Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies

The aim of the study was to estimate the accuracy of standardized uptake values of (18)F-fluorodeoxyglucose ((18)F-FDG) in lung lesions during positron emission tomography combined with computed tomography (PET/CT) imaging, based on phantom studies performed for different PET/CT scanners. MATERIALS...

Descripción completa

Detalles Bibliográficos
Autores principales: Tlostanova, M.S., Chipiga, L.A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Privolzhsky Research Medical University 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8482825/
https://www.ncbi.nlm.nih.gov/pubmed/34603751
http://dx.doi.org/10.17691/stm2021.13.3.02
_version_ 1784576990518444032
author Tlostanova, M.S.
Chipiga, L.A.
author_facet Tlostanova, M.S.
Chipiga, L.A.
author_sort Tlostanova, M.S.
collection PubMed
description The aim of the study was to estimate the accuracy of standardized uptake values of (18)F-fluorodeoxyglucose ((18)F-FDG) in lung lesions during positron emission tomography combined with computed tomography (PET/CT) imaging, based on phantom studies performed for different PET/CT scanners. MATERIALS AND METHODS: The analysis of the PET/CT with (18)F-FDG data was performed for 86 patients newly diagnosed with the lung lesions: malignant tumors (n=37), benign tumors and inflammatory diseases (n=49). The criteria for inclusion in the study were developed considering the recommendations of the Fleischner Society (2017). The characteristics of the lesions on CT met the following requirements: a round shape or close to it; total size of 8 to 30 mm; solid or subsolid structure (with the exception of lesion with ground-glass opacity); a solid part size of ≥8 mm. All the patients had no signs of pleurisy, lymphadenopathy, or cancer history. PET/CT imaging with (18)F-FDG was performed with three scanners: Discovery 690 (General Electric, USA), Biograph mCT 128 (Siemens, Germany), and Biograph mCT 40 (Siemens); the preparation of patients prior to the scan was standardized. To determine the reference accumulation of a radiopharmaceutical in the pathological lesion, four scans of a specialized NEMA IEC PET Body Phantom Set (USA) were performed for each scanner. For each unit, the recovery coefficients (RCs) of radioactivity, maximum and recovered (corrected) standardized uptake values (SUVs) were determined. Statistical relationship between the size of lesions, SUV(max) and SUV(correct) was evaluated. Data processing was performed using MedCalc v. 19.2.0 software. RESULTS: During the phantom study, the underestimation of the radioactivity was determined in the spheres with the diameters of 10 and 13 mm, overestimation was observed in the sphere with the diameter of 28 mm. Both underestimation and overestimation of radioactivity were determined for the spheres with a diameter of 17 and 22 mm. SUV(max) differed from the reference values for 85 patients (98.8%). The underestimation of these values was found for 63 patients (73.2%) due to the partial volume effect. The greatest underestimation was observed for the patients with 8 mm diameter lesions. Depending on the scanner, the underestimation of the SUV(max) in these patients reached up to 54–73%. For 9 patients (25%) with malignant tumors of 9–12 mm, the utility of RC made it possible to avoid false negative results. For the lesions with a diameter of 30 mm, an overestimation of SUV(max) up to 22% was determined due to the negative influence of the reconstruction algorithms. CONCLUSION: The use of RC eliminates the influence of the partial volume effect and reconstruction methods on the accuracy of estimating the SUV(max) in lung lesions, which ensures reproducibility, increase in the information content of the method, as well as the comparability of the results of PET/CT with (18)F-FDG obtained on the different models of PET/CT units with different technological characteristics.
format Online
Article
Text
id pubmed-8482825
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Privolzhsky Research Medical University
record_format MEDLINE/PubMed
spelling pubmed-84828252021-10-01 Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies Tlostanova, M.S. Chipiga, L.A. Sovrem Tekhnologii Med Advanced Researches The aim of the study was to estimate the accuracy of standardized uptake values of (18)F-fluorodeoxyglucose ((18)F-FDG) in lung lesions during positron emission tomography combined with computed tomography (PET/CT) imaging, based on phantom studies performed for different PET/CT scanners. MATERIALS AND METHODS: The analysis of the PET/CT with (18)F-FDG data was performed for 86 patients newly diagnosed with the lung lesions: malignant tumors (n=37), benign tumors and inflammatory diseases (n=49). The criteria for inclusion in the study were developed considering the recommendations of the Fleischner Society (2017). The characteristics of the lesions on CT met the following requirements: a round shape or close to it; total size of 8 to 30 mm; solid or subsolid structure (with the exception of lesion with ground-glass opacity); a solid part size of ≥8 mm. All the patients had no signs of pleurisy, lymphadenopathy, or cancer history. PET/CT imaging with (18)F-FDG was performed with three scanners: Discovery 690 (General Electric, USA), Biograph mCT 128 (Siemens, Germany), and Biograph mCT 40 (Siemens); the preparation of patients prior to the scan was standardized. To determine the reference accumulation of a radiopharmaceutical in the pathological lesion, four scans of a specialized NEMA IEC PET Body Phantom Set (USA) were performed for each scanner. For each unit, the recovery coefficients (RCs) of radioactivity, maximum and recovered (corrected) standardized uptake values (SUVs) were determined. Statistical relationship between the size of lesions, SUV(max) and SUV(correct) was evaluated. Data processing was performed using MedCalc v. 19.2.0 software. RESULTS: During the phantom study, the underestimation of the radioactivity was determined in the spheres with the diameters of 10 and 13 mm, overestimation was observed in the sphere with the diameter of 28 mm. Both underestimation and overestimation of radioactivity were determined for the spheres with a diameter of 17 and 22 mm. SUV(max) differed from the reference values for 85 patients (98.8%). The underestimation of these values was found for 63 patients (73.2%) due to the partial volume effect. The greatest underestimation was observed for the patients with 8 mm diameter lesions. Depending on the scanner, the underestimation of the SUV(max) in these patients reached up to 54–73%. For 9 patients (25%) with malignant tumors of 9–12 mm, the utility of RC made it possible to avoid false negative results. For the lesions with a diameter of 30 mm, an overestimation of SUV(max) up to 22% was determined due to the negative influence of the reconstruction algorithms. CONCLUSION: The use of RC eliminates the influence of the partial volume effect and reconstruction methods on the accuracy of estimating the SUV(max) in lung lesions, which ensures reproducibility, increase in the information content of the method, as well as the comparability of the results of PET/CT with (18)F-FDG obtained on the different models of PET/CT units with different technological characteristics. Privolzhsky Research Medical University 2021 2021-06-28 /pmc/articles/PMC8482825/ /pubmed/34603751 http://dx.doi.org/10.17691/stm2021.13.3.02 Text en https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Advanced Researches
Tlostanova, M.S.
Chipiga, L.A.
Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title_full Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title_fullStr Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title_full_unstemmed Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title_short Evaluation of the Accuracy of Standardized Uptake Values of (18)F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies
title_sort evaluation of the accuracy of standardized uptake values of (18)f-fluorodeoxyglucose in lung lesions based on phantom studies
topic Advanced Researches
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8482825/
https://www.ncbi.nlm.nih.gov/pubmed/34603751
http://dx.doi.org/10.17691/stm2021.13.3.02
work_keys_str_mv AT tlostanovams evaluationoftheaccuracyofstandardizeduptakevaluesof18ffluorodeoxyglucoseinlunglesionsbasedonphantomstudies
AT chipigala evaluationoftheaccuracyofstandardizeduptakevaluesof18ffluorodeoxyglucoseinlunglesionsbasedonphantomstudies