Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system
BACKGROUND: DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including a 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use. METHODS: The analyzer was evaluated according to the Inte...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Society of Medical Biochemists of Serbia, Belgrade
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8451233/ https://www.ncbi.nlm.nih.gov/pubmed/34616226 http://dx.doi.org/10.5937/jomb0-28836 |
_version_ | 1784569800402403328 |
---|---|
author | Velizarova, Milena Yacheva, Teodora Genova, Mariana Svinarov, Dobrin |
author_facet | Velizarova, Milena Yacheva, Teodora Genova, Mariana Svinarov, Dobrin |
author_sort | Velizarova, Milena |
collection | PubMed |
description | BACKGROUND: DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including a 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use. METHODS: The analyzer was evaluated according to the International Council for Standardization in Haematology guideline. The purposes of this study were to assess its analytical performance in comparison to SYSMEX XN 1000 hematology analyzer currently used in our laboratory, as well as to compare the automated and manual WBC differential. RESULTS: Within-run precision in all concentration ranges was very good with coefficients of variation (CVs) between 0.02% and 2.5% except for platelets over 500×109/L (CV 9.5%). Within-batch imprecision showed CVs lower the declared deviation ranges. Accuracy (defined as trueness) was excellent for all CBC and white cell differential parameters, compared with the state of the art%. Linearity was confirmed with excellent regression coefficients (0.999-1.000), even in the lowest values, and carryover was ≤ 1%. Comparison between DYMIND DH76 and SYSMEX XN 1000 was also very good with correlation coefficients (R2) for WBC (1.000), RBC (0.999), hemoglobin (0.999) and PLT over 50×10(9)/L (0.994) and R2 was lower but still acceptable (0.910) for PLT<50×10(9)/L. R2 for neutrophils, lymphocytes, eosinophils, basophils, and monocytes were 0.974, 0.982, 0.957, 0.625, and 0.836, respectively, in the comparison between the manual and DYMIND DH76 automated differential WBC counts. CONCLUSIONS: With excellent analytical performance and acceptable comparative analysis, DYMIND DH76 hematology analyser covered the predefined international standards and requirements and is fully appropriate for clinical application. |
format | Online Article Text |
id | pubmed-8451233 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Society of Medical Biochemists of Serbia, Belgrade |
record_format | MEDLINE/PubMed |
spelling | pubmed-84512332021-10-05 Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system Velizarova, Milena Yacheva, Teodora Genova, Mariana Svinarov, Dobrin J Med Biochem Original Paper BACKGROUND: DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including a 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use. METHODS: The analyzer was evaluated according to the International Council for Standardization in Haematology guideline. The purposes of this study were to assess its analytical performance in comparison to SYSMEX XN 1000 hematology analyzer currently used in our laboratory, as well as to compare the automated and manual WBC differential. RESULTS: Within-run precision in all concentration ranges was very good with coefficients of variation (CVs) between 0.02% and 2.5% except for platelets over 500×109/L (CV 9.5%). Within-batch imprecision showed CVs lower the declared deviation ranges. Accuracy (defined as trueness) was excellent for all CBC and white cell differential parameters, compared with the state of the art%. Linearity was confirmed with excellent regression coefficients (0.999-1.000), even in the lowest values, and carryover was ≤ 1%. Comparison between DYMIND DH76 and SYSMEX XN 1000 was also very good with correlation coefficients (R2) for WBC (1.000), RBC (0.999), hemoglobin (0.999) and PLT over 50×10(9)/L (0.994) and R2 was lower but still acceptable (0.910) for PLT<50×10(9)/L. R2 for neutrophils, lymphocytes, eosinophils, basophils, and monocytes were 0.974, 0.982, 0.957, 0.625, and 0.836, respectively, in the comparison between the manual and DYMIND DH76 automated differential WBC counts. CONCLUSIONS: With excellent analytical performance and acceptable comparative analysis, DYMIND DH76 hematology analyser covered the predefined international standards and requirements and is fully appropriate for clinical application. Society of Medical Biochemists of Serbia, Belgrade 2021-09-03 2021-09-03 /pmc/articles/PMC8451233/ /pubmed/34616226 http://dx.doi.org/10.5937/jomb0-28836 Text en 2021 Milena Velizarova, Teodora Yacheva, Mariana Genova, Dobrin Svinarov, published by CEON/CEES https://creativecommons.org/licenses/by/4.0/This work is licensed under the Creative Commons Attribution 4.0 License. |
spellingShingle | Original Paper Velizarova, Milena Yacheva, Teodora Genova, Mariana Svinarov, Dobrin Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title | Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title_full | Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title_fullStr | Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title_full_unstemmed | Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title_short | Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system |
title_sort | evaluation of automated hematology analyzer dymind dh76 compared to sysmex xn 1000 system |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8451233/ https://www.ncbi.nlm.nih.gov/pubmed/34616226 http://dx.doi.org/10.5937/jomb0-28836 |
work_keys_str_mv | AT velizarovamilena evaluationofautomatedhematologyanalyzerdyminddh76comparedtosysmexxn1000system AT yachevateodora evaluationofautomatedhematologyanalyzerdyminddh76comparedtosysmexxn1000system AT genovamariana evaluationofautomatedhematologyanalyzerdyminddh76comparedtosysmexxn1000system AT svinarovdobrin evaluationofautomatedhematologyanalyzerdyminddh76comparedtosysmexxn1000system |