Stability of Intraocular Lens With Different Haptic Design: A Swept-Source Optical Coherence Tomography Study

Purpose: To investigate the stability of intraocular lens (IOLs) with different haptics by swept-source anterior-segment optical coherence tomography (AS-OCT). Methods: Sixty-eight eyes from 55 patients received the implantation of Rayner 920H (Closed C-loop Group), Zeiss 509M (Plate Group) or Lenst...

Descripción completa

Detalles Bibliográficos
Autores principales: Xiao, Zixuan, Wang, Geng, Zhen, Miaoru, Zhao, Zifeng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8455909/
https://www.ncbi.nlm.nih.gov/pubmed/34568368
http://dx.doi.org/10.3389/fmed.2021.705873
Descripción
Sumario:Purpose: To investigate the stability of intraocular lens (IOLs) with different haptics by swept-source anterior-segment optical coherence tomography (AS-OCT). Methods: Sixty-eight eyes from 55 patients received the implantation of Rayner 920H (Closed C-loop Group), Zeiss 509M (Plate Group) or Lenstec SOFTEC HD (C-loop Group) IOLs. The tilt and decentration of IOLs were evaluated using AS-OCT at least 1 month postoperatively. Results: Mean decentration and tilt of IOLs were 0.18 ± 0.12 mm (range 0.02 to 0.59 mm) and 5.63 ± 1.65° (range 2.2 to 9.6°) respectively. Decentration was significantly smaller in the plate haptic group (0.12 ± 0.06 mm) as compared to the C-loop group (0.22 ± 0.13 mm, P = 0.02). The tilt of IOL was also significantly smaller in the plate haptic group (4.96 ± 0.89°) as compared to the C-loop group (6.28 ± 1.83°, P = 0.01). There was marginal difference between the Closed C-loop group (5.52 ± 1.74°) and C-loop group (6.28 ± 1.83°, P = 0.07). Conclusions: The Plate-haptic IOLs should have better stability for the decentration and tilt than the C-loop design IOLs.