Cargando…

Error Analysis of the Combined-Scan High-Speed Atomic Force Microscopy

A combined tip-sample scanning architecture can improve the imaging speed of atomic force microscopy (AFM). However, the nonorthogonality between the three scanners and the nonideal response of each scanner cause measurement errors. In this article, the authors systematically analyze the influence o...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Lu, Kong, Ming, Wu, Sen, Xu, Xinke, Wang, Daodang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8471258/
https://www.ncbi.nlm.nih.gov/pubmed/34577346
http://dx.doi.org/10.3390/s21186139
Descripción
Sumario:A combined tip-sample scanning architecture can improve the imaging speed of atomic force microscopy (AFM). However, the nonorthogonality between the three scanners and the nonideal response of each scanner cause measurement errors. In this article, the authors systematically analyze the influence of the installation and response errors of the combined scanning architecture. The experimental results show that when the probe in the homemade high-speed AFM moves with the Z-scanner, the spot position on the four-quadrant detector changes, thus introducing measurement error. Comparing the experimental results with the numerical and theoretical results shows that the undesired motion of the Z-scanner introduces a large error. The authors believe that this significant error occurs because the piezoelectric actuator not only stretches along the polarization direction but also swings under nonuniform multifield coupling. This article proposes a direction for further optimizing the instrument and provides design ideas for similar high-speed atomic force microscopes.