Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks
Background: Night-time heart rate (HR) is expected to reflect more accurately the cardiac autonomic function of modulating cardiovascular activity. Few studies have been conducted on the predictive values of night-time HR in relation to cardioverter-defibrillator therapies. Aims: To explore the asso...
Autores principales: | , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8450398/ https://www.ncbi.nlm.nih.gov/pubmed/34552968 http://dx.doi.org/10.3389/fcvm.2021.739889 |
_version_ | 1784569636552966144 |
---|---|
author | Sun, Xuerong Zhou, Bin Chen, Keping Hua, Wei Su, Yangang Xu, Wei Wang, Fang Fan, Xiaohan Niu, Hongxia Dai, Yan Liu, Zhimin Zhao, Shuang Zhang, Shu |
author_facet | Sun, Xuerong Zhou, Bin Chen, Keping Hua, Wei Su, Yangang Xu, Wei Wang, Fang Fan, Xiaohan Niu, Hongxia Dai, Yan Liu, Zhimin Zhao, Shuang Zhang, Shu |
author_sort | Sun, Xuerong |
collection | PubMed |
description | Background: Night-time heart rate (HR) is expected to reflect more accurately the cardiac autonomic function of modulating cardiovascular activity. Few studies have been conducted on the predictive values of night-time HR in relation to cardioverter-defibrillator therapies. Aims: To explore the associations of night-time HR with the ventricular tachyarrhythmias (VTAs), appropriate and inappropriate implantable cardioverter-defibrillator (ICD) shocks. Methods: Patients from the SUMMIT registry receiving ICD or cardiac resynchronization therapy with defibrillator (CRT-D) implantation were retrospectively analyzed using archived home monitoring data. Night-time HR was recorded from 2:00 am to 6:00 am during the first 30 to 60 days after implantation. VTA events and ICD shocks were identified using the intracardiac electrograms by two independent physicians. Restricted cubic splines and smooth curve fitting were conducted to address the non-linear associations between night-time HR and adjusted hazards for clinical outcomes. Results: Over a mean follow-up duration of 55.8 ± 22.7 months, 187 deaths were observed among 730 patients. VTAs, appropriate and inappropriate ICD shocks were observed in 422 (57.8%), 293 (40.1%), and 72 (10.0%) patients, respectively. Apparent U-shaped non-linear associations of night-time HR with VTAs (P for non-linearity = 0.007), appropriate ICD shocks (P for non-linearity = 0.003) and inappropriate ICD shocks (P for non-linearity = 0.014) were detected. When night-time HR was beyond 60 bpm, every 1 bpm increase in night-time HR could result in 3.2, 3.3, and 4.9% higher risks of VTAs and appropriate and inappropriate ICD shocks, respectively; when night-time HR was lower than 60 bpm, every 1 bpm increase in night-time HR could result in 6.0 and 10.7% lower risks of appropriate and inappropriate ICD shocks. Compared to night-time HR of ≤ 50 or ≥70 bpm, night-time HR of 50–70 bpm was associated with 24.9, 30.2, 63.5, and 31.5% reduced incidences of VTA events, appropriate ICD shocks, inappropriate ICD shocks, and all-cause mortality, respectively. Conclusion: Apparent non-linear associations of night-time HR with VTAs and ICD shocks were detected. An increasing incidence of VTAs and ICD shocks was observed at both low and high levels of night-time HR. Night-time HR of 50–70 bpm might be the optimal therapeutics target for the management of ICD/CRT-D recipients. |
format | Online Article Text |
id | pubmed-8450398 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-84503982021-09-21 Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks Sun, Xuerong Zhou, Bin Chen, Keping Hua, Wei Su, Yangang Xu, Wei Wang, Fang Fan, Xiaohan Niu, Hongxia Dai, Yan Liu, Zhimin Zhao, Shuang Zhang, Shu Front Cardiovasc Med Cardiovascular Medicine Background: Night-time heart rate (HR) is expected to reflect more accurately the cardiac autonomic function of modulating cardiovascular activity. Few studies have been conducted on the predictive values of night-time HR in relation to cardioverter-defibrillator therapies. Aims: To explore the associations of night-time HR with the ventricular tachyarrhythmias (VTAs), appropriate and inappropriate implantable cardioverter-defibrillator (ICD) shocks. Methods: Patients from the SUMMIT registry receiving ICD or cardiac resynchronization therapy with defibrillator (CRT-D) implantation were retrospectively analyzed using archived home monitoring data. Night-time HR was recorded from 2:00 am to 6:00 am during the first 30 to 60 days after implantation. VTA events and ICD shocks were identified using the intracardiac electrograms by two independent physicians. Restricted cubic splines and smooth curve fitting were conducted to address the non-linear associations between night-time HR and adjusted hazards for clinical outcomes. Results: Over a mean follow-up duration of 55.8 ± 22.7 months, 187 deaths were observed among 730 patients. VTAs, appropriate and inappropriate ICD shocks were observed in 422 (57.8%), 293 (40.1%), and 72 (10.0%) patients, respectively. Apparent U-shaped non-linear associations of night-time HR with VTAs (P for non-linearity = 0.007), appropriate ICD shocks (P for non-linearity = 0.003) and inappropriate ICD shocks (P for non-linearity = 0.014) were detected. When night-time HR was beyond 60 bpm, every 1 bpm increase in night-time HR could result in 3.2, 3.3, and 4.9% higher risks of VTAs and appropriate and inappropriate ICD shocks, respectively; when night-time HR was lower than 60 bpm, every 1 bpm increase in night-time HR could result in 6.0 and 10.7% lower risks of appropriate and inappropriate ICD shocks. Compared to night-time HR of ≤ 50 or ≥70 bpm, night-time HR of 50–70 bpm was associated with 24.9, 30.2, 63.5, and 31.5% reduced incidences of VTA events, appropriate ICD shocks, inappropriate ICD shocks, and all-cause mortality, respectively. Conclusion: Apparent non-linear associations of night-time HR with VTAs and ICD shocks were detected. An increasing incidence of VTAs and ICD shocks was observed at both low and high levels of night-time HR. Night-time HR of 50–70 bpm might be the optimal therapeutics target for the management of ICD/CRT-D recipients. Frontiers Media S.A. 2021-09-06 /pmc/articles/PMC8450398/ /pubmed/34552968 http://dx.doi.org/10.3389/fcvm.2021.739889 Text en Copyright © 2021 Sun, Zhou, Chen, Hua, Su, Xu, Wang, Fan, Niu, Dai, Liu, Zhao and Zhang. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Cardiovascular Medicine Sun, Xuerong Zhou, Bin Chen, Keping Hua, Wei Su, Yangang Xu, Wei Wang, Fang Fan, Xiaohan Niu, Hongxia Dai, Yan Liu, Zhimin Zhao, Shuang Zhang, Shu Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title | Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title_full | Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title_fullStr | Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title_full_unstemmed | Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title_short | Association of Night-Time Heart Rate With Ventricular Tachyarrhythmias, Appropriate and Inappropriate Implantable Cardioverter-Defibrillator Shocks |
title_sort | association of night-time heart rate with ventricular tachyarrhythmias, appropriate and inappropriate implantable cardioverter-defibrillator shocks |
topic | Cardiovascular Medicine |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8450398/ https://www.ncbi.nlm.nih.gov/pubmed/34552968 http://dx.doi.org/10.3389/fcvm.2021.739889 |
work_keys_str_mv | AT sunxuerong associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT zhoubin associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT chenkeping associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT huawei associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT suyangang associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT xuwei associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT wangfang associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT fanxiaohan associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT niuhongxia associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT daiyan associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT liuzhimin associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT zhaoshuang associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks AT zhangshu associationofnighttimeheartratewithventriculartachyarrhythmiasappropriateandinappropriateimplantablecardioverterdefibrillatorshocks |