Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging
Purpose: To compare listening ability (speech reception thresholds) and real-life listening experience in users with a percutaneous bone conduction device (BCD) with two listening programs differing only in high-frequency gain. In situ real-life experiences were recorded with ecological momentary as...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432250/ https://www.ncbi.nlm.nih.gov/pubmed/34501371 http://dx.doi.org/10.3390/jcm10173923 |
_version_ | 1783751121639047168 |
---|---|
author | Bosman, Arjan J. Christensen, Jeppe Høy Rosenbom, Tove Patou, François Janssen, Arno Hol, Myrthe K. S. |
author_facet | Bosman, Arjan J. Christensen, Jeppe Høy Rosenbom, Tove Patou, François Janssen, Arno Hol, Myrthe K. S. |
author_sort | Bosman, Arjan J. |
collection | PubMed |
description | Purpose: To compare listening ability (speech reception thresholds) and real-life listening experience in users with a percutaneous bone conduction device (BCD) with two listening programs differing only in high-frequency gain. In situ real-life experiences were recorded with ecological momentary assessment (EMA) techniques combined with real-time acoustical data logging and standard retrospective questionnaires. Methods: Nineteen experienced BCD users participated in this study. They all used a Ponto 4 BCD from Oticon Medical during a 4-week trial period. Environmental data and device parameters (i.e., device usage and volume control) were logged in real-time on an iPhone via a custom iOS research app. At the end of the trial period, subjects filled in APHAB, SSQ, and preference questionnaires. Listening abilities with the two programs were evaluated with speech reception threshold tests. Results: The APHAB and SSQ questionnaires did not reveal any differences between the two listening programs. The EMAs revealed group-level effects, indicating that in speech and noisy listening environments, subjects preferred the default listening program, and found the program with additional high-frequency gain too loud. This finding was corroborated by the volume log—subjects avoided the higher volume control setting and reacted more to changes in environmental sound pressure levels when using the high-frequency gain program. Finally, day-to-day changes in EMAs revealed acclimatization effects in the listening experience for ratings of “sound quality” and “program suitability” of the BCD, but not for ratings of “loudness perception” and “speech understanding”. The acclimatization effect did not differ among the listening programs. Conclusion: Adding custom high-frequency amplification to the BCD target-gain prescription improves speech reception in laboratory tests under quiet conditions, but results in poorer real-life listening experiences due to loudness. |
format | Online Article Text |
id | pubmed-8432250 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-84322502021-09-11 Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging Bosman, Arjan J. Christensen, Jeppe Høy Rosenbom, Tove Patou, François Janssen, Arno Hol, Myrthe K. S. J Clin Med Article Purpose: To compare listening ability (speech reception thresholds) and real-life listening experience in users with a percutaneous bone conduction device (BCD) with two listening programs differing only in high-frequency gain. In situ real-life experiences were recorded with ecological momentary assessment (EMA) techniques combined with real-time acoustical data logging and standard retrospective questionnaires. Methods: Nineteen experienced BCD users participated in this study. They all used a Ponto 4 BCD from Oticon Medical during a 4-week trial period. Environmental data and device parameters (i.e., device usage and volume control) were logged in real-time on an iPhone via a custom iOS research app. At the end of the trial period, subjects filled in APHAB, SSQ, and preference questionnaires. Listening abilities with the two programs were evaluated with speech reception threshold tests. Results: The APHAB and SSQ questionnaires did not reveal any differences between the two listening programs. The EMAs revealed group-level effects, indicating that in speech and noisy listening environments, subjects preferred the default listening program, and found the program with additional high-frequency gain too loud. This finding was corroborated by the volume log—subjects avoided the higher volume control setting and reacted more to changes in environmental sound pressure levels when using the high-frequency gain program. Finally, day-to-day changes in EMAs revealed acclimatization effects in the listening experience for ratings of “sound quality” and “program suitability” of the BCD, but not for ratings of “loudness perception” and “speech understanding”. The acclimatization effect did not differ among the listening programs. Conclusion: Adding custom high-frequency amplification to the BCD target-gain prescription improves speech reception in laboratory tests under quiet conditions, but results in poorer real-life listening experiences due to loudness. MDPI 2021-08-30 /pmc/articles/PMC8432250/ /pubmed/34501371 http://dx.doi.org/10.3390/jcm10173923 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Bosman, Arjan J. Christensen, Jeppe Høy Rosenbom, Tove Patou, François Janssen, Arno Hol, Myrthe K. S. Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title | Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title_full | Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title_fullStr | Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title_full_unstemmed | Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title_short | Investigating Real-World Benefits of High-Frequency Gain in Bone-Anchored Users with Ecological Momentary Assessment and Real-Time Data Logging |
title_sort | investigating real-world benefits of high-frequency gain in bone-anchored users with ecological momentary assessment and real-time data logging |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8432250/ https://www.ncbi.nlm.nih.gov/pubmed/34501371 http://dx.doi.org/10.3390/jcm10173923 |
work_keys_str_mv | AT bosmanarjanj investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging AT christensenjeppehøy investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging AT rosenbomtove investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging AT patoufrancois investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging AT janssenarno investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging AT holmyrtheks investigatingrealworldbenefitsofhighfrequencygaininboneanchoreduserswithecologicalmomentaryassessmentandrealtimedatalogging |