High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its imp...
Autores principales: | , , , , , , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432131/ https://www.ncbi.nlm.nih.gov/pubmed/34501258 http://dx.doi.org/10.3390/jcm10173811 |
_version_ | 1783751092866121728 |
---|---|
author | Kovacs, Boldizsar Burri, Haran Buehler, Andres Reek, Sven Sticherling, Christian Schaer, Beat Linka, Andre Ammann, Peter Müller, Andreas S. Dzemali, Omer Kobza, Richard Schindler, Matthias Haegeli, Laurent Mayer, Kurt Eriksson, Urs Herrera-Siklody, Claudia Reichlin, Tobias Steffel, Jan Saguner, Ardan M. Duru, Firat |
author_facet | Kovacs, Boldizsar Burri, Haran Buehler, Andres Reek, Sven Sticherling, Christian Schaer, Beat Linka, Andre Ammann, Peter Müller, Andreas S. Dzemali, Omer Kobza, Richard Schindler, Matthias Haegeli, Laurent Mayer, Kurt Eriksson, Urs Herrera-Siklody, Claudia Reichlin, Tobias Steffel, Jan Saguner, Ardan M. Duru, Firat |
author_sort | Kovacs, Boldizsar |
collection | PubMed |
description | Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its impact on clinical outcomes. Methods: Patients from the nationwide Swiss WCD Registry were included. Clinical characteristics and data were obtained from the WCDs. Arrhythmia recordings ≥30 s in length were analysed and categorized as VT/VF, atrial fibrillation (AF), supraventricular tachycardia (SVT) or artefact. Results: A total of 10653 device alarms were documented in 324 of 456 patients (71.1%) over a mean WCD wear-time of 2.0 ± 1.6 months. Episode duration was 30 s or more in 2996 alarms (28.2%). One hundred and eleven (3.7%) were VT/VF episodes. The remaining recordings were inappropriate detections (2736 (91%) due to artefacts; 117 (3.7%) AF; 48 (1.6%) SVT). Two-hundred and seven patients (45%) had three or more alarms per month. Obesity was significantly associated with three or more alarms per month (p = 0.01, 27.7% vs. 15.9%). High alarm burden was not associated with a lower average daily wear time (20.8 h vs. 20.7 h, p = 0.785) or a decreased implantable cardioverter defibrillator implantation rate after stopping WCD use (48% vs. 47.3%, p = 0.156). Conclusions: In patients using WCDs, alarms emitted by the device and impending inappropriate shocks were frequent and most commonly caused by artefacts. A high alarm burden was associated with obesity but did not lead to a decreased adherence. |
format | Online Article Text |
id | pubmed-8432131 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-84321312021-09-11 High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry Kovacs, Boldizsar Burri, Haran Buehler, Andres Reek, Sven Sticherling, Christian Schaer, Beat Linka, Andre Ammann, Peter Müller, Andreas S. Dzemali, Omer Kobza, Richard Schindler, Matthias Haegeli, Laurent Mayer, Kurt Eriksson, Urs Herrera-Siklody, Claudia Reichlin, Tobias Steffel, Jan Saguner, Ardan M. Duru, Firat J Clin Med Article Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its impact on clinical outcomes. Methods: Patients from the nationwide Swiss WCD Registry were included. Clinical characteristics and data were obtained from the WCDs. Arrhythmia recordings ≥30 s in length were analysed and categorized as VT/VF, atrial fibrillation (AF), supraventricular tachycardia (SVT) or artefact. Results: A total of 10653 device alarms were documented in 324 of 456 patients (71.1%) over a mean WCD wear-time of 2.0 ± 1.6 months. Episode duration was 30 s or more in 2996 alarms (28.2%). One hundred and eleven (3.7%) were VT/VF episodes. The remaining recordings were inappropriate detections (2736 (91%) due to artefacts; 117 (3.7%) AF; 48 (1.6%) SVT). Two-hundred and seven patients (45%) had three or more alarms per month. Obesity was significantly associated with three or more alarms per month (p = 0.01, 27.7% vs. 15.9%). High alarm burden was not associated with a lower average daily wear time (20.8 h vs. 20.7 h, p = 0.785) or a decreased implantable cardioverter defibrillator implantation rate after stopping WCD use (48% vs. 47.3%, p = 0.156). Conclusions: In patients using WCDs, alarms emitted by the device and impending inappropriate shocks were frequent and most commonly caused by artefacts. A high alarm burden was associated with obesity but did not lead to a decreased adherence. MDPI 2021-08-25 /pmc/articles/PMC8432131/ /pubmed/34501258 http://dx.doi.org/10.3390/jcm10173811 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Kovacs, Boldizsar Burri, Haran Buehler, Andres Reek, Sven Sticherling, Christian Schaer, Beat Linka, Andre Ammann, Peter Müller, Andreas S. Dzemali, Omer Kobza, Richard Schindler, Matthias Haegeli, Laurent Mayer, Kurt Eriksson, Urs Herrera-Siklody, Claudia Reichlin, Tobias Steffel, Jan Saguner, Ardan M. Duru, Firat High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title | High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title_full | High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title_fullStr | High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title_full_unstemmed | High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title_short | High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry |
title_sort | high incidence of inappropriate alarms in patients with wearable cardioverter-defibrillators: findings from the swiss wcd registry |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432131/ https://www.ncbi.nlm.nih.gov/pubmed/34501258 http://dx.doi.org/10.3390/jcm10173811 |
work_keys_str_mv | AT kovacsboldizsar highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT burriharan highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT buehlerandres highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT reeksven highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT sticherlingchristian highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT schaerbeat highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT linkaandre highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT ammannpeter highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT mullerandreass highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT dzemaliomer highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT kobzarichard highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT schindlermatthias highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT haegelilaurent highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT mayerkurt highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT erikssonurs highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT herrerasiklodyclaudia highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT reichlintobias highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT steffeljan highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT sagunerardanm highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry AT durufirat highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry |