Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR meas...
Autores principales: | , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley & Sons, Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8453920/ https://www.ncbi.nlm.nih.gov/pubmed/33951285 http://dx.doi.org/10.1002/ccd.29678 |
_version_ | 1784570379518345216 |
---|---|
author | Li, Chenguang Yang, Junqing Dong, Shaohong Dong, Liang Chen, Jiyan Shen, Li Zhang, Feng Li, Changling Liu, Huadong Hu, Xinyang Hau, William Kongto Qian, Juying Jeremias, Allen Wang, Jian'an Ge, Junbo |
author_facet | Li, Chenguang Yang, Junqing Dong, Shaohong Dong, Liang Chen, Jiyan Shen, Li Zhang, Feng Li, Changling Liu, Huadong Hu, Xinyang Hau, William Kongto Qian, Juying Jeremias, Allen Wang, Jian'an Ge, Junbo |
author_sort | Li, Chenguang |
collection | PubMed |
description | OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. METHODS: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. RESULTS: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43% less (9.5 vs. 16.7%, p = .014). CONCLUSIONS: Our study showed the MEMS‐PMC is safe to use and has a minimal bias equal to the resolution of current FFR systems. Given the MEMS‐PMC's high measurement accuracy and rapid‐exchange nature, it may become an attractive new tool facilitating routine coronary physiology assessment. |
format | Online Article Text |
id | pubmed-8453920 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley & Sons, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-84539202021-09-27 Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement Li, Chenguang Yang, Junqing Dong, Shaohong Dong, Liang Chen, Jiyan Shen, Li Zhang, Feng Li, Changling Liu, Huadong Hu, Xinyang Hau, William Kongto Qian, Juying Jeremias, Allen Wang, Jian'an Ge, Junbo Catheter Cardiovasc Interv Coronary Artery Disease OBJECTIVES: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. BACKGROUND: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. METHODS: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. RESULTS: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43% less (9.5 vs. 16.7%, p = .014). CONCLUSIONS: Our study showed the MEMS‐PMC is safe to use and has a minimal bias equal to the resolution of current FFR systems. Given the MEMS‐PMC's high measurement accuracy and rapid‐exchange nature, it may become an attractive new tool facilitating routine coronary physiology assessment. John Wiley & Sons, Inc. 2021-05-05 2021-08-01 /pmc/articles/PMC8453920/ /pubmed/33951285 http://dx.doi.org/10.1002/ccd.29678 Text en © 2021 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Coronary Artery Disease Li, Chenguang Yang, Junqing Dong, Shaohong Dong, Liang Chen, Jiyan Shen, Li Zhang, Feng Li, Changling Liu, Huadong Hu, Xinyang Hau, William Kongto Qian, Juying Jeremias, Allen Wang, Jian'an Ge, Junbo Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title | Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title_full | Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title_fullStr | Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title_full_unstemmed | Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title_short | Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
title_sort | multicenter clinical evaluation of a piezoresistive‐mems‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement |
topic | Coronary Artery Disease |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8453920/ https://www.ncbi.nlm.nih.gov/pubmed/33951285 http://dx.doi.org/10.1002/ccd.29678 |
work_keys_str_mv | AT lichenguang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT yangjunqing multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT dongshaohong multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT dongliang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT chenjiyan multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT shenli multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT zhangfeng multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT lichangling multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT liuhuadong multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT huxinyang multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT hauwilliamkongto multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT qianjuying multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT jeremiasallen multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT wangjianan multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT gejunbo multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement AT multicenterclinicalevaluationofapiezoresistivememssensorrapidexchangepressuremicrocathetersystemforfractionalflowreservemeasurement |