Dose rate correction for a silicon diode detector array
PURPOSE: A signal dependence on dose rate was reported for the ArcCHECK array due to recombination processes within the diodes. The purpose of our work was to quantify the necessary correction and apply them to quality assurance measurements. METHODS: Static 10 × 10 cm(2) 6‐MV fields delivered by a...