Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement

OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR meas...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Chenguang, Yang, Junqing, Dong, Shaohong, Dong, Liang, Chen, Jiyan, Shen, Li, Zhang, Feng, Li, Changling, Liu, Huadong, Hu, Xinyang, Hau, William Kongto, Qian, Juying, Jeremias, Allen, Wang, Jian'an, Ge, Junbo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley & Sons, Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8453920/
https://www.ncbi.nlm.nih.gov/pubmed/33951285
http://dx.doi.org/10.1002/ccd.29678
_version_ 1784570379518345216
author Li, Chenguang
Yang, Junqing
Dong, Shaohong
Dong, Liang
Chen, Jiyan
Shen, Li
Zhang, Feng
Li, Changling
Liu, Huadong
Hu, Xinyang
Hau, William Kongto
Qian, Juying
Jeremias, Allen
Wang, Jian'an
Ge, Junbo
author_facet Li, Chenguang
Yang, Junqing
Dong, Shaohong
Dong, Liang
Chen, Jiyan
Shen, Li
Zhang, Feng
Li, Changling
Liu, Huadong
Hu, Xinyang
Hau, William Kongto
Qian, Juying
Jeremias, Allen
Wang, Jian'an
Ge, Junbo
author_sort Li, Chenguang
collection PubMed
description OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. METHODS: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. RESULTS: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43% less (9.5 vs. 16.7%, p = .014). CONCLUSIONS: Our study showed the MEMS‐PMC is safe to use and has a minimal bias equal to the resolution of current FFR systems. Given the MEMS‐PMC's high measurement accuracy and rapid‐exchange nature, it may become an attractive new tool facilitating routine coronary physiology assessment.
format Online
Article
Text
id pubmed-8453920
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley & Sons, Inc.
record_format MEDLINE/PubMed
spelling pubmed-84539202021-09-27 Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement Li, Chenguang Yang, Junqing Dong, Shaohong Dong, Liang Chen, Jiyan Shen, Li Zhang, Feng Li, Changling Liu, Huadong Hu, Xinyang Hau, William Kongto Qian, Juying Jeremias, Allen Wang, Jian'an Ge, Junbo Catheter Cardiovasc Interv Coronary Artery Disease OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. METHODS: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. RESULTS: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43% less (9.5 vs. 16.7%, p = .014). CONCLUSIONS: Our study showed the MEMS‐PMC is safe to use and has a minimal bias equal to the resolution of current FFR systems. Given the MEMS‐PMC's high measurement accuracy and rapid‐exchange nature, it may become an attractive new tool facilitating routine coronary physiology assessment. John Wiley & Sons, Inc. 2021-05-05 2021-08-01 /pmc/articles/PMC8453920/ /pubmed/33951285 http://dx.doi.org/10.1002/ccd.29678 Text en © 2021 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Coronary Artery Disease
Li, Chenguang
Yang, Junqing
Dong, Shaohong
Dong, Liang
Chen, Jiyan
Shen, Li
Zhang, Feng
Li, Changling
Liu, Huadong
Hu, Xinyang
Hau, William Kongto
Qian, Juying
Jeremias, Allen
Wang, Jian'an
Ge, Junbo
Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title_full Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title_fullStr Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title_full_unstemmed Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title_short Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
title_sort multicenter clinical evaluation of a piezoresistive‐mems‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
topic Coronary Artery Disease
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8453920/
https://www.ncbi.nlm.nih.gov/pubmed/33951285
http://dx.doi.org/10.1002/ccd.29678
work_keys_str_mv AT lichenguang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT yangjunqing multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT dongshaohong multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT dongliang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT chenjiyan multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT shenli multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT zhangfeng multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT lichangling multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT liuhuadong multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT huxinyang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT hauwilliamkongto multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT qianjuying multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT jeremiasallen multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT wangjianan multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT gejunbo multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement
AT multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement