Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy

Patient breathing during lung cancer radiotherapy reduces the ability to keep a sharp dose gradient between tumor and normal tissues. To mitigate detrimental effects, accurate information about the tumor position is required. In this work, we evaluate the errors in modeled tumor positions over sever...

Descripción completa

Detalles Bibliográficos
Autores principales: Wikström, Kenneth A., Isacsson, Ulf M., Nilsson, Kristina M., Ahnesjö, Anders
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8425865/
https://www.ncbi.nlm.nih.gov/pubmed/34258853
http://dx.doi.org/10.1002/acm2.13351
_version_ 1783749927824785408
author Wikström, Kenneth A.
Isacsson, Ulf M.
Nilsson, Kristina M.
Ahnesjö, Anders
author_facet Wikström, Kenneth A.
Isacsson, Ulf M.
Nilsson, Kristina M.
Ahnesjö, Anders
author_sort Wikström, Kenneth A.
collection PubMed
description Patient breathing during lung cancer radiotherapy reduces the ability to keep a sharp dose gradient between tumor and normal tissues. To mitigate detrimental effects, accurate information about the tumor position is required. In this work, we evaluate the errors in modeled tumor positions over several fractions of a simple tumor motion model driven by a surface surrogate measure and its time derivative. The model is tested with respect to four different surface surrogates and a varying number of surrogate and image acquisitions used for model training. Fourteen patients were imaged 100 times with cine CT, at three sessions mimicking a planning session followed by two treatment fractions. Patient body contours were concurrently detected by a body surface laser scanning system BSLS from which four surface surrogates were extracted; thoracic point TP, abdominal point AP, the radial distance mean RDM, and a surface derived volume SDV. The motion model was trained on session 1 and evaluated on sessions 2 and 3 by comparing modeled tumor positions with measured positions from the cine images. The number of concurrent surrogate and image acquisitions used in the training set was varied, and its impact on the final result was evaluated. The use of AP as a surface surrogate yielded the smallest error in modeled tumor positions. The mean deviation between modeled and measured tumor positions was 1.9 mm. The corresponding deviations for using the other surrogates were 2.0 mm (RDM), 2.8 mm (SDV), and 3.0 mm (TP). To produce a motion model that accurately models the tumor position over several fractions requires at least 10 simultaneous surrogate and image acquisitions over 1–2 minutes.
format Online
Article
Text
id pubmed-8425865
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-84258652021-09-13 Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy Wikström, Kenneth A. Isacsson, Ulf M. Nilsson, Kristina M. Ahnesjö, Anders J Appl Clin Med Phys Radiation Oncology Physics Patient breathing during lung cancer radiotherapy reduces the ability to keep a sharp dose gradient between tumor and normal tissues. To mitigate detrimental effects, accurate information about the tumor position is required. In this work, we evaluate the errors in modeled tumor positions over several fractions of a simple tumor motion model driven by a surface surrogate measure and its time derivative. The model is tested with respect to four different surface surrogates and a varying number of surrogate and image acquisitions used for model training. Fourteen patients were imaged 100 times with cine CT, at three sessions mimicking a planning session followed by two treatment fractions. Patient body contours were concurrently detected by a body surface laser scanning system BSLS from which four surface surrogates were extracted; thoracic point TP, abdominal point AP, the radial distance mean RDM, and a surface derived volume SDV. The motion model was trained on session 1 and evaluated on sessions 2 and 3 by comparing modeled tumor positions with measured positions from the cine images. The number of concurrent surrogate and image acquisitions used in the training set was varied, and its impact on the final result was evaluated. The use of AP as a surface surrogate yielded the smallest error in modeled tumor positions. The mean deviation between modeled and measured tumor positions was 1.9 mm. The corresponding deviations for using the other surrogates were 2.0 mm (RDM), 2.8 mm (SDV), and 3.0 mm (TP). To produce a motion model that accurately models the tumor position over several fractions requires at least 10 simultaneous surrogate and image acquisitions over 1–2 minutes. John Wiley and Sons Inc. 2021-07-14 /pmc/articles/PMC8425865/ /pubmed/34258853 http://dx.doi.org/10.1002/acm2.13351 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Wikström, Kenneth A.
Isacsson, Ulf M.
Nilsson, Kristina M.
Ahnesjö, Anders
Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title_full Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title_fullStr Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title_full_unstemmed Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title_short Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
title_sort evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8425865/
https://www.ncbi.nlm.nih.gov/pubmed/34258853
http://dx.doi.org/10.1002/acm2.13351
work_keys_str_mv AT wikstromkennetha evaluationoffoursurfacesurrogatesformodelinglungtumorpositionsoverseveralfractionsinradiotherapy
AT isacssonulfm evaluationoffoursurfacesurrogatesformodelinglungtumorpositionsoverseveralfractionsinradiotherapy
AT nilssonkristinam evaluationoffoursurfacesurrogatesformodelinglungtumorpositionsoverseveralfractionsinradiotherapy
AT ahnesjoanders evaluationoffoursurfacesurrogatesformodelinglungtumorpositionsoverseveralfractionsinradiotherapy