A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices

Purpose Variation among aqueous humor outflow from venting slits performed on glaucoma drainage device tubing often occurs even when physician technique and equipment are held constant. Our hypothesis is that there are dimensional differences within the tubing, even among the same make and model of...

Descripción completa

Detalles Bibliográficos
Autores principales: Terrell, Jesse A, Blake, Charles Richard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8494156/
https://www.ncbi.nlm.nih.gov/pubmed/34659982
http://dx.doi.org/10.7759/cureus.17771
_version_ 1784579251555532800
author Terrell, Jesse A
Blake, Charles Richard
author_facet Terrell, Jesse A
Blake, Charles Richard
author_sort Terrell, Jesse A
collection PubMed
description Purpose Variation among aqueous humor outflow from venting slits performed on glaucoma drainage device tubing often occurs even when physician technique and equipment are held constant. Our hypothesis is that there are dimensional differences within the tubing, even among the same make and model of glaucoma drainage device (GDD) implants. Methods Prior to surgical implantation, excess glaucoma drainage tubing was collected for analysis. The tubing samples were sliced horizontally, and the external tube, internal lumen, and wall dimension measurements were collected microscopically. Groups were divided based upon brand and model and then statistically analyzed using an independent t-test. A total of 28 tubes were analyzed, consisting of 7 Molteno and 21 Baerveldt implants. Results The mean external diameter for the Molteno group was 656 ± 20µm, significantly larger than the Baerveldt external diameter of 620 ± 13µm (P<0.05). The mean internal diameter among Molteno lumens was 344 ± 13µm, also statistically larger than the mean internal diameter of 309 ± 18µm for Baerveldt tubes (P<0.05). The Molteno luminal wall width varied significantly less than the Baerveldt wall, 18% versus 28%, respectively (P<0.05). The tubings’ wall widths variation translated into highly significant off-centered lumens among both brands. Conclusion Our findings suggest that there are significant variations among glaucoma implant dimensions between and within the multiple makes and models. The discrepancies among tubal wall thickness and off-centered lumens are undetectable to the naked eye. Importantly, this may result in significant aqueous humor outflow variation following the creation of venting slits secondary to the found irregular luminal diameters and tube wall thicknesses.
format Online
Article
Text
id pubmed-8494156
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-84941562021-10-14 A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices Terrell, Jesse A Blake, Charles Richard Cureus Ophthalmology Purpose Variation among aqueous humor outflow from venting slits performed on glaucoma drainage device tubing often occurs even when physician technique and equipment are held constant. Our hypothesis is that there are dimensional differences within the tubing, even among the same make and model of glaucoma drainage device (GDD) implants. Methods Prior to surgical implantation, excess glaucoma drainage tubing was collected for analysis. The tubing samples were sliced horizontally, and the external tube, internal lumen, and wall dimension measurements were collected microscopically. Groups were divided based upon brand and model and then statistically analyzed using an independent t-test. A total of 28 tubes were analyzed, consisting of 7 Molteno and 21 Baerveldt implants. Results The mean external diameter for the Molteno group was 656 ± 20µm, significantly larger than the Baerveldt external diameter of 620 ± 13µm (P<0.05). The mean internal diameter among Molteno lumens was 344 ± 13µm, also statistically larger than the mean internal diameter of 309 ± 18µm for Baerveldt tubes (P<0.05). The Molteno luminal wall width varied significantly less than the Baerveldt wall, 18% versus 28%, respectively (P<0.05). The tubings’ wall widths variation translated into highly significant off-centered lumens among both brands. Conclusion Our findings suggest that there are significant variations among glaucoma implant dimensions between and within the multiple makes and models. The discrepancies among tubal wall thickness and off-centered lumens are undetectable to the naked eye. Importantly, this may result in significant aqueous humor outflow variation following the creation of venting slits secondary to the found irregular luminal diameters and tube wall thicknesses. Cureus 2021-09-06 /pmc/articles/PMC8494156/ /pubmed/34659982 http://dx.doi.org/10.7759/cureus.17771 Text en Copyright © 2021, Terrell et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Ophthalmology
Terrell, Jesse A
Blake, Charles Richard
A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title_full A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title_fullStr A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title_full_unstemmed A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title_short A Novel Study Quantifying Intrinsic Dimensional Variation Among Glaucoma Drainage Devices
title_sort novel study quantifying intrinsic dimensional variation among glaucoma drainage devices
topic Ophthalmology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8494156/
https://www.ncbi.nlm.nih.gov/pubmed/34659982
http://dx.doi.org/10.7759/cureus.17771
work_keys_str_mv AT terrelljessea anovelstudyquantifyingintrinsicdimensionalvariationamongglaucomadrainagedevices
AT blakecharlesrichard anovelstudyquantifyingintrinsicdimensionalvariationamongglaucomadrainagedevices
AT terrelljessea novelstudyquantifyingintrinsicdimensionalvariationamongglaucomadrainagedevices
AT blakecharlesrichard novelstudyquantifyingintrinsicdimensionalvariationamongglaucomadrainagedevices