Cargando…

Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial

IMPORTANCE: Data regarding upper extremity midline catheter (MC)–related thrombosis (CRT) are sparse, with some evidence indicating that MCs have a high rate of CRT. OBJECTIVE: To compare 2 MCs with differing antithrombogenic mechanisms for this outcome. DESIGN, SETTING, AND PARTICIPANTS: In this pa...

Descripción completa

Detalles Bibliográficos
Autores principales: Bahl, Amit, Diloreto, Emily, Jankowski, David, Hijazi, Mahmoud, Chen, Nai-Wei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Medical Association 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8495531/
https://www.ncbi.nlm.nih.gov/pubmed/34613402
http://dx.doi.org/10.1001/jamanetworkopen.2021.27836
_version_ 1784579574022012928
author Bahl, Amit
Diloreto, Emily
Jankowski, David
Hijazi, Mahmoud
Chen, Nai-Wei
author_facet Bahl, Amit
Diloreto, Emily
Jankowski, David
Hijazi, Mahmoud
Chen, Nai-Wei
author_sort Bahl, Amit
collection PubMed
description IMPORTANCE: Data regarding upper extremity midline catheter (MC)–related thrombosis (CRT) are sparse, with some evidence indicating that MCs have a high rate of CRT. OBJECTIVE: To compare 2 MCs with differing antithrombogenic mechanisms for this outcome. DESIGN, SETTING, AND PARTICIPANTS: In this parallel, 2-arm randomized clinical trial, 496 adult patients hospitalized at a tertiary care suburban academic medical center who received an MC were assessed for eligibility between January 1, 2019, and October 31, 2020, and 212 were randomized. INTERVENTIONS: Inpatients were randomized to receive a 4F antithrombotic MC (MC-AT) or a 4.5F antithrombotic and antimicrobial MC (MC-AT-AM). MAIN OUTCOMES AND MEASURES: The primary outcome was symptomatic midline CRT inclusive of deep vein thrombosis or superficial venous thrombophlebitis within 30 days after insertion. Secondary outcomes included catheter-associated bloodstream infection and catheter failure. RESULTS: A total of 191 patients (mean [SD] age, 60.2 [16.7] years; 114 [59.7%] female) were included in the final analysis: 94 patients in the MC-AT group and 97 in the MC-AT-AM group. Symptomatic midline CRT occurred in 7 patients (7.5%) in the MC-AT group and 11 (11.3%) in the MC-AT-AM group (P = .46). Deep vein thrombosis occurred in 5 patients (5.3%) in the MC-AT group and 5 patients (5.2%) in the MC-AT-AM group (P > .99). Pulmonary embolism occurred in 1 patient in the MC-AT group. No catheter-associated bloodstream infection occurred in either group. Premature catheter failure occurred in 22 patients (23.4%) in the MC-AT group and 20 (20.6%) in the MC-AT-AM group (P = .64). In Cox proportional hazards regression analysis, no statistically significant difference was found between groups for the risk of catheter failure (hazard ratio, 1.27; 95% CI, 0.67-2.43; P = .46). CONCLUSIONS AND RELEVANCE: No difference was found in thrombosis in MCs with 2 distinct antithrombogenic mechanisms; however, the risk of CRT in both groups was high. Practitioners should strongly consider the safety risks associated with MCs when determining the appropriate vascular access device. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT03725293
format Online
Article
Text
id pubmed-8495531
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher American Medical Association
record_format MEDLINE/PubMed
spelling pubmed-84955312021-10-19 Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial Bahl, Amit Diloreto, Emily Jankowski, David Hijazi, Mahmoud Chen, Nai-Wei JAMA Netw Open Original Investigation IMPORTANCE: Data regarding upper extremity midline catheter (MC)–related thrombosis (CRT) are sparse, with some evidence indicating that MCs have a high rate of CRT. OBJECTIVE: To compare 2 MCs with differing antithrombogenic mechanisms for this outcome. DESIGN, SETTING, AND PARTICIPANTS: In this parallel, 2-arm randomized clinical trial, 496 adult patients hospitalized at a tertiary care suburban academic medical center who received an MC were assessed for eligibility between January 1, 2019, and October 31, 2020, and 212 were randomized. INTERVENTIONS: Inpatients were randomized to receive a 4F antithrombotic MC (MC-AT) or a 4.5F antithrombotic and antimicrobial MC (MC-AT-AM). MAIN OUTCOMES AND MEASURES: The primary outcome was symptomatic midline CRT inclusive of deep vein thrombosis or superficial venous thrombophlebitis within 30 days after insertion. Secondary outcomes included catheter-associated bloodstream infection and catheter failure. RESULTS: A total of 191 patients (mean [SD] age, 60.2 [16.7] years; 114 [59.7%] female) were included in the final analysis: 94 patients in the MC-AT group and 97 in the MC-AT-AM group. Symptomatic midline CRT occurred in 7 patients (7.5%) in the MC-AT group and 11 (11.3%) in the MC-AT-AM group (P = .46). Deep vein thrombosis occurred in 5 patients (5.3%) in the MC-AT group and 5 patients (5.2%) in the MC-AT-AM group (P > .99). Pulmonary embolism occurred in 1 patient in the MC-AT group. No catheter-associated bloodstream infection occurred in either group. Premature catheter failure occurred in 22 patients (23.4%) in the MC-AT group and 20 (20.6%) in the MC-AT-AM group (P = .64). In Cox proportional hazards regression analysis, no statistically significant difference was found between groups for the risk of catheter failure (hazard ratio, 1.27; 95% CI, 0.67-2.43; P = .46). CONCLUSIONS AND RELEVANCE: No difference was found in thrombosis in MCs with 2 distinct antithrombogenic mechanisms; however, the risk of CRT in both groups was high. Practitioners should strongly consider the safety risks associated with MCs when determining the appropriate vascular access device. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT03725293 American Medical Association 2021-10-06 /pmc/articles/PMC8495531/ /pubmed/34613402 http://dx.doi.org/10.1001/jamanetworkopen.2021.27836 Text en Copyright 2021 Bahl A et al. JAMA Network Open. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article distributed under the terms of the CC-BY-NC-ND License.
spellingShingle Original Investigation
Bahl, Amit
Diloreto, Emily
Jankowski, David
Hijazi, Mahmoud
Chen, Nai-Wei
Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title_full Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title_fullStr Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title_full_unstemmed Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title_short Comparison of 2 Midline Catheter Devices With Differing Antithrombogenic Mechanisms for Catheter-Related Thrombosis: A Randomized Clinical Trial
title_sort comparison of 2 midline catheter devices with differing antithrombogenic mechanisms for catheter-related thrombosis: a randomized clinical trial
topic Original Investigation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8495531/
https://www.ncbi.nlm.nih.gov/pubmed/34613402
http://dx.doi.org/10.1001/jamanetworkopen.2021.27836
work_keys_str_mv AT bahlamit comparisonof2midlinecatheterdeviceswithdifferingantithrombogenicmechanismsforcatheterrelatedthrombosisarandomizedclinicaltrial
AT diloretoemily comparisonof2midlinecatheterdeviceswithdifferingantithrombogenicmechanismsforcatheterrelatedthrombosisarandomizedclinicaltrial
AT jankowskidavid comparisonof2midlinecatheterdeviceswithdifferingantithrombogenicmechanismsforcatheterrelatedthrombosisarandomizedclinicaltrial
AT hijazimahmoud comparisonof2midlinecatheterdeviceswithdifferingantithrombogenicmechanismsforcatheterrelatedthrombosisarandomizedclinicaltrial
AT chennaiwei comparisonof2midlinecatheterdeviceswithdifferingantithrombogenicmechanismsforcatheterrelatedthrombosisarandomizedclinicaltrial