Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study

BACKGROUND: The digital cadmium–zinc–telluride (CZT)-based SPECT system has many advantages, including better spatial and energy resolution. However, the impacts of different acquisition and reconstruction parameters on CZT SPECT quantification might still need to be validated. This study aimed to e...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Ruyi, Wang, Miao, Zhou, Yaqian, Wang, Shen, Shen, Yiming, Li, Ning, Wang, Peng, Tan, Jian, Meng, Zhaowei, Jia, Qiang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8473509/
https://www.ncbi.nlm.nih.gov/pubmed/34568990
http://dx.doi.org/10.1186/s40658-021-00412-4
_version_ 1784575007047811072
author Zhang, Ruyi
Wang, Miao
Zhou, Yaqian
Wang, Shen
Shen, Yiming
Li, Ning
Wang, Peng
Tan, Jian
Meng, Zhaowei
Jia, Qiang
author_facet Zhang, Ruyi
Wang, Miao
Zhou, Yaqian
Wang, Shen
Shen, Yiming
Li, Ning
Wang, Peng
Tan, Jian
Meng, Zhaowei
Jia, Qiang
author_sort Zhang, Ruyi
collection PubMed
description BACKGROUND: The digital cadmium–zinc–telluride (CZT)-based SPECT system has many advantages, including better spatial and energy resolution. However, the impacts of different acquisition and reconstruction parameters on CZT SPECT quantification might still need to be validated. This study aimed to evaluate the impacts of acquisition parameters (the main energy window and acquisition time per frame) and reconstruction parameters (the number of iterations, subsets in iterative reconstruction, post-filter, and image correction methods) on the technetium quantification of CZT SPECT/CT. METHODS: A phantom (PET NEMA/IEC image quality, USA) was filled with four target-to-background (T/B) ratios (32:1, 16:1, 8:1, and 4:1) of technetium. Mean uptake values (the calculated mean concentrations for spheres) were measured to evaluate the recovery coefficient (RC) changes under different acquisition and reconstruction parameters. The corresponding standard deviations of mean uptake values were also measured to evaluate the quantification error. Image quality was evaluated using the National Electrical Manufacturers Association (NEMA) NU 2–2012 standard. RESULTS: For all T/B ratios, significant correlations were found between iterations and RCs (r = 0.62–0.96 for 1–35 iterations, r = 0.94–0.99 for 35–90 iterations) as well as between the full width at half maximum (FWHM) of the Gaussian filter and RCs (r = − 0.86 to − 1.00, all P values < 0.05). The regression coefficients of 1–35 iterations were higher than those of 35–90 iterations (0.51–1.60 vs. 0.02–0.19). RCs calculated with AC (attenuation correction) + SC (scatter correction) + RR (resolution recovery correction) combination were more accurate (53.82–106.70%) than those calculated with other combinations (all P values < 0.05). No significant statistical differences (all P values > 0.05) were found between the 15% and 20% energy windows except for the 32:1 T/B ratio (P value = 0.023) or between the 10 s/frame and 120 s/frame acquisition times except for the 4:1 T/B ratio (P value = 0.015) in terms of RCs. CONCLUSIONS: CZT-SPECT/CT of technetium resulted in good quantification accuracy. The favourable acquisition parameters might be a 15% energy window and 40 s/frame of acquisition time. The favourable reconstruction parameters might be 35 iterations, 20 subsets, the AC + SC + RR correction combination, and no filter. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s40658-021-00412-4.
format Online
Article
Text
id pubmed-8473509
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-84735092021-10-08 Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study Zhang, Ruyi Wang, Miao Zhou, Yaqian Wang, Shen Shen, Yiming Li, Ning Wang, Peng Tan, Jian Meng, Zhaowei Jia, Qiang EJNMMI Phys Original Research BACKGROUND: The digital cadmium–zinc–telluride (CZT)-based SPECT system has many advantages, including better spatial and energy resolution. However, the impacts of different acquisition and reconstruction parameters on CZT SPECT quantification might still need to be validated. This study aimed to evaluate the impacts of acquisition parameters (the main energy window and acquisition time per frame) and reconstruction parameters (the number of iterations, subsets in iterative reconstruction, post-filter, and image correction methods) on the technetium quantification of CZT SPECT/CT. METHODS: A phantom (PET NEMA/IEC image quality, USA) was filled with four target-to-background (T/B) ratios (32:1, 16:1, 8:1, and 4:1) of technetium. Mean uptake values (the calculated mean concentrations for spheres) were measured to evaluate the recovery coefficient (RC) changes under different acquisition and reconstruction parameters. The corresponding standard deviations of mean uptake values were also measured to evaluate the quantification error. Image quality was evaluated using the National Electrical Manufacturers Association (NEMA) NU 2–2012 standard. RESULTS: For all T/B ratios, significant correlations were found between iterations and RCs (r = 0.62–0.96 for 1–35 iterations, r = 0.94–0.99 for 35–90 iterations) as well as between the full width at half maximum (FWHM) of the Gaussian filter and RCs (r = − 0.86 to − 1.00, all P values < 0.05). The regression coefficients of 1–35 iterations were higher than those of 35–90 iterations (0.51–1.60 vs. 0.02–0.19). RCs calculated with AC (attenuation correction) + SC (scatter correction) + RR (resolution recovery correction) combination were more accurate (53.82–106.70%) than those calculated with other combinations (all P values < 0.05). No significant statistical differences (all P values > 0.05) were found between the 15% and 20% energy windows except for the 32:1 T/B ratio (P value = 0.023) or between the 10 s/frame and 120 s/frame acquisition times except for the 4:1 T/B ratio (P value = 0.015) in terms of RCs. CONCLUSIONS: CZT-SPECT/CT of technetium resulted in good quantification accuracy. The favourable acquisition parameters might be a 15% energy window and 40 s/frame of acquisition time. The favourable reconstruction parameters might be 35 iterations, 20 subsets, the AC + SC + RR correction combination, and no filter. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s40658-021-00412-4. Springer International Publishing 2021-09-26 /pmc/articles/PMC8473509/ /pubmed/34568990 http://dx.doi.org/10.1186/s40658-021-00412-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Zhang, Ruyi
Wang, Miao
Zhou, Yaqian
Wang, Shen
Shen, Yiming
Li, Ning
Wang, Peng
Tan, Jian
Meng, Zhaowei
Jia, Qiang
Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title_full Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title_fullStr Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title_full_unstemmed Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title_short Impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based SPECT/CT system: a phantom study
title_sort impacts of acquisition and reconstruction parameters on the absolute technetium quantification of the cadmium–zinc–telluride-based spect/ct system: a phantom study
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8473509/
https://www.ncbi.nlm.nih.gov/pubmed/34568990
http://dx.doi.org/10.1186/s40658-021-00412-4
work_keys_str_mv AT zhangruyi impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT wangmiao impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT zhouyaqian impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT wangshen impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT shenyiming impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT lining impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT wangpeng impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT tanjian impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT mengzhaowei impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy
AT jiaqiang impactsofacquisitionandreconstructionparametersontheabsolutetechnetiumquantificationofthecadmiumzinctelluridebasedspectctsystemaphantomstudy