This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four‐field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, an...

Descripción completa

Detalles Bibliográficos
Autores principales: Rojas, Diana M. C., Pavoni, Juliana F., Arruda, Gustavo V., Baffa, Oswaldo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8504585/
https://www.ncbi.nlm.nih.gov/pubmed/34436819
http://dx.doi.org/10.1002/acm2.13403
Descripción
Sumario:This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four‐field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, and thermoluminescent dosimeters (TLD) were used for point‐dose measurements outside it. Both results were compared to the treatment planning system (TPS) dose calculation without using heterogeneity corrections to evaluate the influence of the metal in the dose distribution. MAGIC‐f gel dosimeter (Methacrylic and Ascorbic acid in Gelatin Initiated by Copper with Formaldehyde) associated with magnetic resonance imaging was used. TLD were positioned at several points at the bone metal interface and the sacrum region. The comparison of the gel measured and the TPS calculated dose distributions were done using gamma analysis (3%/3 mm), and a pass rate of 93% was achieved. The TLD dose values at the bone‐metal interface showed variations from the planned dose. However, at the sacrum region, where the beams did not intercept the prosthesis, there was a good agreement between TPS planning and TLD measurements. Our results show how the combination of 3D dosimetry and measurements at specific points in the phantom allowed a comprehensive view of the dose distribution and identified that care must also be paid to regions outside the PTV.