Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium

The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis, 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genita...

Descripción completa

Detalles Bibliográficos
Autores principales: Herrmann, Björn, Malm, Karin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8449771/
https://www.ncbi.nlm.nih.gov/pubmed/33723736
http://dx.doi.org/10.1007/s10096-020-04135-9
Descripción
Sumario:The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis, 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genitalium, 22 (7.9%) of 279 samples were positive in the Alinity assay and 18 (6.5%) in the m2000 assay, for which DNA extraction was performed on an m2000sp instrument combined with in-house real-time PCR. The Alinity assay has at least the same sensitivity as the m2000 assay. The specificity was evaluated by discrepancy analysis. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s10096-020-04135-9.