Cargando…

Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System

Background: Closed loop bi-hormonal artificial pancreas systems, such as the artificial pancreas (AP™) developed by Inreda Diabetic B.V., control blood glucose levels of type 1 diabetes mellitus patients via closed loop regulation. As the AP™ currently does not classify postures and movements to est...

Descripción completa

Detalles Bibliográficos
Autores principales: Sawaryn, Ben, Klaassen, Michel, van Beijnum, Bert-Jan, Zwart, Hans, Veltink, Peter H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8434663/
https://www.ncbi.nlm.nih.gov/pubmed/34502845
http://dx.doi.org/10.3390/s21175954
_version_ 1783751653324750848
author Sawaryn, Ben
Klaassen, Michel
van Beijnum, Bert-Jan
Zwart, Hans
Veltink, Peter H.
author_facet Sawaryn, Ben
Klaassen, Michel
van Beijnum, Bert-Jan
Zwart, Hans
Veltink, Peter H.
author_sort Sawaryn, Ben
collection PubMed
description Background: Closed loop bi-hormonal artificial pancreas systems, such as the artificial pancreas (AP™) developed by Inreda Diabetic B.V., control blood glucose levels of type 1 diabetes mellitus patients via closed loop regulation. As the AP™ currently does not classify postures and movements to estimate metabolic energy consumption to correct hormone administration levels, considerable improvements to the system can be made. Therefore, this research aimed to investigate the possibility to use the current system to identify several postures and movements. Methods: seven healthy participants took part in an experiment where sequences of postures and movements were performed to train and assess a computationally sparing algorithm. Results: Using accelerometers, one on the hip and two on the abdomen, user-specific models achieved classification accuracies of 86.5% using only the hip sensor and 87.3% when including the abdomen sensors. With additional accelerometers on the sternum and upper leg for identification, 90.0% of the classified postures and movements were correct. Conclusions: The current hardware configuration of the AP™ poses no limitation to the identification of postures and movements. If future research shows that identification can still be done accurately in a daily life setting, this algorithm may be an improvement for the AP™ to sense physical activity.
format Online
Article
Text
id pubmed-8434663
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-84346632021-09-12 Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System Sawaryn, Ben Klaassen, Michel van Beijnum, Bert-Jan Zwart, Hans Veltink, Peter H. Sensors (Basel) Article Background: Closed loop bi-hormonal artificial pancreas systems, such as the artificial pancreas (AP™) developed by Inreda Diabetic B.V., control blood glucose levels of type 1 diabetes mellitus patients via closed loop regulation. As the AP™ currently does not classify postures and movements to estimate metabolic energy consumption to correct hormone administration levels, considerable improvements to the system can be made. Therefore, this research aimed to investigate the possibility to use the current system to identify several postures and movements. Methods: seven healthy participants took part in an experiment where sequences of postures and movements were performed to train and assess a computationally sparing algorithm. Results: Using accelerometers, one on the hip and two on the abdomen, user-specific models achieved classification accuracies of 86.5% using only the hip sensor and 87.3% when including the abdomen sensors. With additional accelerometers on the sternum and upper leg for identification, 90.0% of the classified postures and movements were correct. Conclusions: The current hardware configuration of the AP™ poses no limitation to the identification of postures and movements. If future research shows that identification can still be done accurately in a daily life setting, this algorithm may be an improvement for the AP™ to sense physical activity. MDPI 2021-09-05 /pmc/articles/PMC8434663/ /pubmed/34502845 http://dx.doi.org/10.3390/s21175954 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Sawaryn, Ben
Klaassen, Michel
van Beijnum, Bert-Jan
Zwart, Hans
Veltink, Peter H.
Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title_full Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title_fullStr Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title_full_unstemmed Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title_short Identification of Movements and Postures Using Wearable Sensors for Implementation in a Bi-Hormonal Artificial Pancreas System
title_sort identification of movements and postures using wearable sensors for implementation in a bi-hormonal artificial pancreas system
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8434663/
https://www.ncbi.nlm.nih.gov/pubmed/34502845
http://dx.doi.org/10.3390/s21175954
work_keys_str_mv AT sawarynben identificationofmovementsandposturesusingwearablesensorsforimplementationinabihormonalartificialpancreassystem
AT klaassenmichel identificationofmovementsandposturesusingwearablesensorsforimplementationinabihormonalartificialpancreassystem
AT vanbeijnumbertjan identificationofmovementsandposturesusingwearablesensorsforimplementationinabihormonalartificialpancreassystem
AT zwarthans identificationofmovementsandposturesusingwearablesensorsforimplementationinabihormonalartificialpancreassystem
AT veltinkpeterh identificationofmovementsandposturesusingwearablesensorsforimplementationinabihormonalartificialpancreassystem