Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010

Background: Electronic devices for measuring blood pressure (BP) need to go through independent clinical validation as recommended by different authorities, both in general and specific populations. The aim of this study was to assess the validity of the Omron RS6 (HEM-6221-E) wrist oscillometric de...

Descripción completa

Detalles Bibliográficos
Autores principales: Ali, Elrazi A., Omar, Saeed M., Ibrahim, Yassin, Al-Wutayd, Osama, Adam, Ishag
Formato: Online Artículo Texto
Lenguaje:English
Publicado: F1000 Research Limited 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8503788/
https://www.ncbi.nlm.nih.gov/pubmed/34691394
http://dx.doi.org/10.12688/f1000research.26442.1
_version_ 1784581201962467328
author Ali, Elrazi A.
Omar, Saeed M.
Ibrahim, Yassin
Al-Wutayd, Osama
Adam, Ishag
author_facet Ali, Elrazi A.
Omar, Saeed M.
Ibrahim, Yassin
Al-Wutayd, Osama
Adam, Ishag
author_sort Ali, Elrazi A.
collection PubMed
description Background: Electronic devices for measuring blood pressure (BP) need to go through independent clinical validation as recommended by different authorities, both in general and specific populations. The aim of this study was to assess the validity of the Omron RS6 (HEM-6221-E) wrist oscillometric devices in obese Sudanese patients. Methods: Of 90 obese individuals invited for recruitment, 33 were included in the study, and had their BP at the level of the wrist measured using Omron RS6 and standard mercury sphygmomanometer. Two observations were made and the mean was taken. BP differences between the two methods for the 33 participants were classified into three categories (≤5, ≤10, and ≤15 mmHg), according to the European Society of Hypertension-International Protocol revision 2010 (ESH-IP2) criteria. This was then used to assess the validity of the tested Omron RS6 device. Results: Participants had a mean age of 56.97 years (standard deviation (SD), 8.75; range, 36-79). Average systolic blood pressure (SBP) was 146.21 mmHg (SD, 23.07; range, 107-182), and average diastolic blood pressure (DBP) was 93.82 mmHg (SD, 16.06; range, 67-128). There was a good agreement between the two observations using the OMRON RS6 and the standard sphygmomanometer: −4 to + 3 mmHg for SBP and −4 to +4 mmHg for DBP, with the mean difference of 1.73±1.11 mmHg for SBP and 1.49±1.02 mmHg for DBP. Conclusion: Thus, the Omron RS6 (HEM-6221-E) is a valid and suitable measure of BP according to ESH-IP2.
format Online
Article
Text
id pubmed-8503788
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher F1000 Research Limited
record_format MEDLINE/PubMed
spelling pubmed-85037882021-10-22 Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010 Ali, Elrazi A. Omar, Saeed M. Ibrahim, Yassin Al-Wutayd, Osama Adam, Ishag F1000Res Research Article Background: Electronic devices for measuring blood pressure (BP) need to go through independent clinical validation as recommended by different authorities, both in general and specific populations. The aim of this study was to assess the validity of the Omron RS6 (HEM-6221-E) wrist oscillometric devices in obese Sudanese patients. Methods: Of 90 obese individuals invited for recruitment, 33 were included in the study, and had their BP at the level of the wrist measured using Omron RS6 and standard mercury sphygmomanometer. Two observations were made and the mean was taken. BP differences between the two methods for the 33 participants were classified into three categories (≤5, ≤10, and ≤15 mmHg), according to the European Society of Hypertension-International Protocol revision 2010 (ESH-IP2) criteria. This was then used to assess the validity of the tested Omron RS6 device. Results: Participants had a mean age of 56.97 years (standard deviation (SD), 8.75; range, 36-79). Average systolic blood pressure (SBP) was 146.21 mmHg (SD, 23.07; range, 107-182), and average diastolic blood pressure (DBP) was 93.82 mmHg (SD, 16.06; range, 67-128). There was a good agreement between the two observations using the OMRON RS6 and the standard sphygmomanometer: −4 to + 3 mmHg for SBP and −4 to +4 mmHg for DBP, with the mean difference of 1.73±1.11 mmHg for SBP and 1.49±1.02 mmHg for DBP. Conclusion: Thus, the Omron RS6 (HEM-6221-E) is a valid and suitable measure of BP according to ESH-IP2. F1000 Research Limited 2020-10-30 /pmc/articles/PMC8503788/ /pubmed/34691394 http://dx.doi.org/10.12688/f1000research.26442.1 Text en Copyright: © 2020 Ali EA et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Ali, Elrazi A.
Omar, Saeed M.
Ibrahim, Yassin
Al-Wutayd, Osama
Adam, Ishag
Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title_full Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title_fullStr Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title_full_unstemmed Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title_short Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010
title_sort validation of the wrist blood pressure measuring device omron rs6 (hem-6221-e) among obese sudanese patients according to the european society of hypertension international protocol revision 2010
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8503788/
https://www.ncbi.nlm.nih.gov/pubmed/34691394
http://dx.doi.org/10.12688/f1000research.26442.1
work_keys_str_mv AT alielrazia validationofthewristbloodpressuremeasuringdeviceomronrs6hem6221eamongobesesudanesepatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT omarsaeedm validationofthewristbloodpressuremeasuringdeviceomronrs6hem6221eamongobesesudanesepatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT ibrahimyassin validationofthewristbloodpressuremeasuringdeviceomronrs6hem6221eamongobesesudanesepatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT alwutaydosama validationofthewristbloodpressuremeasuringdeviceomronrs6hem6221eamongobesesudanesepatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT adamishag validationofthewristbloodpressuremeasuringdeviceomronrs6hem6221eamongobesesudanesepatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocolrevision2010