchandra variationindosimetercalibrationfactorndwoveraperiodof20years AT singhalsakshi variationindosimetercalibrationfactorndwoveraperiodof20years AT kumarashok variationindosimetercalibrationfactorndwoveraperiodof20years AT kaushikaruna variationindosimetercalibrationfactorndwoveraperiodof20years |