Cargando…

A comparative biomechanical study of the Distal Tibia Nail against compression plating for the osteosynthesis of supramalleolar corrective osteotomies

The Distal Tibia Nail (DTN; Mizuho, Japan) has demonstrated higher biomechanical stiffness to locking plates in previous research for A3 distal tibia fractures. It is here investigated as a fixation option for supramalleolar corrective osteotomies (SMOT). Sixteen Sawbones tibiae were implanted with...

Descripción completa

Detalles Bibliográficos
Autores principales: Greenfield, Julia, Appelmann, Philipp, Lafon, Yoann, Bruyère-Garnier, Karine, Rommens, Pol Maria, Kuhn, Sebastian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8458440/
https://www.ncbi.nlm.nih.gov/pubmed/34552116
http://dx.doi.org/10.1038/s41598-021-97968-z
Descripción
Sumario:The Distal Tibia Nail (DTN; Mizuho, Japan) has demonstrated higher biomechanical stiffness to locking plates in previous research for A3 distal tibia fractures. It is here investigated as a fixation option for supramalleolar corrective osteotomies (SMOT). Sixteen Sawbones tibiae were implanted with either a DTN (n = 8) or Medial Distal Tibia Plate (MDTP; n = 8) and a SMOT simulated. Two surgical outcome scenarios were envisaged: “best-case” representing an intact lateral cortex, and “worst-case” representing a fractured lateral cortex. All samples were subjected to compressive (350 N, 700 N) and torsional (± 4 Nm, ± 8 Nm) testing. Samples were evaluated using calculated construct stiffness from force–displacement data, interfragmentary movement and Von Mises’ strain distribution. The DTN demonstrated a greater compressive stiffness for the best-case surgical scenario, whereas the MDTP showed higher stiffness (p < 0.05) for the worst-case surgical scenario. In torsional testing, the DTN proved more resistant to torsion in the worst-case surgical setup (p < 0.05) for both ± 4 Nm and ± 8 Nm. The equivalent stiffness of the DTN against the MDTP supports the use of this implant for SMOT fixation and should be considered as a treatment option particularly in patients presenting vascularisation problems where the MDTP is an inappropriate choice.