A gain series method for accurate EMCCD calibration
Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model mo...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8443689/ https://www.ncbi.nlm.nih.gov/pubmed/34526588 http://dx.doi.org/10.1038/s41598-021-97759-6 |
Sumario: | Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model more accurately estimates the gain response of the camera. Furthermore, we develop a method to compare the results from different calibration approaches. Gain-series calibration outperforms all other methods in this self-consistency test. |
---|