Within-session variability as quality control for oscillometry in health and disease
Oscillometry is increasingly adopted in respiratory clinics, but many recommendations regarding measurement settings and quality control remain subjective. The aim of this study was to investigate the optimal number of measurements and acceptable within-session coefficient of variation (CoV) in heal...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
European Respiratory Society
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8502940/ https://www.ncbi.nlm.nih.gov/pubmed/34651039 http://dx.doi.org/10.1183/23120541.00074-2021 |
_version_ | 1784581002143727616 |
---|---|
author | Harkness, Louise M. Patel, Kieran Sanai, Farid Rutting, Sandra Cottee, Alice M. Farah, Claude S. Schoeffel, Robin E. King, Gregory G. Thamrin, Cindy |
author_facet | Harkness, Louise M. Patel, Kieran Sanai, Farid Rutting, Sandra Cottee, Alice M. Farah, Claude S. Schoeffel, Robin E. King, Gregory G. Thamrin, Cindy |
author_sort | Harkness, Louise M. |
collection | PubMed |
description | Oscillometry is increasingly adopted in respiratory clinics, but many recommendations regarding measurement settings and quality control remain subjective. The aim of this study was to investigate the optimal number of measurements and acceptable within-session coefficient of variation (CoV) in health, asthma and COPD. 15 healthy, 15 asthma and 15 COPD adult participants were recruited. Eight consecutive 30-s measurements were made using an oscillometry device, from which resistance at 5 Hz (R(rs(5))) was examined. The effect of progressively including a greater number of measurements on R(rs(5)) and its within-session CoV was investigated. Data were analysed using one-way repeated-measures ANOVA with Bonferroni post hoc test. The CoV(R(rs(5))) of the first three measurements was 6.7±4.7%, 9.7±5.7% and 12.6±11.2% in healthy, asthma and COPD participants, respectively. Both mean R(rs(5)) and CoV(R(rs(5))) were not statistically different when progressively including four to eight measurements. Selecting the three closest R(rs(5)) values over an increasing number of measurements progressively decreased the CoV(R(rs(5))). In order for ≥95% of participants to fall within a target CoV(R(rs(5))) of 10%, four or more, five and six measurements were needed in health, asthma and COPD, respectively. Within-session variability of oscillometry is increased in disease. Furthermore, the higher number of measurements required to achieve a set target for asthma and COPD patients may not be practical in a clinical setting. Provided technical acceptability of measurements is established, i.e. by removing artefacts and outliers, then a CoV of 10% is a marker of quality in most patients, but we suggest higher CoVs up to 15–20% should still be reportable. |
format | Online Article Text |
id | pubmed-8502940 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | European Respiratory Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-85029402021-10-13 Within-session variability as quality control for oscillometry in health and disease Harkness, Louise M. Patel, Kieran Sanai, Farid Rutting, Sandra Cottee, Alice M. Farah, Claude S. Schoeffel, Robin E. King, Gregory G. Thamrin, Cindy ERJ Open Res Original Research Articles Oscillometry is increasingly adopted in respiratory clinics, but many recommendations regarding measurement settings and quality control remain subjective. The aim of this study was to investigate the optimal number of measurements and acceptable within-session coefficient of variation (CoV) in health, asthma and COPD. 15 healthy, 15 asthma and 15 COPD adult participants were recruited. Eight consecutive 30-s measurements were made using an oscillometry device, from which resistance at 5 Hz (R(rs(5))) was examined. The effect of progressively including a greater number of measurements on R(rs(5)) and its within-session CoV was investigated. Data were analysed using one-way repeated-measures ANOVA with Bonferroni post hoc test. The CoV(R(rs(5))) of the first three measurements was 6.7±4.7%, 9.7±5.7% and 12.6±11.2% in healthy, asthma and COPD participants, respectively. Both mean R(rs(5)) and CoV(R(rs(5))) were not statistically different when progressively including four to eight measurements. Selecting the three closest R(rs(5)) values over an increasing number of measurements progressively decreased the CoV(R(rs(5))). In order for ≥95% of participants to fall within a target CoV(R(rs(5))) of 10%, four or more, five and six measurements were needed in health, asthma and COPD, respectively. Within-session variability of oscillometry is increased in disease. Furthermore, the higher number of measurements required to achieve a set target for asthma and COPD patients may not be practical in a clinical setting. Provided technical acceptability of measurements is established, i.e. by removing artefacts and outliers, then a CoV of 10% is a marker of quality in most patients, but we suggest higher CoVs up to 15–20% should still be reportable. European Respiratory Society 2021-10-11 /pmc/articles/PMC8502940/ /pubmed/34651039 http://dx.doi.org/10.1183/23120541.00074-2021 Text en Copyright ©The authors 2021 https://creativecommons.org/licenses/by-nc/4.0/This version is distributed under the terms of the Creative Commons Attribution Non-Commercial Licence 4.0. For commercial reproduction rights and permissions contact permissions@ersnet.org (mailto:permissions@ersnet.org) |
spellingShingle | Original Research Articles Harkness, Louise M. Patel, Kieran Sanai, Farid Rutting, Sandra Cottee, Alice M. Farah, Claude S. Schoeffel, Robin E. King, Gregory G. Thamrin, Cindy Within-session variability as quality control for oscillometry in health and disease |
title | Within-session variability as quality control for oscillometry in health and disease |
title_full | Within-session variability as quality control for oscillometry in health and disease |
title_fullStr | Within-session variability as quality control for oscillometry in health and disease |
title_full_unstemmed | Within-session variability as quality control for oscillometry in health and disease |
title_short | Within-session variability as quality control for oscillometry in health and disease |
title_sort | within-session variability as quality control for oscillometry in health and disease |
topic | Original Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8502940/ https://www.ncbi.nlm.nih.gov/pubmed/34651039 http://dx.doi.org/10.1183/23120541.00074-2021 |
work_keys_str_mv | AT harknesslouisem withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT patelkieran withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT sanaifarid withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT ruttingsandra withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT cotteealicem withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT farahclaudes withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT schoeffelrobine withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT kinggregoryg withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease AT thamrincindy withinsessionvariabilityasqualitycontrolforoscillometryinhealthanddisease |