Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8454201/ https://www.ncbi.nlm.nih.gov/pubmed/34557630 http://dx.doi.org/10.1093/ehjcr/ytab198 |
_version_ | 1784570439491649536 |
---|---|
author | Patel, Bhavisha Monkhouse, Christopher Manisty, Charlotte Papageorgiou, Nikos |
author_facet | Patel, Bhavisha Monkhouse, Christopher Manisty, Charlotte Papageorgiou, Nikos |
author_sort | Patel, Bhavisha |
collection | PubMed |
description | BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. CASE SUMMARY: A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. DISCUSSION: We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication. |
format | Online Article Text |
id | pubmed-8454201 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-84542012021-09-22 Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report Patel, Bhavisha Monkhouse, Christopher Manisty, Charlotte Papageorgiou, Nikos Eur Heart J Case Rep Case Report BACKGROUND: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. CASE SUMMARY: A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. DISCUSSION: We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication. Oxford University Press 2021-09-21 /pmc/articles/PMC8454201/ /pubmed/34557630 http://dx.doi.org/10.1093/ehjcr/ytab198 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Case Report Patel, Bhavisha Monkhouse, Christopher Manisty, Charlotte Papageorgiou, Nikos Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title | Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title_full | Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title_fullStr | Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title_full_unstemmed | Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title_short | Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report |
title_sort | temporary device malfunction of an mr conditional cardiac resynchronization defibrillator when undergoing mri without appropriate re-programming: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8454201/ https://www.ncbi.nlm.nih.gov/pubmed/34557630 http://dx.doi.org/10.1093/ehjcr/ytab198 |
work_keys_str_mv | AT patelbhavisha temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport AT monkhousechristopher temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport AT manistycharlotte temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport AT papageorgiounikos temporarydevicemalfunctionofanmrconditionalcardiacresynchronizationdefibrillatorwhenundergoingmriwithoutappropriatereprogrammingacasereport |