High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry

Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its imp...

Descripción completa

Detalles Bibliográficos
Autores principales: Kovacs, Boldizsar, Burri, Haran, Buehler, Andres, Reek, Sven, Sticherling, Christian, Schaer, Beat, Linka, Andre, Ammann, Peter, Müller, Andreas S., Dzemali, Omer, Kobza, Richard, Schindler, Matthias, Haegeli, Laurent, Mayer, Kurt, Eriksson, Urs, Herrera-Siklody, Claudia, Reichlin, Tobias, Steffel, Jan, Saguner, Ardan M., Duru, Firat
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432131/
https://www.ncbi.nlm.nih.gov/pubmed/34501258
http://dx.doi.org/10.3390/jcm10173811
_version_ 1783751092866121728
author Kovacs, Boldizsar
Burri, Haran
Buehler, Andres
Reek, Sven
Sticherling, Christian
Schaer, Beat
Linka, Andre
Ammann, Peter
Müller, Andreas S.
Dzemali, Omer
Kobza, Richard
Schindler, Matthias
Haegeli, Laurent
Mayer, Kurt
Eriksson, Urs
Herrera-Siklody, Claudia
Reichlin, Tobias
Steffel, Jan
Saguner, Ardan M.
Duru, Firat
author_facet Kovacs, Boldizsar
Burri, Haran
Buehler, Andres
Reek, Sven
Sticherling, Christian
Schaer, Beat
Linka, Andre
Ammann, Peter
Müller, Andreas S.
Dzemali, Omer
Kobza, Richard
Schindler, Matthias
Haegeli, Laurent
Mayer, Kurt
Eriksson, Urs
Herrera-Siklody, Claudia
Reichlin, Tobias
Steffel, Jan
Saguner, Ardan M.
Duru, Firat
author_sort Kovacs, Boldizsar
collection PubMed
description Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its impact on clinical outcomes. Methods: Patients from the nationwide Swiss WCD Registry were included. Clinical characteristics and data were obtained from the WCDs. Arrhythmia recordings ≥30 s in length were analysed and categorized as VT/VF, atrial fibrillation (AF), supraventricular tachycardia (SVT) or artefact. Results: A total of 10653 device alarms were documented in 324 of 456 patients (71.1%) over a mean WCD wear-time of 2.0 ± 1.6 months. Episode duration was 30 s or more in 2996 alarms (28.2%). One hundred and eleven (3.7%) were VT/VF episodes. The remaining recordings were inappropriate detections (2736 (91%) due to artefacts; 117 (3.7%) AF; 48 (1.6%) SVT). Two-hundred and seven patients (45%) had three or more alarms per month. Obesity was significantly associated with three or more alarms per month (p = 0.01, 27.7% vs. 15.9%). High alarm burden was not associated with a lower average daily wear time (20.8 h vs. 20.7 h, p = 0.785) or a decreased implantable cardioverter defibrillator implantation rate after stopping WCD use (48% vs. 47.3%, p = 0.156). Conclusions: In patients using WCDs, alarms emitted by the device and impending inappropriate shocks were frequent and most commonly caused by artefacts. A high alarm burden was associated with obesity but did not lead to a decreased adherence.
format Online
Article
Text
id pubmed-8432131
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-84321312021-09-11 High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry Kovacs, Boldizsar Burri, Haran Buehler, Andres Reek, Sven Sticherling, Christian Schaer, Beat Linka, Andre Ammann, Peter Müller, Andreas S. Dzemali, Omer Kobza, Richard Schindler, Matthias Haegeli, Laurent Mayer, Kurt Eriksson, Urs Herrera-Siklody, Claudia Reichlin, Tobias Steffel, Jan Saguner, Ardan M. Duru, Firat J Clin Med Article Background: The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts. Objective: The aim of this study is to assess the alarm burden in patients and its impact on clinical outcomes. Methods: Patients from the nationwide Swiss WCD Registry were included. Clinical characteristics and data were obtained from the WCDs. Arrhythmia recordings ≥30 s in length were analysed and categorized as VT/VF, atrial fibrillation (AF), supraventricular tachycardia (SVT) or artefact. Results: A total of 10653 device alarms were documented in 324 of 456 patients (71.1%) over a mean WCD wear-time of 2.0 ± 1.6 months. Episode duration was 30 s or more in 2996 alarms (28.2%). One hundred and eleven (3.7%) were VT/VF episodes. The remaining recordings were inappropriate detections (2736 (91%) due to artefacts; 117 (3.7%) AF; 48 (1.6%) SVT). Two-hundred and seven patients (45%) had three or more alarms per month. Obesity was significantly associated with three or more alarms per month (p = 0.01, 27.7% vs. 15.9%). High alarm burden was not associated with a lower average daily wear time (20.8 h vs. 20.7 h, p = 0.785) or a decreased implantable cardioverter defibrillator implantation rate after stopping WCD use (48% vs. 47.3%, p = 0.156). Conclusions: In patients using WCDs, alarms emitted by the device and impending inappropriate shocks were frequent and most commonly caused by artefacts. A high alarm burden was associated with obesity but did not lead to a decreased adherence. MDPI 2021-08-25 /pmc/articles/PMC8432131/ /pubmed/34501258 http://dx.doi.org/10.3390/jcm10173811 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Kovacs, Boldizsar
Burri, Haran
Buehler, Andres
Reek, Sven
Sticherling, Christian
Schaer, Beat
Linka, Andre
Ammann, Peter
Müller, Andreas S.
Dzemali, Omer
Kobza, Richard
Schindler, Matthias
Haegeli, Laurent
Mayer, Kurt
Eriksson, Urs
Herrera-Siklody, Claudia
Reichlin, Tobias
Steffel, Jan
Saguner, Ardan M.
Duru, Firat
High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title_full High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title_fullStr High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title_full_unstemmed High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title_short High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry
title_sort high incidence of inappropriate alarms in patients with wearable cardioverter-defibrillators: findings from the swiss wcd registry
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432131/
https://www.ncbi.nlm.nih.gov/pubmed/34501258
http://dx.doi.org/10.3390/jcm10173811
work_keys_str_mv AT kovacsboldizsar highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT burriharan highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT buehlerandres highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT reeksven highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT sticherlingchristian highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT schaerbeat highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT linkaandre highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT ammannpeter highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT mullerandreass highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT dzemaliomer highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT kobzarichard highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT schindlermatthias highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT haegelilaurent highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT mayerkurt highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT erikssonurs highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT herrerasiklodyclaudia highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT reichlintobias highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT steffeljan highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT sagunerardanm highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry
AT durufirat highincidenceofinappropriatealarmsinpatientswithwearablecardioverterdefibrillatorsfindingsfromtheswisswcdregistry