Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report

BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed...

Descripción completa

Detalles Bibliográficos
Autores principales: Patel, Bhavisha, Monkhouse, Christopher, Manisty, Charlotte, Papageorgiou, Nikos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8454201/
https://www.ncbi.nlm.nih.gov/pubmed/34557630
http://dx.doi.org/10.1093/ehjcr/ytab198
_version_ 1784570439491649536
author Patel, Bhavisha
Monkhouse, Christopher
Manisty, Charlotte
Papageorgiou, Nikos
author_facet Patel, Bhavisha
Monkhouse, Christopher
Manisty, Charlotte
Papageorgiou, Nikos
author_sort Patel, Bhavisha
collection PubMed
description BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. CASE SUMMARY: A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. DISCUSSION: We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication.
format Online
Article
Text
id pubmed-8454201
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-84542012021-09-22 Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report Patel, Bhavisha Monkhouse, Christopher Manisty, Charlotte Papageorgiou, Nikos Eur Heart J Case Rep Case Report BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. CASE SUMMARY: A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. DISCUSSION: We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication. Oxford University Press 2021-09-21 /pmc/articles/PMC8454201/ /pubmed/34557630 http://dx.doi.org/10.1093/ehjcr/ytab198 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Case Report
Patel, Bhavisha
Monkhouse, Christopher
Manisty, Charlotte
Papageorgiou, Nikos
Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title_full Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title_fullStr Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title_full_unstemmed Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title_short Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
title_sort temporary device malfunction of an mr conditional cardiac resynchronization defibrillator when undergoing mri without appropriate re-programming: a case report
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8454201/
https://www.ncbi.nlm.nih.gov/pubmed/34557630
http://dx.doi.org/10.1093/ehjcr/ytab198
work_keys_str_mv AT patelbhavisha temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport
AT monkhousechristopher temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport
AT manistycharlotte temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport
AT papageorgiounikos temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport