Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis
This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four‐field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, an...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8504585/ https://www.ncbi.nlm.nih.gov/pubmed/34436819 http://dx.doi.org/10.1002/acm2.13403 |
_version_ | 1784581348298588160 |
---|---|
author | Rojas, Diana M. C. Pavoni, Juliana F. Arruda, Gustavo V. Baffa, Oswaldo |
author_facet | Rojas, Diana M. C. Pavoni, Juliana F. Arruda, Gustavo V. Baffa, Oswaldo |
author_sort | Rojas, Diana M. C. |
collection | PubMed |
description | This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four‐field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, and thermoluminescent dosimeters (TLD) were used for point‐dose measurements outside it. Both results were compared to the treatment planning system (TPS) dose calculation without using heterogeneity corrections to evaluate the influence of the metal in the dose distribution. MAGIC‐f gel dosimeter (Methacrylic and Ascorbic acid in Gelatin Initiated by Copper with Formaldehyde) associated with magnetic resonance imaging was used. TLD were positioned at several points at the bone metal interface and the sacrum region. The comparison of the gel measured and the TPS calculated dose distributions were done using gamma analysis (3%/3 mm), and a pass rate of 93% was achieved. The TLD dose values at the bone‐metal interface showed variations from the planned dose. However, at the sacrum region, where the beams did not intercept the prosthesis, there was a good agreement between TPS planning and TLD measurements. Our results show how the combination of 3D dosimetry and measurements at specific points in the phantom allowed a comprehensive view of the dose distribution and identified that care must also be paid to regions outside the PTV. |
format | Online Article Text |
id | pubmed-8504585 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-85045852021-10-18 Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis Rojas, Diana M. C. Pavoni, Juliana F. Arruda, Gustavo V. Baffa, Oswaldo J Appl Clin Med Phys Radiation Measurements This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four‐field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, and thermoluminescent dosimeters (TLD) were used for point‐dose measurements outside it. Both results were compared to the treatment planning system (TPS) dose calculation without using heterogeneity corrections to evaluate the influence of the metal in the dose distribution. MAGIC‐f gel dosimeter (Methacrylic and Ascorbic acid in Gelatin Initiated by Copper with Formaldehyde) associated with magnetic resonance imaging was used. TLD were positioned at several points at the bone metal interface and the sacrum region. The comparison of the gel measured and the TPS calculated dose distributions were done using gamma analysis (3%/3 mm), and a pass rate of 93% was achieved. The TLD dose values at the bone‐metal interface showed variations from the planned dose. However, at the sacrum region, where the beams did not intercept the prosthesis, there was a good agreement between TPS planning and TLD measurements. Our results show how the combination of 3D dosimetry and measurements at specific points in the phantom allowed a comprehensive view of the dose distribution and identified that care must also be paid to regions outside the PTV. John Wiley and Sons Inc. 2021-08-26 /pmc/articles/PMC8504585/ /pubmed/34436819 http://dx.doi.org/10.1002/acm2.13403 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Measurements Rojas, Diana M. C. Pavoni, Juliana F. Arruda, Gustavo V. Baffa, Oswaldo Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title | Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title_full | Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title_fullStr | Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title_full_unstemmed | Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title_short | Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
title_sort | gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis |
topic | Radiation Measurements |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8504585/ https://www.ncbi.nlm.nih.gov/pubmed/34436819 http://dx.doi.org/10.1002/acm2.13403 |
work_keys_str_mv | AT rojasdianamc gelandthermoluminescencedosimetryfordoseverificationsofarealanatomysimulatedprostateconformalradiationtreatmentinthepresenceofmetallicfemoralprosthesis AT pavonijulianaf gelandthermoluminescencedosimetryfordoseverificationsofarealanatomysimulatedprostateconformalradiationtreatmentinthepresenceofmetallicfemoralprosthesis AT arrudagustavov gelandthermoluminescencedosimetryfordoseverificationsofarealanatomysimulatedprostateconformalradiationtreatmentinthepresenceofmetallicfemoralprosthesis AT baffaoswaldo gelandthermoluminescencedosimetryfordoseverificationsofarealanatomysimulatedprostateconformalradiationtreatmentinthepresenceofmetallicfemoralprosthesis |